Skip to main content
. 2022 Oct 27;94(44):15440–15447. doi: 10.1021/acs.analchem.2c03422

Table 1. Operating Parameters of the Miniaturized High-Performance Drift Tube IMS.

temperature T 32.0–33.5 °C
pressure p 1005–1020 mbar
drift gas synthetic air (< 2 ppmv H2O, < 1 ppmv CO2)
drift gas flow 120 mLs/mina
sample gas purified air (about 3% RH and 400 ppmv CO2) containing specified traces of CWA
sample gas flow 10 mLs/mina
drift region length L 40 mm
drift region voltage UD 2.5 kV (pos.)
  –2.5 kV (neg.)
reaction region length 2 mm
reaction region voltage 550 V (pos.)
  –550 V (neg.)
ionization source X-rays
  filament current: 630 mA
  acceleration voltage: –3.4 kV
injection time 100 μs
repetition rate 40 Hz
a

mLs/min: standard milliliter per minute, mass flow at reference conditions 20 °C and 1013.25 mbar.