Skip to main content
. 2022 Oct 29;14(21):5335. doi: 10.3390/cancers14215335

Table 1.

Ultrasound parameters set for standardized MicroV IOUS image acquisition on Esaote MyLab Twice ultrasound machine according to the proposed protocol.

B-Mode
Depth 44 mm
Gain 85%
Power 80%
Mechanical index 0.4
Probe frequency 6–8 MHz (tissue-enhancement imaging mode: low resolution)
Dynamic range/sharpness/density 16/1/2
PRS (persistence) 4
Post-processing 2
XView +6
Color (to be set in addition to former B-mode parameters)
Mechanical index 0.2
Color frequency 5 MHz
Color PRF 1.0 KHz
Color gain 50%
Color PRC M/2
Color PRS (persistency) 6
Doppler (to be set in addition to former Color mode parameters)
Mechanical index 0.3
SV 3/22 mmθ
Doppler Frequency 4.2 MHz
Doppler PRF 3.0 KHz
Doppler PRC 6
MicroV Doppler (to be set in addition to former Doppler mode parameters)
Color PRS (persistency) 5
HD-CFM 2
MAP COLOR MV0 or alternatively MV2
DIM-TP 56°
SENS 5
SMOOTH Medium
Clip duration Unlimited
Filter 3
Plane 0