Volume heating simulation statistics for each patient model.

Model T50 (°C) T10 (°C) T90 (°C) CEM43-T10 (min) Vext (cm3)
M1          
 Baseline 41.0 43.0 39.3 8.1 6.3
 Balloon 41.0 42.9 39.3 7.5 3.5
M2          
 Baseline 41.0 42.5 39.4 4.6 2.3
 Balloon 41.0 42.5 39.4 5.1 0.9
M3          
 Baseline 41.0 42.7 39.4 6.1 4.0
 Balloon 41.0 42.8 39.4 7.4 1.6

T50: time averaged median temperature of target volume in cycle 2.

T10, T90: time averaged temperature of warmest 10% and 90% of the target volume in cycle 2.

CEM43-T10: cumulative thermal dose for cycles 1 and 2 calculated based on T10 time history.

Vext: volume of tissue outside the target volume ≥ 39.5 °C in cycle 2.