Table 3.

Serial Quantitative Coronary Angiographic Results

Shellac + Vitamin E–Based DCB Iopromide-Based DCB P Value
Pre-procedure n = 39 n = 39
 LL, mm 20.10 ± 6.03 21.79 ± 8.61 0.318
 RVD, mm 2.41 ± 0.48 2.49 ± 0.54 0.466
 MLD, mm 0.87 ± 0.28 0.87 ± 0.34 0.936
 DS, % 62.18 ± 14.02 64.95 ± 10.72 0.331
Post-balloon angioplasty n = 39 n = 39
 RVD in-segment, mm 2.64 ± 0.46 2.64 ± 0.47 0.980
 RVD in-stent, mm 2.64 ± 0.46 2.64 ± 0.47 0.980
 MLD in-segment, mm 1.92 ± 0.42 1.90 ± 0.40 0.821
 MLD in-stent, mm 1.92 ± 0.41 1.90 ± 0.40 0.821
 DS in-segment, % 27.32 ± 11.18 27.73 ± 10.81 0.874
 DS in-stent, % 27.35 ± 11.16 27.73 ± 10.81 0.882
Post-DCB n = 39 n = 39
 RVD in-segment, mm 2.69 ± 0.54 2.80 ± 0.46 0.343
 RVD in-stent, mm 2.68 ± 0.54 2.80 ± 0.46 0.310
 MLD in segment, mm 2.13 ± 0.44 2.24 ± 0.36 0.220
 MLD in-stent, mm 2.13 ± 0.44 2.24 ± 0.36 0.220
 DS in-segment, % 20.49 ± 8.76 19.51 ± 8.86 0.625
 DS in-stent, % 20.49 ± 8.76 19.51 ± 8.86 0.625
Acute lumen gain in-segment, mm 1.26 ± 0.38 1.37 ± 0.43 0.250
Acute lumen gain in-stent, mm 1.26 ± 0.38 1.37 ± 0.43 0.249
6-mo follow-up n = 29 n = 31
 Follow-up, d 184.03 ± 24.48 187.77 ± 20.27 0.299
 RVD in-segment, mm 2.71 ± 0.50 2.65 ± 0.43 0.640
 RVD in-stent, mm 2.71 ± 0.50 2.66 ± 0.42 0.707
 MLD in-segment, mm 2.07 ± 0.44 1.90 ± 0.55 0.196
 MLD in-stent, mm 2.07 ± 0.44 1.87 ± 0.47 0.107
 DS in-segment, % 23.53 ± 10.93 30.43 ± 11.17 0.020a
 DS in-stent, % 23.53 ± 10.93 30.23 ± 10.82 0.022a
LLL in-segment, mm 0.15 ± 0.43 0.24 ± 0.39 0.299
LLL in-stent, mm 0.15 ± 0.43 0.26 ± 0.36 0.201
Binary restenosis in-segment 0 2 (6.45) 0.164
Binary restenosis rate in-stent 0 1 (3.22) 0.329

Values are mean ± SD or n (%).

DCB = drug-coated balloon; DS = diameter stenosis; LL = lesion length; LLL = late lumen loss; MLD = minimal lumen diameter; RVD = reference vessel diameter.

a

P < 0.05.