Performance of Wantai SARS-CoV-2 antibody ELISA test compared to the Roche Elecsys anti-SARS-CoV-2 ELISA test as the gold standard method.

All women N = 500 HIV-positive N = 131 HIV-negative N = 360
Wantai +/ Roche +
Sensitivity (95% CI)
270/292
92.5 (88.8-95.2)
61/64
95.3 (86.9-99.0)
204/222
91.9 (87.5-95.1)
Wantai -/Roche-
Specificity (95% CI)
180/208
86.5 (81.1-90.9)
55/67
82.1 (70.8-90.4)
122/138
88.4 (81.9-93.2)
Roche+/Wantai+
Positive predictive value (95% CI)
270/298
90.6 (86.7-93.7)
61/73
83.6 (73-91.2)
204/220
92.7 (88.5-95.8)
Roche-/Wantai-
Negative predictive value (95% CI)
180/202
89.1 (84.0-93.0)
55/58
94.8 (85.6-98.9)
122/140
87.1 (80.4-92.2)