Table 4. Performance of AIDRScreening with different camera types among the included eyes.

RDR detection Topcon TRC-50DX Zeiss Visucam FF450 Topcon TRC-NW400
Value 95% CI Value 95% CI Value 95% CI
Sensitivity 83.33% 62.62–95.26% 86.38% 82.00–90.77% 87.86% 82.45–93.27%
Specificity 96.15% 80.36–99.90% 96.42% 94.58–98.26% 95.21% 91.74–98.67%
Accuracy 90.00% 78.19–96.67% 92.65% 90.61–94.70% 91.61% 88.40–94.82%
PPV 95.24% 76.18–99.88% 93.55% 90.28–96.82% 94.62% 90.74–98.50%
NPV 86.21% 68.34–96.11% 92.18% 89.57–94.78% 89.10% 84.21–93.99%
PLR 21.67 3.14–149.31 24.13 14.39–40.45 18.32 8.87–37.85
NLR 0.17 0.07–0.43 0.14 0.10–0.19 0.13 0.08–0.20
OR 127.47 172.36 140.92

CI, confidence interval; PPV, positive predictive value; NPV, negative predictive value; PLR, positive likely ratio; NLP, negative likely ratio; OR, odds ratio.