Table 1. Summary of the preclinical trial.

Variables Robot-assisted group Traditional manual group P-value
Data are presented as means ± SD.
Contrast volume, mL 24.80 ± 5.88 31.20 ± 6.43 0.172
Cumulative exposure time, min 6.60 ± 2.58 45.00 ± 10.56 0.009
Air kerma dose, mGy 4.40 ± 1.36 7.20 ± 1.72 0.034
Percutaneous coronary intervention procedure time, min 6.60 ± 1.74 3.00 ± 0.00 0.005