TABLE 1.

Diagnostic accuracies of IgE anti-latex assays at cutoffs yielding maximum diagnostic efficiency

Parameter Diagnostic accuracya
ALA (<0.35 kU/literb) CAP (<0.35 kUA/liter) HY-TEC (0.11 kU/liter)
Sensitivity 73.3 (65.7–80.9) 76.3 (69.1–83.6) 83.2∗∗ (76.8–89.6)c
Specificity 97.2 (94.8–99.6) 96.7 (94.0–99.3) 92.8 (89.0–96.6)
Positive predictive value 95.1 (90.8–99.3) 94.3 (89.9–98.7) 89.3 (83.9–94.8)
Negative predictive value 83.3 (78.3–88.4) 84.9 (80.0–89.8) 88.4 (83.8–92.9)
Efficiency 87.1 (83.4–90.9) 88.1 (84.5–91.7) 88.7 (85.2–92.3)
a

Percent; the numbers in parentheses are the 95% confidence interval (n = 311). 

b

Cutoff value yielding maximal diagnostic efficiency. 

c

Performance parameters of HY-TEC assay based on PST are significantly different from those displayed by the CAP and AlaSTAT systems. ∗, P < 0.05; ∗∗, P < 0.001.