Skip to main content
. 2022 Oct;10(20):1133. doi: 10.21037/atm-22-4589

Table 6. AUC of FT3, FT4 and TSH in hyperthyroidism patients by Siemens and Mindray.

Index System AUC P value Criterion Sensitivity% Specificity%
FT3 Siemens 0.9449 P=0.6985>0.05 >6.33 (pmol/L) 81.72 100
Mindray 0.9440 >5.41 (pmol/L) 90.65 100
FT4 Siemens 0.9289 P=0.9321>0.05 >21.11 (pmol/L) 83.18 100
Mindray 0.9279 >23.56 (pmol/L) 78.50 100
TSH Siemens 0.9999 P=0.9989>0.05 ≤0.27 (μIU/mL) 99.07 100
Mindray 0.9999 ≤0.343 (μIU/mL) 99.07 100

AUC, area under the curve of receiver operating characteristic (ROC); FT3, free triiodothyronine; FT4, free thyroxine; TSH, thyroid stimulating hormone.