Skip to main content
. 2022 Jul 13;15(12):2292–2299. doi: 10.1093/ckj/sfac167

Table 2.

Comparison of dialysis parameters in the six study sessions

FX80 Cordiax HD Phylther 17SD HDx Vie-18X HDx Elisio HX19 HDx Theranova 400 HDx FX80 Cordiax HDF
Blood processed (L) 124.6 ± 9.6 123.6 ± 9.9 124.2 ± 9.9 123.8 ± 9.5 124.2 ± 9.8 123.6 ± 9.7
Recirculation (%) 14.3 ± 5.5 14.9 ± 4.1 13.7 ± 4.6 15.4 ± 4.6 15.3 ± 4.8 14.2 ± 4.3
Real dialysis time (min) 283.3 ± 16.7 282.9 ± 17.8 283.0 ± 18.1 282.2 ± 17.3 282.4 ± 17.6 281.2 ± 17.6
Initial weight (kg) 69.7 ± 17.5 70.2 ± 17.6 70.0 ± 17.4 69.8 ± 17.4 70.0 ± 17.6 69.9 ± 17.6
Final weight (kg) 67.3 ± 17.1 67.7 ± 17.2 67.4 ± 17.2 67.2 ± 17.0 67.4 ± 17.3 67.5 ± 17.4
Weight gain (kg) 2.44 ± 1.06 2.59 ± 1.05 2.60 ± 0.81 2.56 ± 0.92 2.63 ± 0.99 2.38 ± 0.96
Initial hematocrit (%) 30.8 ± 4.9 29.7 ± 4.7 30.9 ± 4.7 30.8 ± 4.3 30.6 ± 4.5 30.9 ± 4.8
Final hematocrit (%) 36.2 ± 5.5 35.7 ± 5.8 36.7 ± 5.8 36.5 ± 5.3 36.9 ± 5.6 36.9 ± 6.2
Arterial pressure. (mmHg) –229 ± 24 –227 ± 22 –230 ± 22 –229 ± 22 –228 ± 25 –228 ± 25
Venous pressure. (mmHg) 199 ± 23 192 ± 24 197 ± 23 202 ± 28 201 ± 21 200 ± 23
TMP (mmHg) 30.0 ± 5.1 31.4 ± 5.0 29.2 ± 3.5 29.0 ± 4.2 28.5 ± 2.6 191.3 ± 32.1a
Substitution volume (L) Not applicable Not applicable Not applicable Not applicable Not applicable 30.4 ± 4.1
a

P < .001 vs all others dialysis treatments. TMP, transmembrane pressure. Data are presented as mean ± standard deviation.