Table 1.

Scan paremeters for 3 kinds of MR cisternography.

Sequence name Type Repetition time (ms) Echo time (ms) Flip angle (degree) Section thickness/re construction step (mm) Pixel size (mm) Number of slices Echo spacing (ms) Echo train length Echo train duration (ms) Phase partial fourier (%) Field of view (mm) Matrix size Number of excitations GRAPPA Bandwidth (Hz/pixel) Slice oversampling (%) Scan time (min:s)
MR cisternography 1 3D T2 SPACE 15,000 3,200 90/ constant160 2.0/1.0 0.5×0.5 80 7.23 599 4,345 77 1 65 × 196 324 × 384 1.4 3 434 100 6:00
MR cisternography 2 3D T2 SPACE 15,000 2,000 90/ constant160 2.0/1.0 0.5×0.5 80 7.23 599 4,157 96 165 × 196 324 × 384 1.4 3 434 100 6:00
MR cisternography 3 3D T2 SPACE 4,000 540 90/ constant130 2.0/1.0 0.5×0.5 80 6.11 272 1,369 82 165 × 196 324 × 384 1.4 3 434 100 3:28

SPACE, Sampling perfection with application-optimized contrasts using different flip angle evolutions.

GRAPPA, GeneRalized autocalibrating partial parallel acquisition.

All three sequences utilize a restore magnetization pulse and slab selective excitation pulse.

3D slab is set in an axial orientation.