. 2022 Oct 15;13(10):5205–5218. doi: 10.19102/icrm.2022.13102

Figure 6:

Figure 6:

Anteroposterior and 40° left anterior oblique fluoroscopic ventricular lead positions during cardiac contractility modulation (CCM) implant. The implantable cardioverter-defibrillator lead was previously implanted high in the outflow tract. The initial CCM ventricular lead positions show dyssynchronous movement, and the tips move apart with beating (A and B). Repositioning the ventricular leads results in tandem movement without tip separation (C and D). The superior CCM lead is felt to be appropriately distanced from the base of the heart. See Videos 2A2D for more information.