. 2022 Oct 22;11(21):6234. doi: 10.3390/jcm11216234

Table 1.

Comparison of intraocular pressure (IOP) measurement obtained in different positions both with Icare PRO (ICP) and Tono-Pen AVIA (TPA), with evaluation of difference of IOP measurements between ICP and TPA in each position. CI 95% = 95% confidence interval from the mean; SD = standard deviation; Min/Max = minimum and maximum measurement. p value = level of significance obtained by Paired T-test. Δ= Difference between ICP and TPA. All IOP parameters were reported in mmHg.

Sitting Supine Standing 1 Standing 2
ICP TPA Δ ICP TPA Δ ICP TPA Δ ICP TPA Δ
Mean 14.97 14.34 0.57 16.58 15.61 0.93 14.45 14.03 0.37 15.70 14.88 0.73
CI 95% [14.30;
15.65]
[13.57;
15.12]
[0.03;
1.12]
[15.88;
17.30]
[14.75;
16.47]
[0.29;
1.58]
[13.83;
15.07]
[13.31;
14.75]
[−0.13;
0.88]
[14.94;
16.45]
[14.04;
15.71]
[0.20;
1.24]
SD 2.71 3.10 2.10 2.83 3.44 2.49 2.49 2.89 1.96 3.04 3.36 2.03
Min 8.60 8.00 −3.80 9.10 10.00 −4.50 8.30 9.00 −5.20 8.50 9.00 −4.50
Max 22.80 22.00 6.60 23.90 22.00 7.10 20.40 22.00 5.00 23.00 23.00 6.40
p value - - 0.019 - - 0.003 - - 0.102 - - 0.002