Functional lifespan (hours:minutes) of FreeStyle Libre (Abbott) sensors during the first experiment

Item n LSM SE P-value
By period1
 HHC 13 14:52 5:53 <0.01
 EC 10 49:38 6:37
By location2
 Ear 13 48:37 5:53 <0.01
 Leg 10 15:52 6:37
1

Sensors were placed at the beginning of hyperinsulinemic-hypoglycemic clamps (HHC) and euglycemic clamps (EC).

2

Sensors were placed near the ear or on the rear upper leg.