Skip to main content
. 2022 Nov 17:1–12. Online ahead of print. doi: 10.1007/s11325-022-02740-w

Table 2.

Specificity and sensitivity values for AHI > 5 and AHI > 10

AHI > 10
PSG 1A vs Device
AHI > 10
PSG 1B vs Device
AHI > 5
PSG 1A vs Device
AHI > 5
PSG 1B vs Device
AHI > 10
1A vs 1B
AHI > 5
1A vs 1B
Sensitivity 0.556 0.667 0.903 0.905 0.353 0.625
Specificity 0.969 0.818 0.842 0.621 1.000 1.000
Positive predictive value 0.909 0.333 0.903 0.633 1.000 1.000
Negative predictive value 0.795 0.947 0.842 0.900 0.750 0.600