Table 5. T1 and T2 Values Obtained for Measurements of Biphasic Anthropomorphic Phantoms (in Figure 11) after 1, 2, and 8 months postfabrication.
| GM | WM | |||||
|---|---|---|---|---|---|---|
| T1 (s) | ||||||
| time after fabrication in months | 1 | 2 | 8 | 1 | 2 | 8 |
| sample 1 | 0.35 (1) | 0.32 (2) | 0.22–0.36 | 0.19 (1) | 0.22 (3) | 0.23 (3) |
| sample 2 | 0.35–0.42 | 0.27 (2) | 0.23 (1) | 0.17 (2) | 0.22 (2) | 0.23–0.37 |
| sample 3 | 0.37 (2) | 0.26–0.36 | 0.23–0.39 | 0.145 (10) | 0.18 (2) | 0.20 (5) |
| T2 (s) | ||||||
| time after fabrication in months | 1 | 2 | 8 | 1 | 2 | 8 |
| sample 1 | 44 (2) | 43 (1) | 41 (2) | 27 (2) | 31 (1) | 34 (2) |
| sample 2 | 46 (4) | 42 (2) | 37 (4) | 23 (3) | 28 (4) | 38 (4) |
| sample 3 | 50 (7) | 45 (5) | 42 (5) | 22 (2) | 26 (2) | 33 (2) |