Skip to main content
. 2022 Nov 8;12:19020. doi: 10.1038/s41598-022-21317-x

Table 2.

Analytical performance of the evaluated sVNT assays using the pre-defined cutoff values and the newly calculated cutoff values determined using Youden's index.

Assay GenScript sVNT (cPass) Dynamiker sVNT Mindray NTAb sVNT
Pre-defined cutoff 30% 30% 33.1 IU/ml
Sensitivity in the infected group (95% CI) 96.2% (90.6–98.5) 97.0% (94.1–98.5) 97.1% (94.4–98.5)
Sensitivity in the vaccinated group (95% CI) 100% (86.2–100%) 97.0% (94.1–98.5) 97.1% (94.4–98.5)
Specificity (95% CI) 100% (94.8–100) 100% (92.1–100) 100% (94.9–100)
New cutoff for infected group 17.1% 14.4% 51.3 IU/ml
Sensitivity using the new cutoff (95% CI) 100% (96.5–100%) 99.2% (95.7–100) 97.6% (87.7–99.9)
New cutoff for the vaccinated group 31.4% 19.9% 21.4 IU/ml
Sensitivity using the new cutoff (95% CI) 100% (86.2–100%) 100% (97.6–100) 99.4% (96.4–100)
Specificity using the new cutoff (95% CI) 100% (94.8–100) 100% (92.1–100) 100% (95.0–100)