Table 3. Results of Method Evaluationa.

  Precision
Accuracy
Calibration Statistics
Analyte Repeat. Intra-Assay Inter-Assay % Recovery % Error LOD LOQ Linearity (R2)
E1 12.2 11.2 9.6 102.0 8.5 0.09±0.03 0.27±0.09 0.991 ± 0.005
            24 ± 8 72 ± 24  
E2 11.5 10.4 8.9 103.3 7.1 0.10±0.04 0.30±0.13 0.988 ± 0.009
            27 ± 12 82 ± 36  
BODIPY-E1 1.0 1.1 1.3 96.5 4.3 0.02±0.01 0.07±0.03 0.999 ± 0.001
            14 ± 7 42 ± 21  
BODIPY-E2 1.2 1.2 1.4 95.9 4.8 0.02±0.01 0.07±0.03 0.999 ± 0.001
            13 ± 6 40 ± 19  
a

Precision values are percent relative standard deviation. The limit of detection (LOD) and limit of quantitation (LOQ) are reported in micromolar (bold, top values) and μg/L (bottom values). The linearity and limit values are the means of the three different batches (n = 3) and include standard deviation (±value).