Skip to main content
. 2022 Aug 8;27(6):402–407. doi: 10.1097/MBP.0000000000000615

Table 3.

Overall performance of the Senbiosys device over the 10-S epochs

SBP DBP MBP
N = 708 N = 708 N = 708
MAE (mmHg) 5.6 2.8 2.9
ME (mmHg) 2.3 0.5 1.1
SDE (mmHg) 7.0 3.5 3.5
≤5 mmHg (%) 55 86 85
≤10 mmHg (%) 84 99 99
≤15 mmHg (%) 94 100 100
R 0.93 0.94 0.94
RMSE (mmHg) 7.3 3.6 3.6

MAE, mean absolute error; MBP, mean blood pressure; ME, mean error; r, correlation; RMSE, root mean square error; SDE, standard deviation of error.