. 2022 Nov 15;17(11):e0277300. doi: 10.1371/journal.pone.0277300

Multicenter validation of a machine learning phase space electro-mechanical pulse wave analysis to predict elevated left ventricular end diastolic pressure at the point-of-care

Sanjeev P Bhavnani 1,*, Rola Khedraki 2, Travis J Cohoon 1, Frederick J Meine III 3, Thomas D Stuckey 4, Thomas McMinn 5, Jeremiah P Depta 6, Brett Bennett 7, Thomas McGarry 8, William Carroll 9, David Suh 10, John A Steuter 11, Michael Roberts 12, Horace R Gillins 13, Ian Shadforth 13, Emmanuel Lange 14,15, Abhinav Doomra 14,15, Mohammad Firouzi 14,15, Farhad Fathieh 14,15, Timothy Burton 14,15, Ali Khosousi 14,15, Shyam Ramchandani 14,15, William E Sanders Jr 13, Frank Smart 16
Editor: Chim C Lang17
PMCID: PMC9665374  PMID: 36378672

Abstract

Background

Phase space is a mechanical systems approach and large-scale data representation of an object in 3-dimensional space. Whether such techniques can be applied to predict left ventricular pressures non-invasively and at the point-of-care is unknown.

Objective

This study prospectively validated a phase space machine-learned approach based on a novel electro-mechanical pulse wave method of data collection through orthogonal voltage gradient (OVG) and photoplethysmography (PPG) for the prediction of elevated left ventricular end diastolic pressure (LVEDP).

Methods

Consecutive outpatients across 15 US-based healthcare centers with symptoms suggestive of coronary artery disease were enrolled at the time of elective cardiac catheterization and underwent OVG and PPG data acquisition immediately prior to angiography with signals paired with LVEDP (IDENTIFY; NCT #03864081). The primary objective was to validate a ML algorithm for prediction of elevated LVEDP using a definition of ≥25 mmHg (study cohort) and normal LVEDP ≤ 12 mmHg (control cohort), using AUC as the measure of diagnostic accuracy. Secondary objectives included performance of the ML predictor in a propensity matched cohort (age and gender) and performance for an elevated LVEDP across a spectrum of comparative LVEDP (<12 through 24 at 1 mmHg increments). Features were extracted from the OVG and PPG datasets and were analyzed using machine-learning approaches.

Results

The study cohort consisted of 684 subjects stratified into three LVEDP categories, ≤12 mmHg (N = 258), LVEDP 13–24 mmHg (N = 347), and LVEDP ≥25 mmHg (N = 79). Testing of the ML predictor demonstrated an AUC of 0.81 (95% CI 0.76–0.86) for the prediction of an elevated LVEDP with a sensitivity of 82% and specificity of 68%, respectively. Among a propensity matched cohort (N = 79) the ML predictor demonstrated a similar result AUC 0.79 (95% CI: 0.72–0.8). Using a constant definition of elevated LVEDP and varying the lower threshold across LVEDP the ML predictor demonstrated and AUC ranging from 0.79–0.82.

Conclusion

The phase space ML analysis provides a robust prediction for an elevated LVEDP at the point-of-care. These data suggest a potential role for an OVG and PPG derived electro-mechanical pulse wave strategy to determine if LVEDP is elevated in patients with symptoms suggestive of cardiac disease.

Introduction

‘Phase space’ is a concept based on dynamical systems theory in which possible states of a given object such as position and velocity are represented with each state corresponding to one unique point in phase space [1]. While originating from mechanical systems, it has application to cardiovascular physiology. In one possible application, while systolic dysfunction is characterized by reduced ejection fraction, additional modalities are required to adjudicate dysfunction that is limited to diastole, with the aim of estimating left ventricular (LV) filling pressures. Left ventricular end diastolic pressure (LVEDP) is of distinct interest. The measurement of LVEDP, whether in the presence of reduced or preserved ejection fraction is complex and commonly characterized by multimodality diagnostic imaging. For example, elevation in Brain Naturetic Peptide (BNP) [2, 3] and fixed ratios based on echocardiography (spectral Doppler and Tissue Doppler derived E/e’) [4] are used to classify if left atrial pressure is elevated or not. Several recent studies have aimed to predict diastolic dysfunction (i.e., intracardiac pressure elevation) using ML approaches, such as from CNN analysis of echocardiographic beat variability [5] and clustering of echocardiographic markers to understand the patterns of diastolic dysfunction across patients with symptomatic CVD [6]. While such developments are promising in the characterization of myocardial function, the prediction of LV pressure elevation as a binary classification (elevated or not elevated) across a spectrum of LV pressures that can be used to guide downstream testing and treatment is of value.

In this context, phase space is a continuous measurement that simultaneously captures data related to electromechanical and pulse-wave signals over successive cardiac cycles, with the resultant biopotential plot being a large-scale data representation of myocardial function and is unique for any given person [1]. The benefits of such an approach are that it captures signals of myocardial function and dysfunction through high fidelity, time-series data collection that cannot be quantified by conventional non-invasive imaging or laboratory testing modalities [7, 8].

The physiologic findings of the failing heart that result in elevations in LV filling pressures, LV end diastolic pressure (LVEDP) and left atrial pressure are commonly determined by electrocardiographic and echocardiographic findings of atrial and ventricular remodeling [9], functional changes in diastolic relaxation [10], and changes in flow dynamics [5]. Given a highly heterogenous association between symptoms and the presence of cardiac dysfunction, especially in prevalent conditions such as those with heart failure with preserved EF (HFpEF), new modalities that leverage machine learning (ML) have emerged as potential tools to predict diastolic properties [11, 12] and ejection fraction (EF) [13] through computational approaches including neural network analysis of electrocardiographic intervals and wavelet transformation to predict myocardial function and relaxation.

Similarly, we have previously demonstrated the diagnostic accuracy of a ML approach to predict obstructive coronary artery disease (>70% luminal stenosis) from a cardiac phase space analysis. The predictive algorithm was trained and validated with tomographic, voltage-gradient features that were paired with the degree of coronary stenosis defined at angiography [7]. An approach such as this provided a method to collect data at the time of an outcome of interest and provided a pathway to evaluate cardiac dysfunction at the point-of-care [8, 14, 15]. In this context, we investigated a novel ML algorithm based on electromechanical features that was derived non-invasively from orthogonal voltage gradient (OVG) and photoplethysmography (PPG) to predict an elevated LVEDP among symptomatic patients referred for cardiac catheterization, and herein report the findings from the multi-center, prospective validation cohort.

Results

Study population

Overall, 1,023 study subjects were consecutively enrolled (Fig 1). After exclusions, the final study cohort consisted of 684 subjects with paired signals to LVEDP measurements and stratified into three LVEDP categories, ≤12 mmHg (N = 258), LVEDP 13–24 mmHg (N = 347), and LVEDP ≥25 mmHg (N = 79). Histogram of LVEDP in the study population can be found in Fig 2. Signals were acquired by trained nursing staff in a hospital-based setting (catheterization holding, cardiac unit, or observation area) in 671 (98.1%) or a physician’s clinic in 13 (1.9%), within 7-days prior to angiography [630 (92%) on the same-day; 54 (8%) from 1–7 days].

Fig 1. Flow of subjects in the study.

Fig 1

Fig 2. Histogram of study population LVEDP.

Fig 2

Demographics characteristics for the overall population, study cohort as well as LVEDP groups are listed in Table 1. Within the study cohort, the mean age was 63 years and 45% were women. One third of the population had diabetes with greater than 70% with hypertension and/or hyperlipidemia. The mean EF was 60% with 93% (N = 186) with preserved EF >50%. 38% had obstructive CAD at angiography. Multivariate clinical predictors of an elevated LVEDP can be found in S1 File.

Table 1. Demographics.

Characteristic Overall Study Population Development + Validation N = 1,956 Development Cohort N = 1,272 Validation Cohort N = 684 p-value Development vs. Validation Study Cohort: LVEDP≤12 or LVEDP≥25 N = 337 LVEDP≤12 N = 258 LVEDP≥25 N = 79 LVEDP 13–24 N = 347 p-value LVEDP ≤12 vs ≥25
Clinical Variables
Age (years)
 Mean ± STD 52 ± 19 46 ± 20 63 ± 10 <0.001 63 ± 11 64 ± 10 61 ± 12 62 ± 10 0.04
 ≥ 60 years (%) 880 (45%) 445 (35%) 321 (47%) <0.001 172 (51%) 137 (53%) 34 (43%) 153 (44%) 0.13
 Range 18–91 18–88 30–91 30–90 36–90 30–83 31–91
Ejection Fraction
 Measured (%) 978 (50%) 560 (44%) 428 (63%) -- 199 (59%) 157 (61%) 43 (54%) 229 (66%) 0.60
 Mean ± STD 59 ± 8 59 ± 8 60 ± 7 -- 60 ± 7 60 ± 6 59 ± 10 60 ± 8 0.56
 >50% 892 (91%) 501 (90%) 391 (91%) -- 186 (93%) 148 (94%) 38 (90%) 205 (90%) 0.81
 ≤ 50% 90 (9%) 53 (10%) 37 (9%) -- 13 (7%) 9 (6%) 4 (10%) 24 (10%) 0.81
 ≤ 40% 40 (4%) 27 (5%) 13 (3%) -- 6 (3%) 3 (2%) 3 (7%) 7 (3%) 0.31
 ≤ 30% 19 (2%) 13 (2%) 6 (1%) -- 3 (2%) 1 (1%) 2 (5%) 3 (1%) 0.29
Women (%) 911 (47%) 612 (48%) 299 (44%) 0.070 152 (45%) 101 (39%) 51 (65%) 147 (42%) <0.001
History of Hypertension (%) 1014 (52%) 519 (41%) 495 (72%) <0.001 244 (72%) 184 (71%) 60 (76%) 251 (72%) 0.51
History of Hyperlipidemia (%) 1005 (51%) 512 (40%) 493 (72%) <0.001 246 (73%) 194 (75%) 52 (66%) 247 (71%) 0.12
History of Diabetes (%) 460 (24%) 246 (19%) 214 (31%) <0.001 102 (30%) 75 (29%) 27 (34%) 112 (32%) 0.48
Smoker* (%) 728 (37%) 376 (30%) 352 (51%) <0.001 178 (53%) 139 (54%) 39 (49%) 174 (50%) 0.54
Body Mass Index
 Mean ± STD 30 ± 7 29 ± 7 32 ± 7 <0.001 32 ± 7 30 ± 6 37 ± 8 32 ± 6 <0.001
 ≥ 30 (%) 899 (46%) 520 (41%) 379 (55%) <0.001 176 (52%) 111 (43%) 65 (82%) 203 (59%) <0.001
 <30 1056 (54%) 751 (59%) 305 (45%) <0.001 161 (48%) 147 (57%) 14 (18%) 144 (41%) <0.001
Race/Ethnicity
Caucasian (%) 1440 (74%) 861 (68%) 579 (85%) <0.001 280 (83%) 221 (86%) 59 (75%) 299 (86%) 0
Black or African American (%) 396 (20%) 315 (25%) 81 (12%) <0.001 44 (13%) 27 (10%) 17 (22%) 37 (11%) 0
Other Races (%) 120 (6%) 96 (8%) 24 (4%) <0.001 13 (4%) 10 (4%) 3 (4%) 11 (3%) 0
Symptoms
 Chest Pain During Exercise 577 (29%) 384 (30%) 384 (56%) <0.001 193 (57%) 147 (57%) 46 (58%) 191 (55%) 0.95
 Chest Pain at Rest 404 (21%) 266 (21%) 282 (41%) <0.001 138 (41%) 99 (38%) 39 (49%) 144 (41%) 0.11
 Dyspnea During Exercise 646 (33%) 438 (34%) 426 (62%) <0.001 208 (62%) 153 (59%) 55 (70%) 218 (63%) 0.13
 Dyspnea at Rest 199 (10%) 127 (10%) 145 (21%) <0.001 72 (21%) 49 (19%) 23 (29%) 73 (21%) 0.08
Medications
 ACE inhibitor 378 (19%) 185 (15%) 193 (28%) <0.001 92 (27%) 72 (28%) 20 (25%) 101 (29%) 0.76
 ARB 302 (15%) 148 (12%) 154 (23%) <0.001 77 (23%) 55 (21%) 22 (28%) 77 (22%) 0.29
 Diuretic†† 323 (17%) 187 (15%) 136 (20%) 0.004 65 (19%) 36 (14%) 29 (37%) 71 (20%) <0.001
 Calcium Channel Blocker 313 (16%) 163 (13%) 150 (22%) <0.001 70 (21%) 48 (19%) 22 (28%) 80 (23%) 0.11
 Beta Blocker 563 (29%) 286 (22%) 277 (40%) <0.001 142 (42%) 102 (40%) 40 (51%) 135 (39%) 0.11
 Aldosterone Receptor Antagonist 9 (0%) 0 (0%) 9 (1%) <0.001 4 (1%) 3 (1%) 1 (1%) 5 (1%) 1.00
 Statins 791 (40%) 388 (31%) 403 (59%) <0.001 200 (59%) 158 (61%) 42 (53%) 203 (59%) 0.25
Angiographic Findings
 Presence of obstructive CAD^ 497 (36%) 253 (36%) 244 (36%) -- 129 (38%) 110 (43%) 19 (24%) 115 (33%) 0.005
Hemodynamic Variables
LVEDP
 Mean ± STD 16 ± 7 (N = 1380) 16 ± 7 (N = 696) 16 ± 7 -- 14 ± 9 9 ± 3 29 ± 4 18 ± 3 <0.001
 Range 0–45 0–45 0–38 0–38 0–12 25–38 13–24
LVSP
 Mean ± STD 129 ± 24 (N = 1002) 130 ± 24 (N = 395) 129 ± 24 (N = 607) -- 126 ± 25 (N = 283) 119 ± 21 (N = 208) 143 ± 25 (N = 75) 132 ± 23 (N = 324) <0.001
 Range 15–221 40–221 15–209 22–209 22–209 88–196 15–208

Ejection fraction measured and reported by ventriculography at the time of angiography

*past or present smoker

including combination medications

††Avalide, Dyazide, Furosemide, Hydrochlorothiazide, Maxzide, Triamterene

^at least one coronary lesion with a stenosis of ≥70% and/or FFR<0.8 and/or iFR <0.89 on angiography

Permutation feature importance and exemplar features

A permutation analysis was performed to determine feature importance in LVEDP elevation prediction, and the top 30 most contributive features were grouped by family (S2 File). The most contributive signal-based feature family was PPG indicators, followed by OVG spectral and phase space analysis. The most contributive feature within the PPG indicator family was the maximum of the PPG pulse base (S3 File). Fig 3 illustrates four exemplar features included in the machine-learned predictor, including the atrial depolarization duration, PPG pulse base, and ventricular repolarization. Detailed descriptions of the other features of importance can be found in S3 File.

Fig 3. Exemplar features.

Fig 3

a) the variation in the atrial depolarization duration is a feature extracted from the OVG signal in the time domain, b) quantification of the atrial depolarization vector in phase space, c) quantification of PPG pulse base amplitude and d) ventricular repolarization in band-pass filtered phase space.

Primary outcome—Performance of the machine-learned predictor

All results (primary and secondary) used the ensembled model as a single assessment of algorithm performance on the blinded validation cohort. Testing of the machine-learned predictor as a continuous measurement demonstrated an AUC of 0.81 (95% CI 0.76–0.86) for algorithmic performance (Fig 4) and corresponded to a sensitivity and specificity of 82% (95% CI: 72–90%) and 68% (95% CI: 61–72%), respectively. S4 File contains the 2x2 cross tabulation for the prediction of an elevated LVEDP based on the sensitivity and specificity determined from the primary analysis.

Fig 4. Performance of the machine-learned predictor.

Fig 4

Secondary outcomes

Predictive performance for an elevated LVEDP among a propensity matched cohort

Among a propensity-matched cohort (N = 79 pairs of study subjects) between elevated and non-elevated LVEDP based on age and gender (S5 File), the machine-learned predictor demonstrated a similar result to the primary analysis for the prediction of an elevated LVEDP, AUC 0.79 (95% CI: 0.72–0.86, Fig 5).

Fig 5. Performance of the machine-learned model when propensity matching based on age and gender.

Fig 5

Determine the predictive performance for an elevated LVEDP across a spectrum of comparative LVEDP thresholds

Using a constant elevated definition of LVEDP ≥ 25 mmHg and varying the definition of non-elevated across LVEDP values (in 1 mmHg increments), the machine-learned predictor demonstrated an AUC ranging from 0.79–0.82 (Fig 6a) with corresponding specificity between 59%-69% (using the constant, predefined elevated LVEDP threshold yielding a sensitivity of 82%). Fig 6b illustrates the effect of varying both the definitions of LVEDP elevation and non-elevated, demonstrating a consistent performance, based on AUC, for the prediction of an elevated LVEDP at a threshold value of 25 mmHg.

Fig 6. Performance while varying the definition of LVEDP elevation and non-elevation.

Fig 6

a) the definition of elevation is held constant at ≥25mmHg, while the definition of non-elevation is varied from ≤12mmHg to ≤24mmHg, and the ROC curves assessed for each scenario. b) the definition of both elevation and non-elevation are varied from ≥13-25mmHg and ≤12-24mmHg, respectively, and the corresponding AUCs represented as a heatmap.

Sub-group performance

Among the predefined sub-groups, the machine-learned predictor demonstrated an adequate diagnostic accuracy between group stratifications (Fig 7). While the predictive accuracy for an elevated LVEDP was similar among cohorts with obstructive and non-obstructive (p = 0.31), there was a statistically significant difference with greater predictive accuracy among cohorts with preserved EF compared to low EF (p = 0.03).

Fig 7. Subgroup analyses of the performance of the machine-learned predictor.

Fig 7

Safety and adverse events

Testing of the machine-learned model within the pre-specified safety analysis among a healthy cohort without CV disease, at a threshold sensitivity of 82% reported in the primary analysis, demonstrated a specificity of 95% (95% CI: 90–97%). The corresponding 2x2 tabulation and AUC can be found in S4 File.

No adverse events related to device use were reported during the study.

Bayesian analysis and simulations for net reclassification index between BNP and the machine-learned predictor

A Bayesian analysis of the post-test probability of an elevated LVEDP based on a range of pre-test probabilities and distributed according to the machine learned predictor, and the two simulated BNP performances in a matching population–a BNP threshold of 150 pg/ml and a BNP threshold of 50 pg/ml—are illustrated in Fig 8a–8c; respectively. Using the approximate values of the constraining statistics for the two BNP data sets (non-cardiac symptoms and obese HFpEF, see Methods), the simulation yielded an AUC of 0.69 (95% CI: 0.61–0.71) for BNP prediction of an elevated LVEDP with the corresponding sensitivity, specificity, positive and negative likelihood ratios for the thresholds of a BNP of 150 and 50 pg/ml reported within Fig 8. The NRI of the ML predictor using a posterior probability of a BNP at a threshold of 150 and 50 pg/ml was 0.24 (95% CI: 0.18–0.30) and 0.38 (95% CI: 0.33–0.44), respectively.

Fig 8. Relationship between pre-test (prior) probability and post-test (posterior) probabilities.

Fig 8

a) the machine-learned predictor, b) BNP when greater than 150pg/ml, or c) BNP when greater than 50pg/ml. A positive test is shown in red, and a negative test in green. The diagonal dashed black line represents no change from the pre-test to post-test probability. The vertical dashed black lines represent intermediate to high pre-test probabilities (vertical dashed lines at 30%, 50% and 70%) from left to right. The post-test probabilities were calculated based on a varying pre-test probability, and constant sensitivity, specificity, and corresponding likelihood ratios.

Discussion

There is a growing need for new methods to measure LV filling pressures. Recent studies have used ML approaches that analyze echocardiography data to predict diastolic dysfunction [5, 6]. While the prospect for discovery is promising, the application of any new analytic technique requires robust methodologies for validation. In this context we employed a trial design of prospective validation [16] within a multicenter study. Prospective data collection permitted the validation dataset to be blinded from the training dataset. This is important because blinding may limit common biases such as spectrum bias and measurement bias between training and validation datasets, and to balance clinical characteristics between both datasets [14]. Towards mitigating bias, we recruited a diverse cohort of patients across multiple healthcare centers and geographies and aimed to enroll study subjects that are representative of a real-world population with a clustering of cardiac risk factors, and various ethnicities. Overall, half of the participants recruited were women, nearly 50% with a BMI ≥ 30 (mean of 36) and greater than 90% with preserved EF (mean EF of 61%), a triad of findings where an accurate assessment of LV filling pressures by conventional testing such as BNP [17] and echocardiography [4, 18] have marginal accuracy and vary significantly across those with symptoms of HF. This is particularly true in HFpEF given the heterogeneity of myocardial dysfunction (i.e. ischemic vs non-ischemic etiologies), the cardiopulmonary response to increased afterload and/or preload [19] and the phasic changes in left atrial function that are variable across individuals [9]. Our observations of a high incidence (34%) of an elevated LVEDP among symptomatic patients referred to angiography for the evaluation of ischemic heart disease; however, did not have evidence of obstructive CAD is equally important as it may reflect the underdiagnosis of HF in an ambulatory cohort, and those referred for further cardiovascular testing.

Our hypothesis that electromechanical pulse wave features predict myocardial dysfunction is an extension of the hypothesis that the progression from normal myocardial mechanics to pathologic atrial and ventricular remodeling is a result of rising LV filling pressure. Whereas atrial enlargement and ventricular remodeling from alternations in myocardial tension and strain can be considered mechanical features of a pressure loaded left ventricle [20], we postulate that a high dimensional dataset captured from voltage gradients and photoplethysmography can accurately represent myocardial electromechanical function. In support of this argument, wavelet transformation and the mathematical conversion of an ECG into a normalized energy distribution (depicted by a color spectrum of myocardial energy) has recently emerged as a computational modeling and ML method for the prediction of diastolic dysfunction. Potter [11] and Sengupta [12], paired ECG data with echocardiographic data of diastolic abnormalities such as E/e’> 14, left atrial enlargement, and abnormal LA volume index among 398 and 188 patients at risk of HF, respectively. Using supervised and unsupervised ML including random forest classifiers on 250–650 wavelet features, testing of the ML algorithm on a validation data set demonstrated a high diagnostic accuracy (AUC 0.83–0.91) for the prediction of diastolic dysfunction, with a diagnostic performance that was greater than clinical prediction alone.

The present study extends Potter and Sengupta’s results to a time-series, electromechanical and perfusion dataset with training and validation on direct LVEDP measurements. Within our dataset, in an LV with normal EF that is under high pressure [10], electromechanical features such as variation in atrial depolarization duration, ventricular repolarization in phase space plausibly represents physiologic findings of elevated left atrial pressure [9] and lusitropic changes of diastolic relaxation [11, 12]; respectively. PPG derived feature and the measurement of the pulse wave base may represent a systolic time interval of isovolumetric contraction as this nadir point in the pulse-wave is associated with the lowest photoabsorption immediately prior to onset of systole; a time point of interest that has been associated with HF [21]. Our method of high frequency data capture analyzed in phase space and the corresponding electromechanical features are unique in any given study subject. Once all features are evaluated from a patient’s signal, the machine-learned model processes the feature values to yield a continuous score representing the risk of LVEDP elevation. In contrast to BNP or echocardiography which use a threshold or binary values of elevated or non-elevated LV pressures, the high number of features used in this analysis creates a unique signature of LVEDP that is specific to that individual at the N-of-1 level.

One pertinent question leading from the present analysis, is how our results are translated within the continuum of diagnostic tests to determine the presence of elevated LV pressure among symptomatic patients, particularly those with preserved EF. Several point-of-care diagnostic tests are available to diagnose acute HF including chest radiography, BNP/NT-pro-BNP, handheld echocardiography/lung ultrasound, and bioimpedance [22]. Various studies evaluating such point-of-care tests have largely concluded that lung ultrasound and echocardiography have utility to differentiate HF symptoms from non-HF symptoms [22]. While useful, these tests require trained individuals to acquire and interpret cardiopulmonary images and can have limited diagnostic accuracy as they are dependent on patient characteristics such as body habitus and the user experience with, point-of-care imaging devices. Our findings of high diagnostic accuracy for prediction of elevated LVEDP when compared to a range of non-elevated LVEDP such as normal (≤12 mmHg) and mid-range (13–24 mmHg) is potentially valuable for 2 main reasons: 1) it supports that our method for data capture and analysis of electromechanical data is robust and that our features are those data representations associated with elevated LVEDP; and 2) that LVEDP at a threshold of ≥25 mmHg is representative of an elevation that is clinically relevant as our study population was derived from symptomatic patients requiring cardiac catheterization. The latter is important in the setting of high-risk HF patients such as those at risk for HF re-hospitalizations. If we assume that such patients have, at minimum, a 50% (intermediate) to 70% (high) pre-test probability of an elevated LVEDP, when used sequentially with BNP, our Bayesian simulation demonstrate that the ML predictor reclassifies a BNP of > 150 pg/ml (NRI of 0.24) from 59–77% to a post-test probability of 79–90%, with the greatest reclassification margin within the intermediate (30–50%) pre-test probability group. Such results may have clinical utility to more accurately triage HF patients at the point-of-care for further testing and to identify patients with HFpEF.

Limitations

We identified 14% (36/258) of subjects with non-elevated LVEDP were taking a diuretic at the time of enrollment that may have impacted the performance of the ML predictor. When compared to overall study population the specificity was similar within this cohort (68% vs 64%, p = 0.57) and when the analysis was re-run when excluding this cohort, there was no difference in overall specificity. While we contend that diuretics are an important factor when considering the measurement and prediction of LVEDP, the small number of subjects in this group does not permit us to determine its impact on performance within the study population as presented.

Our study population is intrinsically limited by the recruitment methodology, which was subjects referred to left heart catheterization for assessment of obstructive CAD using coronary angiography, and specifically the subgroup where the treating physician chose to measure the LVEDP. We employed this study methodology to ensure that subjects had a catheterization-confirmed elevated LVEDP, but at the limitation of subjects referred for the evaluation of obstructive CAD. While this may introduce sample bias, we found significant CAD in only 38% of the overall study cohort, a higher incidence of obstructive CAD was observed in subjects with non-elevated LVEDP compared to those with elevated LVEDP (43% vs 24%). Upon subgroup analysis, there was difference in algorithmic performance among those with or without obstructive CAD.

Overfitting, and conversely generalizability, are critical aspects of machine learning and when a large number of features are used for model development. The use of an ensemble, as is the case, does not increase the likelihood of overfitting but rather mitigates it by reducing the dependence on a single model. The methods employed to avoid overfitting could include the use of cross-validation within the development data, using simple models with regularization and penalty terms, and testing the performance of the model on unseen data (doing so only once). With respect to the models in particular, first, each model is exposed to an average of 149 features (with a range of 89–194, S10 File). Second, the model hyperparameters were conservatively designed to mitigate the possibility of overfitting (S10 File). For example, four of the models were Random Forest, which intrinsically limit overfitting by only allowing each component tree access to the square root of the total number of features, and by bootstrap sampling training subjects so that every component tree only has access to a subset of the entire training set. Overfitting was additionally controlled through the use of the maximum tree depth hyperparameter. Deep trees with many splits increase the likelihood of overfitting, and therefore the depth was limited to 3–7. Other model types (Elastic Net and XGBoost) were also designed conservatively. Finally, the ultimate test of overfitting is the performance on unseen blinded data, which yielded an AUC of 0.81. In conclusion, through the analysis of the algorithm, and the performance on unseen blinded data, overfitting did not occur.

Conclusions

We validated a machine learning algorithm of electromechanical pulse wave features to predict an elevated LVEDP among symptomatic patients with a precise measurement of LVEDP. Such techniques to quantify intracardiac pressure with machine learning on large datasets acquired with a portable digital device provides a new method to determine the presence or absence of HF. These data suggest a potential role for a novel OVG and PPG derived electromechanical diagnostic test for the prediction of an elevated LVEDP at the point-of-care.

Methods

Trial design

Enrollment in the overall IDENTIFY trial began on December 10th 2018 with 3,486 participants consecutively enrolled as of April 2021 and was primarily executed to develop a machine-learned predictor to determine the presence of obstructive coronary artery disease (CAD) defined at cardiac catheterization. A cohort analysis (N = 606) using a phase-space ML approach to predict CAD using the same inclusion/exclusion criteria as in this present study has been previous published [7]. The present results are reported according to STARD guidelines [23] (S6 File). The study was approved by a centralized IRB (Western IRB #20183107, now known as WIRB-Copernicus Group). It was initially released on clinicaltrials.gov on January 1st 2019 (NCT #03864081), and was performed at 15 healthcare institutions in the United States (S7 File). A preliminary analysis of the LVEDP development cohort was presented at the 2020 Scientific Sessions of the American College of Cardiology [24].

Study population—Development and validation data

Data sources

The data sources for the study population included patients who were referred for angiography at the discretion of their treating physicians and for the evaluation of symptoms suggestive of CAD. Patients provided written informed consent to participate in the study. Inclusion and exclusion criteria have been previously published [7] and can be found in S8 File.

Development and validation groups

The study population was derived from a pooled individual patient-level analysis stratified by unique time points and into development and validation cohorts. The development cohort was comprised of symptomatic patients referred to cardiac catheterization for the evaluation of CAD and consecutively enrolled between April 2017 –December 2017 (N = 696) and included asymptomatic individuals without CVD (N = 576). A separate validation cohort prospectively consecutively enrolled symptomatic patients referred to cardiac catheterization between March 2019 –November 2019 (N = 1,023). The data sources for the development and validation cohorts were separate with no data from the validation group used for development. Therefore, the validation group is considered blinded.

Primary and secondary study objectives

The primary objective was to develop and validate an ML algorithm for the prediction of an elevated LVEDP ≥ 25 mmHg. LVEDP was measured invasively at the time of cardiac catheterization using conventional techniques for left ventricular pressure assessments. For the primary objective, an analysis using threshold of LVEDP of 25 mmHg (study cohort) was chosen and compared to individuals with a normal LVEDP defined as ≤ 12 mmHg (control cohort). These thresholds were selected to reflect those LVEDP measurements that are likely to be sufficiently high to result in symptoms (elevated LVEDP) or absence of symptoms (normal LVEDP), as they relate to a spectrum of symptoms among CV patients undergoing angiography, and those with obstructive and non-obstructive CAD [25, 26].

Secondary objectives included those analyses to refine the primary objective within the following 5 categories:

  1. Performance of the machine-learned predictor for an elevated LVEDP ≥ 25 mmHg among a propensity-matched cohort (scoring using age and gender).

  2. Performance of the machine-learned predictor for an elevated LVEDP across a spectrum of comparative LVEDP thresholds (<12 through 24 at 1 mmHg increments).

  3. Sub-group performance stratified by age (<60 vs. ≥60), gender, comorbidities, EF (<50% vs ≥50%), CAD status determined at angiography [obstructive CAD (defined as ≥1 lesion with a stenosis of 70%; or ≥1 lesion with a fractional flow reserve of ≤0.80; or ≥1 lesion with an instantaneous wave-free ratio of ≤0.89) vs non-obstructive CAD], ethnicity, and enrollment site (stratified by number of study subjects enrolled N≥50 vs N<50).

  4. Safety analysis and predictive accuracy of the machine-learned predictor using a healthy control cohort to determine the specificity and negative predictive value of the algorithm, and;

  5. Bayesian analysis to determine the post-test (i.e., posterior) probability of the machine-learned predictor based on varying the pre-test probability (i.e., low, intermediate, and high prior probability of elevated LV filling pressures) among symptomatic patients.

Acquisition system description

The acquisition system (CorVista Capture device) simultaneously collects two modalities of time series data: orthogonal voltage gradient (OVG) data, representing cardiac electro-mechanical activity analyzed in phase space, and photoplethysmography (PPG) data representing blood volume changes as a measurement of distal perfusion.

Consecutive study subjects underwent signal acquisition immediately prior to angiography or within seven days prior to the procedure. In addition to OVG and PPG data, the device also captured patient-specific metadata (gender, age, height and weight). Signal data was acquired for 3.5 minutes.

Raw data collection

The OVG signal is collected using electrodes attached to the skin (S9 File). Specifically, the signal is acquired at 8kHz (i.e., 8,000 samples per second, with each consecutive pair of samples separated by 0.000125 seconds) from seven electrodes at an amplitude resolution of 0.024 microvolts. The signal originates from three bipolar pairs of electrodes collecting data from the coronal, sagittal and transverse planes, and the seventh electrode acting as the reference. See S9 File for further details. Similar to existing signal collection methods, the OVG measures the biopotential at the surface of the skin caused by cardiac electrical activity. While the OVG signal acquisition resembles ECG, it differs with greater sampling frequency (conventional ECG sampling frequency of 500-2000Hz) by a factor of 4–16 and provides broader spatial information due to the orthogonal lead configuration and vectors along different planes of the body [27, 28].

The OVG biopotential data is represented within a three-dimensional phase space, where the parameters of the phase space are defined by the three bipolar orthogonal acquisition channels. Specifically, the amplitudes of three voltage gradient data points from the three channels form a three-dimensional coordinate within the phase space. As the signal processes through time, it traces a phase space trajectory. The PPG signal contains red and infrared light components, both collected at 500Hz via a finger clip sensor. The pulse wave is captured as the absorption of light in the tissue varies based on changes in cardiac activity.

Development & validation approach for the machine learning predictor

Development and validation of a machine learning predictor occurred in two distinct phases. In the first phase, the machine learning predictor was trained using the development dataset. Upon completion, the machine-learned predictor was finalized such that no further modifications were permitted. Then, in the second phase, the machine learning predictor was tested in the blinded validation cohort and the performance was assessed.

Signal processing and development of the machine-learned predictor

The sequence for processing a patient’s data to generate the ML prediction occurred in four steps.

Step 1: Confirmation of signal quality

As an initial processing stage, the signal was confirmed to have adequate quality to proceed through the next three steps. Signal quality acts to check for the presence of noise generated by common sources in a clinical environment. The signal quality assessment has been previously published [8], and will be summarized herein. The OVG signal is examined for the presence of powerline noise, which is the electrical noise at the frequency at which alternating current (AC) power is delivered (i.e., to electrical outlets, etc.); specifically, this is 60 cycles per second (Hz) in North America. The OVG signal is also examined for high-frequency noise. A SNR of 57 was considered acceptable for powerline noise, and of 19 for high frequency noise. The PPG signal is examined for sensor saturation, which can occur when the light emitted from the LED on one side of the finger clip directly enters the sensor on the opposite site of the clip without transiting the finger. The light is strong because it isn’t attenuated by the finger, and therefore an optical value is registered that exceeds the maximum measurable value. Excessive occurrence of this situation reduces the physiological information in the signal, and results in a prompt to attempt to reacquire the signal. SNR is not applicable to this score because the occurrence is transient.

Step 2: OVG and PPG feature extraction

After signal quality assessment, features were extracted from the signal. We defined a feature as a characteristic of the data that is automatically measurable on any acquired signal. A spectrum of feature domains of signal characterization were used in the present analysis and include the dynamics of the OVG and PPG signals in isolation; the synchronization dynamics of the OVG and PPG signals, the spectral properties of each signal modality; deviations of the OVG signal from subject-specific models; both conventional time-domain features and variations of those features; phase space features; PPG pulse-wave indicators; and approximation of a patient’s respiration waveform (see S3 File). For example, the variation in the atrial depolarization duration is a feature extracted from the OVG signal in the time domain. Specifically, the duration of the atrial depolarization is measured on each cardiac cycle which forms a distribution of durations across the length of data acquisition. The standard deviation of this distribution is then calculated to represent the variation in the atrial depolarization duration. Examples of other features can be found following in S3 File.

The OVG data represents the entirety of the electrical biopotential signals plotted on axes corresponding to the signal amplitude in millivolts of each channel, where ventricular depolarization and repolarization, and atrial depolarization, appear visually as loops. While this OVG data may appear similar across different people, it is unique for a given individual, therefore generating unique feature vectors in the high-dimensional feature space.

Step 3: Outlier detection

Mathematically outlying subjects were identified based on the signal’s feature values using the Isolation Forest algorithm [29]. Excluding outlying data ensures that the algorithm is not exposed to any data that is significantly differently than the development data.

Step 4: Machine learned model optimization

13 machine-learned models optimized to return high values for elevated LVEDP and low values for non-elevated LVEDP were evaluated given the features as input. Each model was trained individually by varying the subjects, features and thresholds. The models also varied, from Random Forest [30], Extreme Gradient Boosting [31] and Elastic Net [32]. See S10 File for a description of the algorithm, data, and hyperparameters for each model, as well as an explanatory figure for Extreme Gradient Boosting. Each of the 13 models were individually performant based on stratified 5-fold cross-validation repeated for 100 iterations (to vary the train/test folds) within the development data (S11 File), but represent unique analyses of LVEDP assessment. To capture the diversity of each model in a final single prediction, which is intended to eliminate the bias associated with the selection of a single model and thus reduce the likelihood over overfitting on the development data, the 13 model were amalgamated into a single predictive ensemble. The ensemble, composed of an average of the normalized outputs from the constituent models, is intended to on-average outperform any model that we may have selected from the pool of 13 when applied to new data [33].

Statistical analysis

Analyses were performed to determine the diagnostic performance of the machine learned predictor used as a continuous measurement (independent variable) on the prediction of LVEDP (dependent variable) adjudicated as elevated (≥25mmHg) or normal (≤12mmHg). A threshold was then established on the continuous measurement to yield a binary output, after which standard techniques were used to calculate sensitivity, specificity, negative and positive predictive values with each study subject categorized as either a true negative, true positive, false negative, and false positive. R 3.5.2 was used for statistical calculations, including 95% CIs, relevant statistical tests, ROC-AUCs, and propensity matching between the elevated LVEDP and control cohort. CIs were calculated using De Long’s method for AUC, and Clopper Pearson for sensitivity and specificity. Comparisons between ROC-AUCs were computed with DeLong’s test. Independent clinical predictors for an elevated LVEDP were calculated using multivariable logistic regression analyses. Permutation analysis was used to determine feature contribution within the machine learning model [30].

A simulation of BNP performance was performed using the known distributions of measured BNP values [2, 3], and specifically the estimated minimums, 25th percentiles, medians, 75th percentiles and maximums [2, 3] Specifically, the performance of BNP, as a commonly used point-of-care test for the prediction of heart failure, was explored using published studies that included patients with similar clinical characteristics to those in the present study (symptoms suggestive of decompensated HF [3], those with HFpEF, and those with obesity [2]) and to compare the post-test probability of the machine-learned predictor vs BNP. This simulation was then used to explore the sequential prediction of an elevated LVEDP [2] and non-elevated LVEDP [3] (i.e. dyspnea due to non-cardiac causes) using the BNP post-test (i.e., posterior) probability as the pre-test (i.e., prior) probability for the machine-learned predictor. Therefore, to propose how the present results may be used in clinical practice, the sequence of this analysis is the following: pre-test probability → BNP post-test probability used as the machine-learned posterior probability → machine-learned predictor post-test probability → prediction of elevated LVEDP; and to determine a net reclassification index (NRI) within this sequence [34].

The simulation of BNP performance was performed using the distributions of measured BNP values in two relevant publications cohorts [2, 3], specifically the estimated minimums, 25th percentiles, medians, 75th percentiles and maximums using boxplots. The approximate values of the constraining statistics for the two BNP datasets are as follows: 0ng/mL minimum for both datasets, 25th percentile of 75ng/mL for the non-cardiac etiology dataset and 250ng/mL for the obese HFpEF dataset, respective medians of 190ng/ML and 250ng/mL, 75th percentiles of 475ng/mL and 750ng/mL, and maximums of 1075ng/mL and 5000ng/mL.

These statistics were used as constraints to generate a simulated distribution of BNP for each of the cohorts, matching the number of subjects in the non-elevated and elevated LVEDP groups in the present work (258 subjects in the non-cardiac etiology group and 79 subjects in the obese HFpEF group). The performance of BNP was then assessed using this simulated data, from which AUC can be calculated, and thresholds of 50pg/mL and 150pg/mL were applied to calculate any statistics requiring a binary result (i.e., test-negative or test-positive to calculate sensitivity, specificity, PPV, NPV, likelihood ratios). The simulation was repeated for 1000 iterations with the values of the performance statistics averaged and confidence intervals calculated using the distribution of the statistics over the iterations (i.e., values at 2.5th and 97.5th percentiles). The simulation result was analyzed using a Bayesian methodology and to calculate the NRI for the ML predictor based on published methods [34].

Supporting information

S1 File. Multivariate clinical predictors of elevated LVEDP.

(DOCX)

S2 File. Feature permutation importance.

(DOCX)

S3 File. Features.

(DOCX)

S4 File. 2x2 contingency Tables & additional statistics.

(DOCX)

S5 File. Propensity matching.

(DOCX)

S6 File. STARD checklist.

(DOCX)

S7 File. Enrolling healthcare centers.

(DOCX)

S8 File. Inclusion and exclusion criteria.

(DOCX)

S9 File. Placement of the OVG electrodes and PPG sensor.

(DOCX)

S10 File. Model specifications.

(DOCX)

S11 File. Feature values.

(XLSX)

Acknowledgments

The authors are grateful to all patients who participated as study subjects; without you, this work would not have been possible.

Abbreviations

CAD

coronary artery disease

HF

heart failure

LVEDP

left ventricular end diastolic pressure

ML

machine learning

NRI

Net Reclassification Index

OVG

orthogonal voltage gradient

PPG

photoplethysmography

Data Availability

All relevant data are within the paper and its Supporting information files.

Funding Statement

The study sponsor (CorVista Health; Washington, DC & Toronto, Canada - corvista.com) was involved in the study design, data analysis, and interpretation of the results. The corresponding author, Sanjeev Bhavnani, had full access to all the data in the study, data analysis, and had final responsibility for the decision to submit for publication.

References

  • 1.Acharya UR, Krishnan SM, Spaan JAE, Suri JS. Advances in cardiac signal processing. Springer; 2007. [Google Scholar]
  • 2.Khalid U, Wruck LM, Quibrera PM, Bozkurt B, Nambi V, Virani SS, et al. BNP and obesity in acute decompensated heart failure with preserved vs. reduced ejection fraction: the Atherosclerosis Risk in Communities Surveillance Study. Int J Cardiol. 2017;233:61–6. doi: 10.1016/j.ijcard.2017.01.130 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 3.Maisel AS, Krishnaswamy P, Nowak RM, McCord J, Hollander JE, Duc P, et al. Rapid measurement of B-type natriuretic peptide in the emergency diagnosis of heart failure. N Engl J Med. 2002;347(3):161–7. doi: 10.1056/NEJMoa020233 [DOI] [PubMed] [Google Scholar]
  • 4.Sharifov OF, Schiros CG, Aban I, Denney TS Jr, Gupta H. Diagnostic accuracy of tissue doppler index E/è for evaluating left ventricular filling pressure and diastolic dysfunction/heart failure with preserved ejection fraction: A systematic review and meta-analysis. J Am Heart Assoc. 2016;5(1):e002530. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 5.Chiou Y-A, Hung C-L, Lin S-F. AI-Assisted Echocardiographic Prescreening of Heart Failure With Preserved Ejection Fraction on the Basis of Intrabeat Dynamics. JACC Cardiovasc Imaging. 2021. doi: 10.1016/j.jcmg.2021.05.005 [DOI] [PubMed] [Google Scholar]
  • 6.Pandey A, Kagiyama N, Yanamala N, Segar MW, Cho JS, Tokodi M, et al. Deep-Learning Models for the Echocardiographic Assessment of Diastolic Dysfunction. JACC Cardiovasc Imaging. 2021. doi: 10.1016/j.jcmg.2021.04.010 [DOI] [PubMed] [Google Scholar]
  • 7.Stuckey TD, Gammon RS, Goswami R, Depta JP, Steuter JA, Meine FJ, et al. Cardiac phase space tomography: A novel method of assessing coronary artery disease utilizing machine learning. Tang D, editor. PLoS One [Internet]. 2018. Aug 8 [cited 2020 Mar 30];13(8):e0198603. Available from: https://dx.plos.org/10.1371/journal.pone.0198603 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 8.Fathieh F, Paak M, Khosousi A, Burton T, Sanders WE, Doomra A, et al. Predicting cardiac disease from interactions of simultaneously-acquired hemodynamic and cardiac signals. Comput Methods Programs Biomed [Internet]. 2021. [cited 2021 Jun 24];202. Available from: https://reader.elsevier.com/reader/sd/pii/S0169260721000456?token=9E509B01CE407FA6AF5FA1AAE20A7C089C5A9C7B6D32BC4761AC9F82983A13DBC8E72EEBEAE2C3A3788B0360FF382AA1&originRegion=us-east-1&originCreation=20210624151325 doi: 10.1016/j.cmpb.2021.105970 [DOI] [PubMed] [Google Scholar]
  • 9.Thomas L, Marwick TH, Popescu BA, Donal E, Badano LP. Left atrial structure and function, and left ventricular diastolic dysfunction: JACC state-of-the-art review. J Am Coll Cardiol. 2019;73(15):1961–77. doi: 10.1016/j.jacc.2019.01.059 [DOI] [PubMed] [Google Scholar]
  • 10.Prasad A, Hastings JL, Shibata S, Popovic ZB, Arbab-Zadeh A, Bhella PS, et al. Characterization of static and dynamic left ventricular diastolic function in patients with heart failure with a preserved ejection fraction. Circ Hear Fail. 2010;3(5):617–26. doi: 10.1161/CIRCHEARTFAILURE.109.867044 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 11.Potter EL, Rodrigues CHM, Ascher DB, Abhayaratna WP, Sengupta PP, Marwick TH. Machine Learning of ECG Waveforms to Improve Selection for Testing for Asymptomatic Left Ventricular Dysfunction Prompt. JACC Cardiovasc Imaging. 2021. [DOI] [PubMed] [Google Scholar]
  • 12.Sengupta PP, Kulkarni H, Narula J. Prediction of Abnormal Myocardial Relaxation From Signal Processed Surface ECG. J Am Coll Cardiol. 2018;71(15):1650–60. doi: 10.1016/j.jacc.2018.02.024 [DOI] [PubMed] [Google Scholar]
  • 13.Attia ZI, Kapa S, Lopez-Jimenez F, McKie PM, Ladewig DJ, Satam G, et al. Screening for cardiac contractile dysfunction using an artificial intelligence–enabled electrocardiogram. Nat Med. 2019. Jan 1;25(1):70–4. doi: 10.1038/s41591-018-0240-2 [DOI] [PubMed] [Google Scholar]
  • 14.Sengupta PP, Shrestha S, Berthon B, Messas E, Donal E, Tison GH, et al. Proposed requirements for cardiovascular imaging-related machine learning evaluation (PRIME): a checklist: reviewed by the American College of Cardiology Healthcare Innovation Council. Cardiovasc Imaging. 2020;13(9):2017–35. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 15.Bhavnani SP, Parakh K, Atreja A, Druz R, Graham GN, Hayek SS, et al. 2017 Roadmap for innovation—ACC health policy statement on healthcare transformation in the era of digital health, big data, and precision health: a report of the American College of Cardiology Task Force on Health Policy Statements and Systems of Care. J Am Coll Cardiol. 2017;70(21):2696–718. doi: 10.1016/j.jacc.2017.10.018 [DOI] [PubMed] [Google Scholar]
  • 16.Dey D, Slomka PJ, Leeson P, Comaniciu D, Shrestha S, Sengupta PP, et al. Artificial intelligence in cardiovascular imaging: JACC state-of-the-art review. J Am Coll Cardiol. 2019;73(11):1317–35. doi: 10.1016/j.jacc.2018.12.054 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 17.Bachmann KN, Gupta DK, Xu M, Brittain E, Farber-Eger E, Arora P, et al. Unexpectedly low natriuretic peptide levels in patients with heart failure. Heart Fail. 2021;9(3):192–200. doi: 10.1016/j.jchf.2020.10.008 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 18.Jones R, Varian F, Alabed S, Morris P, Rothman A, Swift AJ, et al. Meta-analysis of echocardiographic quantification of left ventricular filling pressure. ESC Hear Fail. 2021;8(1):566–76. doi: 10.1002/ehf2.13119 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 19.Hummel YM, Liu LCY, Lam CSP, Fonseca-Munoz DF, Damman K, Rienstra M, et al. Echocardiographic estimation of left ventricular and pulmonary pressures in patients with heart failure and preserved ejection fraction: a study utilizing simultaneous echocardiography and invasive measurements. Eur J Heart Fail. 2017;19(12):1651–60. doi: 10.1002/ejhf.957 [DOI] [PubMed] [Google Scholar]
  • 20.Bianco CM, Farjo PD, Ghaffar YA, Sengupta PP. Myocardial mechanics in patients with normal LVEF and diastolic dysfunction. JACC Cardiovasc Imaging. 2020;13(1 Part 2):258–71. doi: 10.1016/j.jcmg.2018.12.035 [DOI] [PubMed] [Google Scholar]
  • 21.Alhakak AS, Møgelvang R, Schnohr P, Modin D, Brainin P, Gislason G, et al. The cardiac isovolumetric contraction time is an independent predictor of incident heart failure in the general population. Int J Cardiol. 2020;312:81–6. doi: 10.1016/j.ijcard.2020.03.046 [DOI] [PubMed] [Google Scholar]
  • 22.Martindale JL, Wakai A, Collins SP, Levy PD, Diercks D, Hiestand BC, et al. Diagnosing acute heart failure in the emergency department: a systematic review and meta-analysis. Acad Emerg Med. 2016;23(3):223–42. doi: 10.1111/acem.12878 [DOI] [PubMed] [Google Scholar]
  • 23.Bossuyt PM, Reitsma JB, Bruns DE, Gatsonis CA, Glasziou PP, Irwig L, et al. STARD 2015: an updated list of essential items for reporting diagnostic accuracy studies. Clin Chem. 2015;61(12):1446–52. doi: 10.1373/clinchem.2015.246280 [DOI] [PubMed] [Google Scholar]
  • 24.Khedraki R, Burton T, Cohoon T, Shen C, Khosousi A, Lange E, et al. First-in-man Development of a Machine Learning Cardiac Phase Space Analytic Approach to Predict Elevated Left Ventricular Pressures. J Am Coll Cardiol [Internet]. 2020. Mar 24;75(11_Supplement_1):3495. Available from: doi: 10.1016/S0735-1097(20)34122-X [DOI] [Google Scholar]
  • 25.Flessas AP, Connelly GP, Handa S, Tilney CR, Kloster CK, Rimmer RH Jr, et al. Effects of isometric exercise on the end-diastolic pressure, volumes, and function of the left ventricle in man. Circulation. 1976;53(5):839–47. doi: 10.1161/01.cir.53.5.839 [DOI] [PubMed] [Google Scholar]
  • 26.Bonow RO, Kent KM, Rosing DR, Lipson LC, Bacharach SL, Green MV, et al. Improved left ventricular diastolic filling in patients with coronary artery disease after percutaneous transluminal coronary angioplasty. Circulation. 1982;66(6):1159–67. doi: 10.1161/01.cir.66.6.1159 [DOI] [PubMed] [Google Scholar]
  • 27.Paul K, S. GL, J. BJ, Rory C, J. DB, William HE, et al. Recommendations for the Standardization and Interpretation of the Electrocardiogram. Circulation [Internet]. 2007. Mar 13;115(10):1306–24. Available from: doi: 10.1161/CIRCULATIONAHA.106.180200 [DOI] [PubMed] [Google Scholar]
  • 28.Yang H. Multiscale recurrence quantification analysis of spatial cardiac vectorcardiogram signals. IEEE Trans Biomed Eng. 2010;58(2):339–47. doi: 10.1109/TBME.2010.2063704 [DOI] [PubMed] [Google Scholar]
  • 29.Liu FT, Ting KM, Zhou Z-H. Isolation-based anomaly detection. ACM Trans Knowl Discov from Data. 2012;6(1):1–39. [Google Scholar]
  • 30.Breiman L. Random Forests. Mach Learn [Internet]. 2001. Oct;45(1):5–32. Available from: doi: 10.1023/A:1010933404324 [DOI] [Google Scholar]
  • 31.Chen T, Guestrin C. XGBoost: A scalable tree boosting system. In: Proceedings of the 22nd ACM SIGKDD International Conference on Knowledge Discovery and Data Mining [Internet]. New York, NY, USA: Association for Computing Machinery; 2016. p. 785–794. (KDD ‘16).
  • 32.Zou H, Hastie T. Regularization and variable selection via the elastic net. J R Stat Soc Ser B (Statistical Methodol) [Internet]. 2005. Apr 1;67(2):301–20. Available from: doi: 10.1111/j.1467-9868.2005.00503.x [DOI] [Google Scholar]
  • 33.Dietterich TG. Ensemble methods in machine learning. In: International workshop on multiple classifier systems. Springer; 2000. p. 1–15.
  • 34.Pencina MJ, D’Agostino RB Sr, Steyerberg EW. Extensions of net reclassification improvement calculations to measure usefulness of new biomarkers. Stat Med. 2011;30(1):11–21. doi: 10.1002/sim.4085 [DOI] [PMC free article] [PubMed] [Google Scholar]

Decision Letter 0

Chim C Lang

21 Jun 2022

PONE-D-22-07622Multicenter validation of a machine learning phase space electro-mechanical pulse wave analysis to predict elevated left ventricular end diastolic pressure at the point-of-carePLOS ONE

Dear Dr. Burton

Thank you for submitting your manuscript to PLOS ONE. After careful consideration, we feel that it has merit but does not fully meet PLOS ONE’s publication criteria as it currently stands. Therefore, we invite you to submit a revised version of the manuscript that addresses the points raised during the review process.

Kindly consider the comments made by the comments of the reviewers and respond accordingly.  Please ensure that your decision is justified on PLOS ONE’s publication criteria and not, for example, on novelty or perceived impact.

Please submit your revised manuscript by the Aug 05 2022 11:59PM. If you will need more time than this to complete your revisions, please reply to this message or contact the journal office at plosone@plos.org. When you're ready to submit your revision, log on to https://www.editorialmanager.com/pone/ and select the 'Submissions Needing Revision' folder to locate your manuscript file.

Please include the following items when submitting your revised manuscript:

  • A rebuttal letter that responds to each point raised by the academic editor and reviewer(s). You should upload this letter as a separate file labeled 'Response to Reviewers'.

  • A marked-up copy of your manuscript that highlights changes made to the original version. You should upload this as a separate file labeled 'Revised Manuscript with Track Changes'.

  • An unmarked version of your revised paper without tracked changes. You should upload this as a separate file labeled 'Manuscript'.

If you would like to make changes to your financial disclosure, please include your updated statement in your cover letter. Guidelines for resubmitting your figure files are available below the reviewer comments at the end of this letter.

If applicable, we recommend that you deposit your laboratory protocols in protocols.io to enhance the reproducibility of your results. Protocols.io assigns your protocol its own identifier (DOI) so that it can be cited independently in the future. For instructions see: https://journals.plos.org/plosone/s/submission-guidelines#loc-laboratory-protocols. Additionally, PLOS ONE offers an option for publishing peer-reviewed Lab Protocol articles, which describe protocols hosted on protocols.io. Read more information on sharing protocols at https://plos.org/protocols?utm_medium=editorial-email&utm_source=authorletters&utm_campaign=protocols.

We look forward to receiving your revised manuscript.

Kind regards,

Chim C. Lang

Academic Editor

PLOS ONE

Journal Requirements:

When submitting your revision, we need you to address these additional requirements.

1. Please ensure that your manuscript meets PLOS ONE's style requirements, including those for file naming. The PLOS ONE style templates can be found at

https://journals.plos.org/plosone/s/file?id=wjVg/PLOSOne_formatting_sample_main_body.pdf  and

https://journals.plos.org/plosone/s/file?id=ba62/PLOSOne_formatting_sample_title_authors_affiliations.pdf

2. Please include your tables as part of your main manuscript and remove the individual files. Please note that supplementary tables (should remain/ be uploaded) as separate "supporting information" files.

3. Thank you for stating the following in the Competing Interests section:

“I have read the journal's policy and authors of this manuscript have the following competing interests.

Sanjeev Bhavnani MD is a scientific advisor to Corvista Health and Blumio; consultant to Bristol Meyers Squibb, Pfizer, and Infineon Semiconductor; data safety monitoring board chair at Proteus Digital; has received research support from Scripps Clinic and the Qualcomm Foundation and is member of the healthcare innovation advisory boards at the American College of Cardiology, American Society of Echocardiography, and BIOCOM (all non-profit institutions with all positions voluntary).

Jeremiah P. Depta MD reports consulting fees from Edwards Lifesciences LLC, Boston Scientific, V wave Medical Ltd and Abbot.

Brett Bennett MD reports payment or honoraria for lecture from Philips.

Horace R. Gillins BS, Ian Shadforth EngD, Emmanuel Lange, Abhinav Doomra MScAC, Mohammad Firouzi MSc, Farhad Fathieh PhD, Timothy Burton BComp, Ali Khosousi PhD, Shyam Ramchandami PhD and William E. Sanders Jr. MD report employment by CorVista Health, and stock options in the same.

Frank Smart MD reports grants or contracts from Abbot (GUIDE HF clinical trial), NIH / Ohio State (DCM genetic study), Duke Clinical Research (Transform HF), CorVista Health (Pulmonary Hypertension clinical trial), and participation on a Data Safety Monitoring Board or Advisory Board (Abbott Medical; GUIDE-HF Steering committee).

All other authors report no disclosures.”

Please confirm that this does not alter your adherence to all PLOS ONE policies on sharing data and materials, by including the following statement: "This does not alter our adherence to  PLOS ONE policies on sharing data and materials.” (as detailed online in our guide for authors http://journals.plos.org/plosone/s/competing-interests).  If there are restrictions on sharing of data and/or materials, please state these. Please note that we cannot proceed with consideration of your article until this information has been declared.

Please include your updated Competing Interests statement in your cover letter; we will change the online submission form on your behalf.

4. We note that you have indicated that data from this study are available upon request. PLOS only allows data to be available upon request if there are legal or ethical restrictions on sharing data publicly. For more information on unacceptable data access restrictions, please see http://journals.plos.org/plosone/s/data-availability#loc-unacceptable-data-access-restrictions.

In your revised cover letter, please address the following prompts:

a) If there are ethical or legal restrictions on sharing a de-identified data set, please explain them in detail (e.g., data contain potentially sensitive information, data are owned by a third-party organization, etc.) and who has imposed them (e.g., an ethics committee). Please also provide contact information for a data access committee, ethics committee, or other institutional body to which data requests may be sent.

b) If there are no restrictions, please upload the minimal anonymized data set necessary to replicate your study findings as either Supporting Information files or to a stable, public repository and provide us with the relevant URLs, DOIs, or accession numbers. For a list of acceptable repositories, please see http://journals.plos.org/plosone/s/data-availability#loc-recommended-repositories.

We will update your Data Availability statement on your behalf to reflect the information you provide.

5. We note that Supplementary Appendix Section 9 includes an image of a patient in the study.

As per the PLOS ONE policy (http://journals.plos.org/plosone/s/submission-guidelines#loc-human-subjects-research) on papers that include identifying, or potentially identifying, information, the individual(s) or parent(s)/guardian(s) must be informed of the terms of the PLOS open-access (CC-BY) license and provide specific permission for publication of these details under the terms of this license. Please download the Consent Form for Publication in a PLOS Journal (http://journals.plos.org/plosone/s/file?id=8ce6/plos-consent-form-english.pdf). The signed consent form should not be submitted with the manuscript, but should be securely filed in the individual's case notes. Please amend the methods section and ethics statement of the manuscript to explicitly state that the patient/participant has provided consent for publication: “The individual in this manuscript has given written informed consent (as outlined in PLOS consent form) to publish these case details”.

If you are unable to obtain consent from the subject of the photograph, you will need to remove the figure and any other textual identifying information or case descriptions for this individual.

6. Please include captions for your Supporting Information files at the end of your manuscript, and update any in-text citations to match accordingly. Please see our Supporting Information guidelines for more information: http://journals.plos.org/plosone/s/supporting-information.

Additional Editor Comments:

This paper has been reviewed. A number of comments have been raised that require responses and revision.

[Note: HTML markup is below. Please do not edit.]

Reviewers' comments:

Reviewer's Responses to Questions

Comments to the Author

1. Is the manuscript technically sound, and do the data support the conclusions?

The manuscript must describe a technically sound piece of scientific research with data that supports the conclusions. Experiments must have been conducted rigorously, with appropriate controls, replication, and sample sizes. The conclusions must be drawn appropriately based on the data presented.

Reviewer #1: Yes

Reviewer #2: Partly

**********

2. Has the statistical analysis been performed appropriately and rigorously?

Reviewer #1: Yes

Reviewer #2: Yes

**********

3. Have the authors made all data underlying the findings in their manuscript fully available?

The PLOS Data policy requires authors to make all data underlying the findings described in their manuscript fully available without restriction, with rare exception (please refer to the Data Availability Statement in the manuscript PDF file). The data should be provided as part of the manuscript or its supporting information, or deposited to a public repository. For example, in addition to summary statistics, the data points behind means, medians and variance measures should be available. If there are restrictions on publicly sharing data—e.g. participant privacy or use of data from a third party—those must be specified.

Reviewer #1: Yes

Reviewer #2: No

**********

4. Is the manuscript presented in an intelligible fashion and written in standard English?

PLOS ONE does not copyedit accepted manuscripts, so the language in submitted articles must be clear, correct, and unambiguous. Any typographical or grammatical errors should be corrected at revision, so please note any specific errors here.

Reviewer #1: Yes

Reviewer #2: Yes

**********

5. Review Comments to the Author

Please use the space provided to explain your answers to the questions above. You may also include additional comments for the author, including concerns about dual publication, research ethics, or publication ethics. (Please upload your review as an attachment if it exceeds 20,000 characters)

Reviewer #1: In this manuscript, the authors aimed to verify if phase space ML analysis with orthogonal voltage gradient and photoplethysmography as features can be used to predict elevated left ventricular end diastolic pressure. Overall, the manuscript is well-structured and clearly written. Below I list some comments for this manuscript.

1. The authors can provide more clear description for the definition of the indices used as features (perhaps provide some figure illustrations).

2. Minor comment: since this is a multi-center study, the authors can provide descriptions about the chosen centers. For example, why choosing these centers? Any special subject characteristic for each center?

Reviewer #2: Introduction:

1. It would be nice to include some literature reviews on any previous studies predicting LVEDP using non-invasive measurement, if any.

Results:

1. Are all the results shown in the results section using the ensembled model?

Discussion:

1. The discussion should focus on the results presented in this study, and the literature reviews should be moved to the introduction.

2. In the limitation, the authors mentioned that study subjects taking diuretics might affect the specificity. Have the authors tried to remove those subjects and rerun the model to test that hypothesis?

3. The authors should add the study populations are patients likely to have CAD to the limitation.

4. The authors should also add the possibility of overfitting to the limitation unless they can justify the large number of features they used in the study.

Methods:

1. The study population are patients that are likely to have CAD, which seems to be a biased dataset to predict LVEDP. If the aim is being able to predict the LVEDP values for just patients who are likely to have CAD, then this would be acceptable, but if the aim is to predict LVEDP for a wider range of populations, then it would make sense to include other patients irrelevant to having CAD.

2. Line 336: Is it without CVD or CAD? If it is CVD, then which diseases are considered here? As CVD is a very wide category.

3. Is the validation group also patients likely to have CAD?

4. What type of catheter was used to measure LVEDP, specs?

5. Line 353: it says the following 4 categories, but there are 5 listed

6. Line 368: what are the symptoms are considered here for HF

7. From Line 414 – 420, please give some specific numbers for the cut-offs, e.g. high-frequency noise above XXX Hz?? Maximum measurable value XXX?

8. Could you give a bit more details on how the features are calculated? I imagine you have both OVG and PPG signals with multiple cardiac cycles. Do you calculate the features cycle by cycle? If so, how do you get the final subject level features based on the cycle features? Do you calculate the features over the whole recorded signal? If so, what is the time window you use for each recording? If the time window is constant, then how many seconds? If not, please justify.

9. Could you provide a list of features used in the appendix? And how they are calculated?

10. Is the outlier detection for the cycle level features or in-between subjects as well?

11. The rationale for choosing the 13 machine-learning models is unclear and how the hyper-parameters were chosen is also unclear. The authors claim that the ensemble outputs are intended to outperform any single model, but would they be able to provide some results showing that?

12. In terms of training the model, what type of training was used? Cross-validation?

13. Can the author provide some information about what they did to prevent overfitting the models?

14. Can the author provide some details on how the 95% CIs were calculated?

General question:

1. How do you avoid overfitting the models?

2. Have you investigated the correlations between features?

Figures:

Figure 8: could you add what each colour and lines represent?

Figure 9: This figure is not very clear. Personally, I don’t think it is necessary to have a figure to describe the random forest, as it is a very well known method. I think it is best to describe what setup was used in the Random Forest, such as bootstrapping=True, how many estimators, any limits for the number of trees, leaves etc. Same for the figure of XGBoost in the appendix.

**********

6. PLOS authors have the option to publish the peer review history of their article (what does this mean?). If published, this will include your full peer review and any attached files.

If you choose “no”, your identity will remain anonymous but your review may still be made public.

Do you want your identity to be public for this peer review? For information about this choice, including consent withdrawal, please see our Privacy Policy.

Reviewer #1: No

Reviewer #2: No

**********

[NOTE: If reviewer comments were submitted as an attachment file, they will be attached to this email and accessible via the submission site. Please log into your account, locate the manuscript record, and check for the action link "View Attachments". If this link does not appear, there are no attachment files.]

While revising your submission, please upload your figure files to the Preflight Analysis and Conversion Engine (PACE) digital diagnostic tool, https://pacev2.apexcovantage.com/. PACE helps ensure that figures meet PLOS requirements. To use PACE, you must first register as a user. Registration is free. Then, login and navigate to the UPLOAD tab, where you will find detailed instructions on how to use the tool. If you encounter any issues or have any questions when using PACE, please email PLOS at figures@plos.org. Please note that Supporting Information files do not need this step.

PLoS One. 2022 Nov 15;17(11):e0277300. doi: 10.1371/journal.pone.0277300.r002

Author response to Decision Letter 0


24 Aug 2022

We kindly request that you review the cover letter and response to reviewers documents (editor and reviewers, respectively), as they contain images and text formatting intended to facilitate your review, which are not possible to enter into this plain text field.

Attachment

Submitted filename: Response to Reviewers.pdf

Decision Letter 1

Chim C Lang

12 Sep 2022

PONE-D-22-07622R1Multicenter validation of a machine learning phase space electro-mechanical pulse wave analysis to predict elevated left ventricular end diastolic pressure at the point-of-carePLOS ONE

Dear Sir,

Thank you for submitting your manuscript to PLOS ONE. After careful consideration, we feel that it has merit but does not fully meet PLOS ONE’s publication criteria as it currently stands. Therefore, we invite you to submit a revised version of the manuscript that addresses the points raised during the review process.

==============================

Kindly note the comments of the reviewer to assist in the further revision needed for your manuscript, Please ensure that your decision is justified on PLOS ONE’s publication criteria and not, for example, on novelty or perceived impact.

For Lab, Study and Registered Report Protocols: These article types are not expected to include results but may include pilot data. 

==============================

Please submit your revised manuscript by Oct 27 2022 11:59PM. If you will need more time than this to complete your revisions, please reply to this message or contact the journal office at plosone@plos.org. When you're ready to submit your revision, log on to https://www.editorialmanager.com/pone/ and select the 'Submissions Needing Revision' folder to locate your manuscript file.

Please include the following items when submitting your revised manuscript:

  • A rebuttal letter that responds to each point raised by the academic editor and reviewer(s). You should upload this letter as a separate file labeled 'Response to Reviewers'.

  • A marked-up copy of your manuscript that highlights changes made to the original version. You should upload this as a separate file labeled 'Revised Manuscript with Track Changes'.

  • An unmarked version of your revised paper without tracked changes. You should upload this as a separate file labeled 'Manuscript'.

If you would like to make changes to your financial disclosure, please include your updated statement in your cover letter. Guidelines for resubmitting your figure files are available below the reviewer comments at the end of this letter.

If applicable, we recommend that you deposit your laboratory protocols in protocols.io to enhance the reproducibility of your results. Protocols.io assigns your protocol its own identifier (DOI) so that it can be cited independently in the future. For instructions see: https://journals.plos.org/plosone/s/submission-guidelines#loc-laboratory-protocols. Additionally, PLOS ONE offers an option for publishing peer-reviewed Lab Protocol articles, which describe protocols hosted on protocols.io. Read more information on sharing protocols at https://plos.org/protocols?utm_medium=editorial-email&utm_source=authorletters&utm_campaign=protocols.

We look forward to receiving your revised manuscript.

Kind regards,

Chim C. Lang

Academic Editor

PLOS ONE

Journal Requirements:

Please review your reference list to ensure that it is complete and correct. If you have cited papers that have been retracted, please include the rationale for doing so in the manuscript text, or remove these references and replace them with relevant current references. Any changes to the reference list should be mentioned in the rebuttal letter that accompanies your revised manuscript. If you need to cite a retracted article, indicate the article’s retracted status in the References list and also include a citation and full reference for the retraction notice.

Additional Editor Comments (if provided):

Kindly note the comments of the Reviewer to assist in your resubmission

[Note: HTML markup is below. Please do not edit.]

Reviewers' comments:

Reviewer's Responses to Questions

Comments to the Author

1. If the authors have adequately addressed your comments raised in a previous round of review and you feel that this manuscript is now acceptable for publication, you may indicate that here to bypass the “Comments to the Author” section, enter your conflict of interest statement in the “Confidential to Editor” section, and submit your "Accept" recommendation.

Reviewer #1: All comments have been addressed

Reviewer #2: (No Response)

**********

2. Is the manuscript technically sound, and do the data support the conclusions?

The manuscript must describe a technically sound piece of scientific research with data that supports the conclusions. Experiments must have been conducted rigorously, with appropriate controls, replication, and sample sizes. The conclusions must be drawn appropriately based on the data presented.

Reviewer #1: Yes

Reviewer #2: Partly

**********

3. Has the statistical analysis been performed appropriately and rigorously?

Reviewer #1: Yes

Reviewer #2: Yes

**********

4. Have the authors made all data underlying the findings in their manuscript fully available?

The PLOS Data policy requires authors to make all data underlying the findings described in their manuscript fully available without restriction, with rare exception (please refer to the Data Availability Statement in the manuscript PDF file). The data should be provided as part of the manuscript or its supporting information, or deposited to a public repository. For example, in addition to summary statistics, the data points behind means, medians and variance measures should be available. If there are restrictions on publicly sharing data—e.g. participant privacy or use of data from a third party—those must be specified.

Reviewer #1: Yes

Reviewer #2: No

**********

5. Is the manuscript presented in an intelligible fashion and written in standard English?

PLOS ONE does not copyedit accepted manuscripts, so the language in submitted articles must be clear, correct, and unambiguous. Any typographical or grammatical errors should be corrected at revision, so please note any specific errors here.

Reviewer #1: Yes

Reviewer #2: Yes

**********

6. Review Comments to the Author

Please use the space provided to explain your answers to the questions above. You may also include additional comments for the author, including concerns about dual publication, research ethics, or publication ethics. (Please upload your review as an attachment if it exceeds 20,000 characters)

Reviewer #1: (No Response)

Reviewer #2: 1. About overfitting

a. Generally, a good ratio between sample size and parameters to avoid overfitting is somewhere above 10 to 1. This means for an average of 149 features, you would need about 1490 subjects. If I am not mistaken, the maximum number included in this study is 684 post exclusions. It is true that the tree-based methods (Random Forest and XGB) and the elastic net can help reduce the risk of overfitting, but it does not mean you will not overfit because you are using these models. I have seen many overfitting cases when people use those models. I do see the authors trying to manage overfitting by limiting the maximum depth, but there are other parameters as well that need to be looked into.

b. One of my additional questions here is how did you choose the hyperparameters for your tree-based models? What type of tunning was performed?

c. Similarly, what type of tunning was conducted for the elastic net model? What makes you choose 0.003 or 0.01 for alpha?

d. Also, an AUC of 0.81 from the validation does not prove that overfitting did not occur since it is just one data point. One can easily argue if there is no overfitting, you could have an AUC of 0.95.

e. In summary, it is still not very clear how the hyperparameters were chosen for each model. And it is unfair for the authors to conclude that the overfitting did not occur based on an AUC of 0.81 from validation. Also, please remove the word “high-performing”, as a word like “high” or “low” is subjective and not scientific.

2. Regarding the type of catheter, the authors have replied to Comment #10 that the choice of the catheter was left to the discretion of each cardiologist, which means the device for measuring the ground truth can be different between time and the medical centre. Have the authors analysed the potential impact of this?

3. Is there a reason that the authors can only provide a list of the families of features used in the study instead of a list containing the actual features?

4. What type of cross-validation did the authors use? Can they specify it in the paper?

**********

7. PLOS authors have the option to publish the peer review history of their article (what does this mean?). If published, this will include your full peer review and any attached files.

If you choose “no”, your identity will remain anonymous but your review may still be made public.

Do you want your identity to be public for this peer review? For information about this choice, including consent withdrawal, please see our Privacy Policy.

Reviewer #1: No

Reviewer #2: No

**********

[NOTE: If reviewer comments were submitted as an attachment file, they will be attached to this email and accessible via the submission site. Please log into your account, locate the manuscript record, and check for the action link "View Attachments". If this link does not appear, there are no attachment files.]

While revising your submission, please upload your figure files to the Preflight Analysis and Conversion Engine (PACE) digital diagnostic tool, https://pacev2.apexcovantage.com/. PACE helps ensure that figures meet PLOS requirements. To use PACE, you must first register as a user. Registration is free. Then, login and navigate to the UPLOAD tab, where you will find detailed instructions on how to use the tool. If you encounter any issues or have any questions when using PACE, please email PLOS at figures@plos.org. Please note that Supporting Information files do not need this step.

PLoS One. 2022 Nov 15;17(11):e0277300. doi: 10.1371/journal.pone.0277300.r004

Author response to Decision Letter 1


30 Sep 2022

Dear Reviewer #2,

*Please find these responses in a separate submission file that has the tables formatted for better readability than when copied here*

Thank you for your thorough review and valuable comments. We appreciate and agree with your concern on overfitting and the corresponding risks. We view the recognition of this problem as one of the most significant in the field of machine learning. While we cannot definitively prove that overfitting has not occurred, we believe that the application of our various methods acts to sufficiently mitigate the of likelihood of its occurrence.

Specifically, our most critical control for overfitting was the single evaluation on a test dataset separated from the development set by enrollment date. Further, the probability of our model achieving the performance of 0.81 by chance on a dataset of that size (which could be the case if our development set was overfit) is low, as shown by the comparison between the curve and the random predictor that lies along the diagonal of the ROC space (p<2E-16 by DeLong’s test).

We also implemented a series of additional controls to increase the chance that our algorithm would generalize to an unseen dataset:

Use of modelling methodologies with regularization and penalty terms that are well-suited to high-dimensional spaces.

Cross-validation model development with careful monitoring of fold-to-fold standard deviations

Careful choice of models to ensemble based on hyperparameter stability

Below we provided detailed explanations of our methodologies to address your specific concerns, and we hope that we have addressed them to your satisfaction. Our responses appear below in bold, provided in italics where applicable in the response, and modified manuscript text in track changes.

Sincerely,

Sanjeev Bhavnani MD

Specific Comment #1: In order to clarify our imposed safeguards in reference to overfitting, we have revised the text on lines 350-355.

Author Response: The use of an ensemble, as is the case, does not increase the likelihood of overfitting but rather mitigates it by reducing the dependence on a single model. The methods employed to avoid overfitting include the use of cross-validation within the development data, using simple models with regularization and penalty terms, and testing the performance of the model on unseen data (only doing so only once). With respect to the models in particular; first, each model is exposed to an average of 149 features (with a range of 89-194, supplement section 10).

Generally, a good ratio between sample size and parameters to avoid overfitting is somewhere above 10 to 1. This means for an average of 149 features, you would need about 1490 subjects. If I am not mistaken, the maximum number included in this study is 684 post exclusions. It is true that the tree-based methods (Random Forest and XGB) and the elastic net can help reduce the risk of overfitting, but it does not mean you will not overfit because you are using these models. I have seen many overfitting cases when people use those models. I do see the authors trying to manage overfitting by limiting the maximum depth, but there are other parameters as well that need to be looked into.

Author Response: To clarify the number of subjects – the N=684 that you referenced is the validation cohort, while the models were developed on the N=1272 development cohort (and specifically, each model using a subset of the available cohort, as shown in supplement section 10). With respect to your other comments in this section, we hope that they will be addressed in the responses to your other questions.

One of my additional questions here is how did you choose the hyperparameters for your tree-based models? What type of tunning was performed?

Author Response: The possible hyperparameter values are shown below for XGB and Random Forest and were chosen to provide a range over the conservative values of each hyperparameter, with some exponential changes where appropriate to ensure coverage. Each combination of hyperparameters settings was trained within the development dataset only using cross-validation (recall that the ensemble was locked, and the validation data evaluated only once).

XGB Hyperparameters

Hyperparameter Possible Values

Learning Rate 0.1, 0.3, 0.5

Maximum Tree Depth 2, 3, 5, 7

Minimum Child Weight 1, 3, 5

Number of Trees 5, 10, 50, 100

Regularization Alpha 0.1, 0.3, 0.5

Random Forest Hyperparameters

Hyperparameter Possible Values

Maximum Tree Depth 2, 3, 5, 7

Minimum Samples Per Leaf 1, 25, 50, 100

Number of Trees 5, 10, 50, 100, 500

A heuristic stability assessment was performed during the selection of each model. Specifically, for a given model, the cross-validation performance of each “neighboring” hyperparameter specification was compared with that of the model under consideration, and the difference evaluated, with the goal of determining whether that region of the hyperparameter space generated similar cross-validation performances as a measure of stability. To illustrate the method, consider the toy example below using XGB, which assumes that XBG only requires the learning and minimum child weight hyperparameters. Varying these parameters each generates an AUC value, as described in the following table.

Minimum Child Weight

Learning Rate 1 3 5

0.1 AUC 0.1,1 AUC 0.1,3 AUC 0.1,5

0.3 AUC 0.3,1 AUC 0.3,3 AUC 0.3,5

0.5 AUC 0.5,1 AUC 0.5,3 AUC 0.5,5

The resultant AUC values can be used to assess the stability of learning rate = 0.3 and minimum child weight = 3 in the following manner. Should the stability value be close to zero, then that indicates that the performance at that specific learning rate and minimum child weight is relatively insensitive to changes to adjacent hyperparameter values.

Stability= 〖AUC〗_0.3,3- 1/8 (〖AUC〗_0.1,1+〖AUC〗_0.3,1+ 〖AUC〗_0.5,1+ 〖AUC〗_0.1,3+ 〖AUC〗_0.5,3 〖+ AUC〗_0.1,5+ 〖AUC〗_0.3,5+ 〖AUC〗_0.5,5 )

The extension of the methodology is straightforward to higher dimensions of hyperparameters, where the configuration of interest is compared to each of its neighbors in that high-dimensional space, which are found based on changing a single hyperparameter by one step in the assessed range of that hyperparameter.

Finally, please note that this is a heuristic that we used to guide our selection of the hyperparameters settings, and we do not claim that it is an optimal strategy. For instance, a limitation of this heuristic is the uneven spacing between hyperparameter settings – though we viewed this as necessary to explore the hyperparameter space, it does result in nonuniform distance between neighbors. A possible future extension of this method could correct for that nonuniform distance.

Similarly, what type of tunning was conducted for the elastic net model? What makes you choose 0.003 or 0.01 for alpha?

Author Response: Please see the explored hyperparameter space for Elastic Net in the table below. The same hyperparameter stability assessment performed for Random Forest and XBG (as described above in b.) was also performed for Elastic Net.

Elastic Net Hyperparameters

Hyperparameter Possible Values

Alpha 0, 0.003, 0.01, 0.1, 0.5

Fit Intercept True, False

L1 Ratio 0, 0.01, 0.1, 0.5

Normalize True, False

Also, an AUC of 0.81 from the validation does not prove that overfitting did not occur since it is just one data point. One can easily argue if there is no overfitting, you could have an AUC of 0.95.

Author Response: The point is well taken. Please see discuss on page one regarding AUC.

In summary, it is still not very clear how the hyperparameters were chosen for each model. And it is unfair for the authors to conclude that the overfitting did not occur based on an AUC of 0.81 from validation. Also, please remove the word “high-performing”, as a word like “high” or “low” is subjective and not scientific.

Author Response: We agree that the terms inferring “high/low” are subjective and the text has been altered accordingly on lines 368-369, as shown below. With respect to the hyperparameter selection, we hope that it is now clear from the response to question 1b.

Finally, the ultimate test of overfitting is the performance on unseen blinded data, which yielded an AUC of 0.81.

Specific Comment #2: Regarding the type of catheter, the authors have replied to Comment #10 that the choice of the catheter was left to the discretion of each cardiologist, which means the device for measuring the ground truth can be different between time and the medical centre. Have the authors analysed the potential impact of this?

Author Response: The reviewer makes an excellent point that variations in catheter technique could affect the consistency of measurements. Such sources of variability are an important concern. However, since 6-French to 8-French catheters are the primary options for left ventricular catheterization, recorded pressure should vary minimally as a function of catheter selection. Moreover, since a variety of catheter types and sizes are used in clinical practice, specifying a particular catheter type may impair the generalizability of the resultant algorithm.

Specific Comment #3: Is there a reason that the authors can only provide a list of the families of features used in the study instead of a list containing the actual features?

Author Response: Describing all of the features was out of scope for this paper, and therefore we provided feature description at the level of the family in the interest of brevity. However, we did provide detailed information on several of the features (Fig 3 and Supplement Section 3) to allow the reader to gain an appreciation for the prediction mechanism. In addition, many of the features are proprietary in nature.

Specific Comment #4: What type of cross-validation did the authors use? Can they specify it in the paper?

Author Response: The type of cross-validation used was stratified 5-fold cross-validation repeated for 100 iterations (to vary the train/test folds). Please see the modified text from lines 522-523 below:

Each of the 13 models were individually performant based on stratified 5-fold cross-validation repeated for 100 iterations (to vary the train/test folds) within the development data (S11) but represent unique analyses of LVEDP assessment.

Attachment

Submitted filename: Author Response To Reviewer_SEPT 24.docx

Decision Letter 2

Chim C Lang

25 Oct 2022

Multicenter validation of a machine learning phase space electro-mechanical pulse wave analysis to predict elevated left ventricular end diastolic pressure at the point-of-care

PONE-D-22-07622R2

Dear Dr

We’re pleased to inform you that your manuscript has been judged scientifically suitable for publication and will be formally accepted for publication once it meets all outstanding technical requirements.

Within one week, you’ll receive an e-mail detailing the required amendments. When these have been addressed, you’ll receive a formal acceptance letter and your manuscript will be scheduled for publication.

An invoice for payment will follow shortly after the formal acceptance. To ensure an efficient process, please log into Editorial Manager at http://www.editorialmanager.com/pone/, click the 'Update My Information' link at the top of the page, and double check that your user information is up-to-date. If you have any billing related questions, please contact our Author Billing department directly at authorbilling@plos.org.

If your institution or institutions have a press office, please notify them about your upcoming paper to help maximize its impact. If they’ll be preparing press materials, please inform our press team as soon as possible -- no later than 48 hours after receiving the formal acceptance. Your manuscript will remain under strict press embargo until 2 pm Eastern Time on the date of publication. For more information, please contact onepress@plos.org.

Kind regards,

Chim C. Lang

Academic Editor

PLOS ONE

Additional Editor Comments (optional):

Reviewers' comments:

Acceptance letter

Chim C Lang

5 Nov 2022

PONE-D-22-07622R2

Multicenter validation of a machine learning phase space electro-mechanical pulse wave analysis to predict elevated left ventricular end diastolic pressure at the point-of-care

Dear Dr. Bhavnani:

I'm pleased to inform you that your manuscript has been deemed suitable for publication in PLOS ONE. Congratulations! Your manuscript is now with our production department.

If your institution or institutions have a press office, please let them know about your upcoming paper now to help maximize its impact. If they'll be preparing press materials, please inform our press team within the next 48 hours. Your manuscript will remain under strict press embargo until 2 pm Eastern Time on the date of publication. For more information please contact onepress@plos.org.

If we can help with anything else, please email us at plosone@plos.org.

Thank you for submitting your work to PLOS ONE and supporting open access.

Kind regards,

PLOS ONE Editorial Office Staff

on behalf of

Dr. Chim C. Lang

Academic Editor

PLOS ONE

Associated Data

    This section collects any data citations, data availability statements, or supplementary materials included in this article.

    Supplementary Materials

    S1 File. Multivariate clinical predictors of elevated LVEDP.

    (DOCX)

    S2 File. Feature permutation importance.

    (DOCX)

    S3 File. Features.

    (DOCX)

    S4 File. 2x2 contingency Tables & additional statistics.

    (DOCX)

    S5 File. Propensity matching.

    (DOCX)

    S6 File. STARD checklist.

    (DOCX)

    S7 File. Enrolling healthcare centers.

    (DOCX)

    S8 File. Inclusion and exclusion criteria.

    (DOCX)

    S9 File. Placement of the OVG electrodes and PPG sensor.

    (DOCX)

    S10 File. Model specifications.

    (DOCX)

    S11 File. Feature values.

    (XLSX)

    Attachment

    Submitted filename: Response to Reviewers.pdf

    Attachment

    Submitted filename: Author Response To Reviewer_SEPT 24.docx

    Data Availability Statement

    All relevant data are within the paper and its Supporting information files.


    Articles from PLOS ONE are provided here courtesy of PLOS