Skip to main content
. 2021 Apr 24;2:100013. doi: 10.1016/j.cccb.2021.100013

Table 5.

7T brain MRI protocol in fixed order. CSO= Semioval centre; BG= Basal Ganglia; M1= first segment of the middle cerebral artery; BOLD= Blood oxygenation level-dependent; fMRI= functional magnetic resonance imaging; FOV= Field of view; TE= Echo Time; TI= Inversion Time; TR= Repetition Time. *Total scan time for a heart rate of 80 bpm.

MR sequence Acquired resolution mm3 Time min:s Parameters
T1-weighted 1.0 × 1.0 × 1.0 01:59 FOV 250 × 250 × 190 mm3; TR 4.2 ms; TI 1297 ms; shot interval 3000 ms; flip angle 5°
2D-Qflow CSO 0.3 × 0.3 × 2.0 03:24* FOV 230 × 230 mm2; reconstructed resolution 0.18 × 0.18 mm [2]; TR 29 ms; TE 16 ms; Venc 4 cm/s
T1-weighteda 1.0 × 1.0 × 1.0 00:47 FOV 250 × 250 × 190; TR 4.1 ms; TI 1253 ms; shot interval 3000; flip angle 5°
2D-Qflow BG 0.3 × 0.3 × 2.0 03:47* FOV 170 × 170 mm2; reconstructed resolution 0.18 × 0.18 mm2;TR 28 ms; TE 15 ms; Venc 20 cm/s
2D-Qflow M1 0.5 × 0.5 × 3.0 01:47* FOV 250 × 200 mm2; TR 12 ms; TE 4.3 ms; Venc 100 cm/s
T2* whole brain 0.6 × 0.6 × 0.6 03:02 FOV 239 × 191 × 100 mm3; TR 77 ms; TE 27 ms; flip angle 20°
T2* visual cortex 0.6 × 0.6 × 1.0 02:21 FOV 160 × 169 × 29 mm3; TR 1250 ms; TE 16 ms; flip angle 40°
BOLD fMRI
visual cortexb
1.3 × 1.3 × 1.3 10:05 FOV 140 × 140 × 11 mm3; TR 880 ms; TE 25 ms
T1-weighteda 1.0 × 1.0 × 1.0 00:47 FOV 250 × 250 × 190 mm3; TR 4.1 ms; TI 1253 ms; shot interval 3000; flip angle 5°
BOLD fMRI
whole-brainc
2.0 × 2.0 × 2.0 10:00 FOV 224 × 256 × 101 mm3; TR 3000 ms; TE 25 ms
a

A fast version of the T1-weighted sequence was added multiple times to assess potential subject motion during the scan, particularly after repositioning for attaching the mask for the hypercapnic challenge.

b

The participant is presented with a visual stimulus.

c

The participant undergoes a hypercapnic challenge.