Abstract
PURPOSE
Computer-aided design and manufacturing (CAD-CAM) of implant abutments has been shown to result in surface contamination from site-specific milling and fabrication processes. If not removed, these contaminants can have a potentially adverse effect and may trigger inflammatory responses of the peri-implant tissues. The aim of the present study was to evaluate the bacterial disinfection and cleaning efficacy of ultrasonic reprocessing in approved disinfectants to reduce the microbial load of CAD-CAM abutments.
MATERIALS AND METHODS
Four different types of custom implant abutments (total N = 32) with eight specimens in each test group (type I to IV) were CAD-CAM manufactured. In two separate contamination experiments, specimens were contaminated with heparinized sheep blood alone and with heparinized sheep blood and the test bacterium Enterococcus faecium. Abutments in the test group were processed according to a three-stage ultrasonic protocol and assessed qualitatively and quantitatively by determination of residual protein. Ultrasonicated specimens contaminated with sheep blood and E. faecium were additionally eluted and the dilutions were incubated on agar plates for seven days. The determined bacterial counts were expressed as colony-forming units (CFU).
RESULTS
Ultrasonic reprocessing resulted in a substantial decrease in residual bacterial protein to less than 80 µg and a reduction in microbiota of more than 7 log levels of CFU for all abutment types, exceeding the effect required for disinfection.
CONCLUSION
A three-stage ultrasonic cleaning and disinfection protocol results in effective bacterial decontamination. The procedure is reproducible and complies with the standardized reprocessing and disinfection specifications for one- or two-piece CAD-CAM implant abutments.
Keywords: CAD-CAM abutments, Contamination, Disinfection, Bacterial decontamination, Ultrasonic cleaning
INTRODUCTION
Dental implant abutments are part of the prosthetic superstructure with direct contact with the surrounding oral tissues. In addition to prefabricated stock abutments, computer-aided design and manufacturing (CAD-CAM) enables the fabrication of customized abutments.1 They facilitate the compensation of axial divergences between the implant and the corresponding crown as well as the individualization of the abutment shoulder with an anatomical emergence profile for implant-supported single-tooth restorations.2,3 The peri-implant mucosa is commonly known as hypovascular and hypocellular scar tissue. Immunologically, it is inferior to the periodontal tissue around teeth, as it exhibits lower resistance to bacterial induced infections.4 Surface properties of the abutment such as topography, roughness, hydrophilicity, surface energy, contaminants, and macromolecular conditioning affect the biological response at the hard- and soft tissue interface in multiple ways.5 Hence, recurrent hazards with potential negative impact on the attachment of the peri-implant tissues should be prevented.6,7 CAD-CAM production for custom abutments have been proven to result in surface contamination from microwear particulates, cooling lubricants, and general laboratory debris.8,9 Analysis revealed contamination on their outer and inner surfaces following laboratory procedures.10,11 The presence of these micro-residues at the critical abutment-tissue junction may provoke inflammatory responses of the peri-implant tissues and mechanically compromise the stability of the implant-abutment junction.12 Cleaning and subsequent disinfection of the abutment surface is therefore mandatory. European health regulations and the guidelines of the American Dental Association (ADA) both approved cleaning and disinfection regimens for semi-critical medical devices, including CAD-CAM implant abutments.13,14 This refers to either dry heat sterilization, steam treatment of the components under pressure at 134℃ (autoclaving) or ultrasonic cleaning by means of approved disinfectants. While heat-stable metallic implant abutments (e.g. titanium or titanium nitride) can be safely autoclaved without compromising their material properties,15 sterilization of monotype ceramic or hybrid abutments under moist heat and pressure may lead to permanent damage to the crystal-ceramic framework (degradation) or to the adhesive bond of hybrid abutments and is therefore controversially discussed.16,17,18,19 In addition, it should be noted that the physical process of sterilization by autoclaving using a combination of appropriate heat and pressure is able to kill all viable forms of microbiota, but it cannot effectively remove particulate debris from CAD-CAM abutments. The reprocessing of implant abutments by steam cleaning in the laboratory (vaporization), although frequently employed, is an ineffective method and fails to achieve the normatively required efficacy of disinfection (DIN EN 14885:2018).20,21 In addition to cleanliness, the surface topography within the submucosal region of the abutment structure should be considered. Roughness, wettability and surface energy are important properties in this context. These parameters exhibit some correlation, as roughness can strongly influence wetting behavior.22 It can be assumed that there is a critical threshold for surface roughness, where the accumulation of bacteria and plaques is low, at the same time best supporting the attachment of fibroblasts and the adaptation of the peri-implant mucosa.23 One of the novel cleaning techniques is plasma pretreatment of implant abutments, which has been the subject of promising pilot studies.24,25 However, it should be considered that this is not a validated cleaning method for laboratory-fabricated implant prosthetic components.14,26 Due to the lack of legal validation and the fact that plasma devices are not widely used in dental laboratories or practices,21 the primary focus should be on cleaning methods with a realistic relationship between the technical efforts and costs of the devices and a reliable cleaning efficacy.
Alternatively, recent investigations have demonstrated that a validated three-step ultrasonic cleaning procedure significantly reduces surface contamination of monotype and hybrid CAD-CAM abutments without adversely affecting their tensile bond strength.9,19 The abutments are cleaned in three successive ultrasonic baths, one with an antibacterial cleaning solution, the other with 80% ethyl alcohol and finally with medically pure water (Cleaning System; Bredent GmbH & Co. KG, Senden, Germany) at 30℃ for 5 min each, yielding a reduction in deposit particles and organic and inorganic contaminants.10 While qualitative and semi-quantitative effectiveness in removing these contaminants from CAD-CAM-fabricated surfaces has been demonstrated by microscopic and chemical analyses,9,10 data on the disinfection efficacy of the three-step ultrasonic procedure are lacking. Therefore, the aim of the present study was to evaluate its potential for ultrasonic disinfection to reduce the microbial load of CAD-CAM implant abutments. The null hypothesis tested was that the three-step ultrasonic cleaning protocol results in a normatively required reduction of residual protein ≤ 80 µg per sample27 and a decrease of microbiota by at least 5 decadal logarithms (lg) of colony-forming units (CFU) relative to the positive control and is therefore suitable for disinfection of implant abutments.
MATERIALS AND METHODS
The laboratory procedures for sample production have been described in detail in a previous publication.10 Briefly summarized, a total of 32 CAD-CAM implant abutments were virtually constructed (Implant Studio; 3Shape, Copenhagen, Denmark) and fabricated (CADAbut F and CADAbut D; BEGO Implant Systems GmbH & Co. KG, Bremen, Germany). The master cast of a clinical case involving the replacement of the right maxillary central incisor with an implant (Semados SCX D 4.1/L 11.5; BEGO Implant Systems GmbH & Co. KG, Bremen, Germany) served as the basis for the digital abutment design. The geometry of the virtual design had uniform dimensions of 10.5 mm height and 6.5 mm shoulder width for all abutment specimens. According to AAMI TIR30:2011/R2016, 3 test samples per test setup are required. Due to the validation of the reprocessing efficacy of the process according to EN ISO 17664, it was decided to test representative abutments of 4 different types with a sample size of 3 per setup. The 32 CAD-CAM abutments were divided into four groups (n = 8 each) depending on the material used and CAD-CAM manufacturing process (Fig. 1): monotype abutments (one-piece) and hybrid abutments (two-piece). Lab-processed materials for monotype abutments included zirconia and titanium, while hybrid abutments were comprised of zirconia or lithium disilicate CAD-CAM copings that were bonded to prefabricated titanium bases (Ti-Base D 4.1 mm) (Semados SCX; BEGO Implant Systems GmbH & Co. KG, Bremen, Germany). Accordingly, the following four material groups were employed for testing: one-piece zirconia abutments (type I), two-piece lithium disilicate (LDS) meso-abutments bonded to titanium bases (type II), two-piece zirconia meso-abutments bonded to titanium bases (type III), and one-piece titanium abutments (type IV) (Fig. 2). A list of materials and manufacturers can be found in Table 1. The bonding surfaces of the titanium inserts and the ceramic copings of the two-piece hybrid abutments were grit-blasted (aluminum oxide particles 50 µm; 2 bar/0.25 MPa; 10 s; distance 10 mm) and cleaned with ethanol. Afterwards, the titanium inserts were moistened with a metal primer (GC Metal Primer II; GC EUROPE N.V, Leuven, Belgium), while a bonding agent (Monobond Plus; Ivoclar Vivadent, Schaan, Liechtenstein) was applied to the basal portion of the ceramic copings. Each hybrid abutment was luted with a dimethacrylate/hydroxyethyl methacrylate (DMA/HEMA)-based cement (Multilink Implant; Ivoclar Vivadent, Schaan, Liechtenstein) as recommended by the manufacturer. The excess cement was removed and the adhesive joint was polished with silicone polishers and polishing paste according to a previously documented protocol.28 Upon completion of a defined 3-step ultrasonic cleaning protocol, all test abutment types I-IV were evaluated in terms of both cleaning and disinfection effectiveness.
Fig. 1. Study design.
Fig. 2. Samples of the investigated CAD-CAM abutment types I-IV from left to right. Type I: zirconia monotype abutment (CADAbut F, BeCe CAD Zirconia XH), type II: lithium disilicate hybrid abutment (CADAbut D, IPS e.max CAD LT on Ti-Base), type III: zirconia hybrid abutment (CADAbut D, Zirconia LT CAD on Ti-Base), type IV: titanium monotype abutment (CADAbut F, CAD Titanium).
Table 1. Type, material, manufacturer, and sample size of tested CAD-CAM abutments.
| Abutment type | Abutment design | Material | Product & Manufacturer | No. of samples |
|---|---|---|---|---|
| Type I | Monotype Abutment | Y-TZP Zirconia | CADAbut F, BeCe CAD Zirconia XH, Semados SCX, BEGO Implant Systems | 8 |
| Type II | Hybrid Abutment | Lithium Disilicate Coping/Ti-Base | CADAbut D, IPS e.max CAD LT/Ivoclar Vivadent AG on Ti-Base, Semados SCX, BEGO Implant Systems | 8 |
| Type II | Hybrid Abutment | Y-TZP Zirconia Coping/Ti-Base | CADAbut D, Zirconia LT CAD on Ti-Base, Semados SCX, BEGO Implant Systems | 8 |
| Type IV | Monotype Abutment | Titanium Grade 5 | CADAbut F, CAD Titanium, Semados SCX, BEGO Implant Systems | 8 |
To test cleanability, in a first trial a test soil was prepared from heparinized sheep blood (Fiebig Xebios Diagnostics GmbH, Düsseldorf, Germany, lot number 31406500/01) diluted 1:5 with 0.85% NaCl and reactivated with protamine and 1% mucin according to DIN EN ISO 15883-529 and AAMI TIR 30:2011.30 For contamination, four test specimens of each abutment type (type I-IV, total n = 16) were immersed in the test soil and subsequently dried for one hour at 22 ± 2℃ under laminar room air.
After soiling, three test specimens per abutment type were subjected to a three-stage ultrasonic cleaning protocol using ultra-high frequency waves in combination with disinfecting agents (Finevo Cleaning System; Bredent GmbH & Co. KG, Senden, Germany) (Fig. 3). According to this protocol, the samples were cleansed three times in an ultrasonic device (Finevoclean; Sirius Ceramics, Frankfurt, Germany) in separate glass beakers at 30℃ for 5 min each. The first bath contained an antibacterial cleaning solution (FINEVO 01, serial number 841117230. BEGO Implant Systems GmbH & Co. KG, Bremen, Germany), the second bath contained 80% ethyl alcohol, and the third bath contained medically purified water (Aqua Destilata, Sanismart GmbH, Waltrop, Germany). The total cleaning time amounted to 15 min. Declared by the manufacturer, the cleaning device solution contains 1.5 g of chlorhexidine gluconate and 15 g of cetrimide per 100 g as active ingredients. The fourth test specimen of each abutment type remained uncleaned as a positive control.
Fig. 3. Set-up of type IV titanium monotype abutment for visual inspection after ultrasonic reprocessing.
The cleanliness of this first group of test specimens was qualitatively assessed after ultrasonic reprocessing by visual inspection with a magnifying glass at 400 × magnification (Fig. 3). Visible cleanliness served as the acceptance criterion in accordance with the requirements for reprocessed instruments (Guideline German Society of Hospital Hygiene).31 Specimens visually free of contamination were subsequently subjected to quantitative analysis for protein residues using a modified o-phthalaldehyde (OPA) spectrophotometric method.
Prior to protein analysis, the reprocessed test abutments were subjected to elution according to DIN EN ISO 15883-5:2021-11.29 For this purpose, they were eluted in 5 ml elution solution (1% sodium dodecyl sulfate, SDS, pH 11) in a 15 ml tube on a rotary shaker for 10 min at 300 rpm. Glass beads were added for better retrieval of the test soil. Screening of protein in the SDS eluate was based on the modified OPA method. The principle of the OPA assay is the chemical conversion of o-phthaldialdehyde and free amino groups in the presence of a thiol component to form fluorescent isoindole compounds (absorbance 340 nm; emission 450 nm). Since the thiol reagent mercaptoethanol is not suitable for general use,32 it is replaced by N,N-dimethyl-2-mercaptoethylammonium chloride in the modified OPA method. Advantageous is a more stable extinction behavior compared to mercaptoethanol. The quantity of residual protein was evaluated according to the requirements for reprocessed instruments (Guideline German Society of Hospital Hygiene,27 DIN EN ISO 15883-5,29 and AAMI TIR 30:201130). They define a residual protein quantity of more than 150 µg per sample as threshold value, while a concentration of > 80 to ≤ 150 µg is classified as critical value, and ≤ 80 µg per sample as reference value. The cleaning guidelines based on international standards suggest a residual protein content of < 6.4 µg protein/cm2 of the product. The extent of depletion was calculated in percentage (%) based on the original degree of contamination (positive controls).
An additional four test specimens of each abutment type (Type I-IV, total n = 16) were soiled in a second soiling test according to DIN EN ISO 15883-529 to test the disinfection efficacy. This particular test soiling consisted of heparinized sheep blood (Fiebig Xebios Diagnostics GmbH, Düsseldorf, Germany, charge 31406500/01)) diluted 1:5 with 0.85% NaCl and reactivated with protamine and 1% mucin as well as with the test germ E. faecium. The second passage of the test bacterium cultured on brain-heart infusion agar (BHI) was adjusted to 1.5-5.0 × 109 CFU/ml (colony forming units) in a dilution solution of 0.85% NaCl and 0.1% tryptone. After centrifugation, the dilution solution was decanted and the volume was supplemented with heparinized sheep blood. Thereafter, the suspension was carefully homogenized with glass beads. For contamination, the test specimens were immersed in the test soiling and then dried for 1 h at 22 ± 2℃ in a laminar air flow box. Three test abutments per group were ultrasonically cleaned and disinfected according to the procedure described above. One abutment per product type remained unprocessed and served as a positive control.
The ultrasonically processed test abutments were eluted into tryptic soy broth (TSB). It consists of 3.0% polysorbate 80, 3.0% saponine, 0.1% histidine, and 0.1% cysteine. The elution volume was 10 ml. The test samples were eluted in 10 ml of elution solution in a 15 ml tube on a rotary shaker for 10 min at 300 rpm. Glass beads were added for better recovery of the test soils.
Dilutions were prepared from the positive controls using a dilution solution of 0.1% tryptone and 0.85% sodium chloride and spread to tryptic soy agar (TSA). From each of the processed samples, 1.0 and 0.1 ml of the eluate were plated onto TSA. All samples were incubated at 36 ± 1℃ for seven days. The grown colonies were counted visually and the determined germ counts were expressed as the decadal logarithm (lg) of the colony-forming units (CFU). The remaining elution volume was incubated for 7 days at 37℃ to allow growth of pre-damaged test bacteria that were not completely killed (enrichment). Over the course of the 7 days, this was visible as turbidity of the sample and assessed with a detected colony count of < 10 (resulting in a logarithmic value of < 1), in case no residual germs could be detected on the culture medium. To evaluate the disinfection efficacy of the three-step cleaning protocol, the microbial load of the test abutments was subtracted from the microbial load of the positive controls, i.e., the reduction factor (RF) was calculated according to the following formula: RF = lg CFU/positive control - lg CFU/test specimens. The criterion for acceptance was a reduction of the test organism (bacteria) by ≥ 5 lg steps.
RESULTS
All test abutments (type I-IV) passed the qualitative-visual inspection with a magnifying glass at 400 × magnification for cleanliness after being subjected to the three-stage cleaning procedure. The prerequisite for determining the amount of residual protein was thus met for all test specimens.
For the uncleaned one-piece zirconia abutment (positive control type I), a protein concentration of 670.55 µg bovine serum albumine (BSA) was detected in the 1:1.25 diluted eluate by modified OPA analysis, whereas after ultrasonic cleaning, protein was not detectable in the undiluted eluate for any of the three test type I specimens examined, and the depletion was ≥99.99% (Table 2). The one-piece titanium test abutments (type IV) exhibited a similar high cleaning efficacy based on the initial BSA of 1492.15 µg for the positive control (Table 2). In both two-piece hybrid abutments (type II, type III), minimal traces of BSA were present in the serum after cleaning. The protein concentration was a maximum of 1.83 µg BSA for lithium disilicate hybrid abutments (type II) (Table 2) and 67.74 µg BSA for zirconia hybrid abutments (type III) (Table 2). However, the extent of depletion was over 95% for type III abutments and over 99% for type II abutments. As a result, the test requirements for cleanliness for all investigated one-piece- and two-piece abutments were successfully met (reference value: ≤ 80 µg residual protein quantity per sample).31 The limit of detection was 3.9 µg/ml, resulting in a limit of quantification of 19.5 µg/sample.
Table 2. Results of quantitative protein analysis based on modified o-phthalaldehyde spectrophotometric method (OPA) for: type I abutments (one-piece zirconia abutments); type II abutments (two-piece lithium disilicate abutments); type III abutments (two-piece zirconia abutments); type IV abutments (one-piece titanium abutments).
| Dilution factor | Extinction at 340 nm (OPA) | Protein concentration [µg BSA/TS] | Extend of depletion [%] | Criterion [≤ 80 µg residual protein TS] | ||
|---|---|---|---|---|---|---|
| Type I sample | ||||||
| PC-I | 1:1.25 | 0.293 | 670.55 | - | - | |
| Test I-1 | 1.0 | 0.000 | b.d. | ≥ 99.99 | Pass | |
| Test I-2 | 1.0 | -0.001 | b.d. | ≥ 99.99 | Pass | |
| Test I-3 | 1.0 | 0.000 | b.d. | ≥ 99.99 | Pass | |
| Type II sample | ||||||
| PC-II | 1:1.25 | 0.344 | 787.27 | - | - | |
| Test II-1 | 1.0 | 0.001 | 1.83 | ≥ 99.73 | Pass | |
| Test II-2 | 1.0 | 0.001 | 1.83 | ≥ 99.73 | Pass | |
| Test II-3 | 1.0 | -0.001 | b.d. | ≥ 99.99 | Pass | |
| Type III sample | ||||||
| PC-III | 1:1.25 | 0.652 | 1492.15 | - | - | |
| Test III-1 | 1.0 | 0.037 | 67.74 | 95.46 | Pass | |
| Test III-2 | 1.0 | -0.001 | b.d. | ≥ 99.99 | Pass | |
| Test III-3 | 1.0 | 0.034 | 62.25 | 95.83 | Pass | |
| Type IV sample | ||||||
| PC-IV | 1:1.25 | 0.625 | 1492.15 | - | - | |
| Test IV-1 | 1.0 | -0.002 | b.d. | ≥ 99.99 | Pass | |
| Test IV-2 | 1.0 | -0.003 | b.d. | ≥ 99.99 | Pass | |
| Test IV-3 | 1.0 | 0.000 | b.d. | ≥ 99.99 | Pass | |
Three test specimens (Test X-1,2,3) were subjected to ultrasonic cleaning once. The fourth test specimen remained uncleaned as positive control. BSA = Bovine Serum Albumin, b.d. = below detection, PC = positive control, TS = test sample.
DISCUSSION
Recent in vitro studies have detected contaminants on the surfaces of titanium and zirconia CAD-CAM customized implant abutments from various manufacturers.8,9,10 All tested specimens exhibited process-related roughening, micro wear deposits, and organic and inorganic debris upon delivery. These contaminants can be caused by machining residues, coolant or chemical washing protocols after industrial milling as a remnant of surface processing in centralized production. The fabrication of CAD-CAM abutments in the dental laboratory is basically subject to the same sources of contamination. In addition, the risk of remaining blasting media, excess adhesive, hand grease, polishing agents and rubber residues must be taken into account.9,33 Soft tissues surrounding implants are hypovascular and hypocellular scar tissues with a considerably lower immunologic capacity than periodontal tissues around teeth.12,34 If not removed, these particulate contaminants can have a potentially adverse effect and may trigger inflammatory responses of the peri-implant tissues.24,28,35 Titanium particles have been shown to trigger acute inflammation via increased interleukin (IL)-1 β secretion and IL-1-associated signaling by promoting the NALP3 inflammasome, which is a temporal enzyme of PGE2 synthesis, and RANL/RANKL that differentiates osteoclasts.36,37 In an effort of surface cleaning, the impact of different cleaning, disinfection and sterilization methods for implant abutments is therefore comprehensively discussed.25,26,38,39 However, a solid clinical relationship between abutment cleanliness and peri-implant bone preservation has not yet been demonstrated.40 It is worth noting that both European and North American health regulations have approved only sterilization or high-level ultrasound disinfection procedures for cleaning semi-critical medical devices such as CAD-CAM implant abutments.13,14 Despite the existence of these regulatory guidelines for adequate cleaning or sterilization, there does not appear to be a consistent application.21 Sterilization by autoclaving involves a physical process that uses a combination of appropriate heating and pressure to remove or destroy all viable forms of microorganisms, including bacterial spores. Some concerns have been raised about the use of resin cements with regard to their hydrothermal aging resistance during autoclavation, as a detrimental effect on bond strength (degradation) has been reported.41 A recent systematic review on the bond strength of resin based cements, however, showed that thermo-artificial aging has minimal effect if the adhesive surface was pretreated, such as sandblasted and/or coated with acid monomers.42 More recent in vitro results on the structural integrity and bond strength between zirconia frameworks and titanium-based abutments after autoclave sterilization have also shown no negative influence on bond strength.18,43
While sterilization has the ability to effectively eliminate microbial surface contamination to achieve aseptic and sterile settings, particulate debris on CAD-CAM abutments cannot be removed by this process alone. However, ultrasonic cleaning with high-frequency waves in approved disinfectants is validated and effective in the mechanical removal of foreign bodies and microbiota from the surface of abutments.10,39 In vitro results employing a three-stage ultrasonic cleaning procedure confirm generally good cell viability on titanium specimens and improved cell attachment and reduced inflammatory response of human gingival fibroblasts (HGF) on zirconia surfaces.28,44 Ultrasonic devices are widely available and employed in clinics and dental laboratories and can be conveniently used for abutment hygiene.45,46
While the effectiveness of the three-step ultrasonic cleaning protocol for removing particulate contamination from CAD-CAM-fabricated surfaces has already been conclusively demonstrated,9,10 the present study aimed to evaluate the microbiological disinfection efficacy of this regimen for CAD-CAM implant abutments. The present data show that an antibacterial cleaning solution with chlorhexidine gluconate and cetrimide as active ingredients, followed by ethyl alcohol and medically pure water in sonication treatment, resulted in effective bacterial decontamination. It revealed a decrease in residual protein of less than 80 µg and a reduction in microbiota of more than 7 lg levels of colony-forming units (CFU) for all abutment types, exceeding the effect required for cleaning and disinfection efficacy (Table 3). OPA assays failed to detect residual protein quantity for the ultrasonically processed zirconia (type I) and titanium (type IV) monotype test abutments. Only minimal traces of protein residues were measured in the eluates of the two-piece hybrid abutments (type II, type III) after ultrasonic cleaning and disinfection, amounting to 1.83 µg BSA equivalent for lithium disilicate hybrid abutments (type II) and 67.74 µg BSA for zirconia hybrid abutments (type III) (Table 2). The degree of disinfection, as measured by the residual protein concentration, was thus dependent on the manufacturing process of the respective CAD-CAM abutment. Nevertheless, all specimens successfully met the standard reference value of ≤ 80 µg residual protein per sample.31 Since the titanium and ceramic abutments tested had non-porous surfaces, it can be hypothesized that the materials employed did not have a significant influence on the disinfection effect. It can only be assumed that macro- and micro-design of the abutments such as geometry, cavities, surface roughness, micro texturing and polishing had a potential influence on the efficacy of high-frequency microwaves and the disinfection regime. A possible explanation for the marginally inferior cleaning results of the tested hybrid abutments after sonication processing could be the fact that their manufacturing process is predominantly manual and thus minimal residues of polishing agents and hand grease may not be completely removed by ultrasound and cleaning and disinfection. The residual levels of protein found on hybrid abutments in the current study are consistent with the results of previous investigations. An in vitro study using scanning electron microscopy (SEM), X-ray dispersion analysis (EDX) and computer-assisted planimetric analysis (CAPM) to evaluate the quality and quantity of process-related surface contamination of various customized implant abutments also concluded that hybrid abutments, in contrast to monotype abutments, exhibited minimal amounts of debris after ultrasonic cleaning.10 Whether the adhesive gap and bonding material of hybrid abutments have an influence on the removal of protein contamination remains to be investigated. The specific influence of the abutment material and the post-processing measures on the cleanliness of the abutment before and after ultrasonic cleaning has yet to be determined in this contest. For the detection of residual protein, the modified OPA assay was chosen because it is considered a suitable method for the quantitative determination of free and terminal amino groups (sensitivity: 0.03 - 1 µg/ml).47 In practice, this method was proven to be easy to adopt, relatively simple to use, and readily implementable under the given conditions. The normative acceptance criterion for confirming disinfection efficacy with a reduction of test germs by at least 5 lg levels compared to the positive control was clearly exceeded in the present study. Consequently, the null hypothesis can be considered accepted, as the ultrasonic protocol successfully met the test requirements for disinfection of all investigated one- and two-piece implant abutments.
Table 3. Results of disinfection test for: type I abutments (one-piece zirconia abutments); type II abutments (two-piece lithium disilicate abutments); type III abutments (two-piece zirconia abutments); type IV abutments (one-piece titanium abutments).
| Holding time | CFU/TS | Enrichment* | log/TS | RF | Criterion | ||
|---|---|---|---|---|---|---|---|
| Type I sample | |||||||
| PC-I | - | 1.45 × 108 | - | 8.19 | - | - | |
| Test I-1 | < 10 | + | < 1.00 | > 7.19 | Pass | ||
| Test I-2 | 5 min. | < 10 | + | < 1.00 | > 7.19 | Pass | |
| Test I-3 | 0 | - | 0.00 | ≥ 8.19 | Pass | ||
| Type II sample | |||||||
| PC-II | - | 4.4 × 108 | - | 8.64 | - | - | |
| Test II-1 | 5 min. | 0 | - | 0.00 | ≥ 8.64 | Pass | |
| Test II-2 | 0 | - | 0.00 | ≥ 8.64 | Pass | ||
| Test II-3 | 0 | - | 0.00 | ≥ 8.64 | Pass | ||
| Type III sample | |||||||
| PC-III | - | 1.47 × 108 | - | 8.17 | - | - | |
| Test III-1 | 5 min. | 0 | - | 0.00 | ≥ 8.17 | Pass | |
| Test III-2 | 0 | - | 0.00 | ≥ 8.17 | Pass | ||
| Test III-3 | 0 | - | 0.00 | ≥ 8.17 | Pass | ||
| Type IV sample | |||||||
| PC-IV | - | 1.33 × 108 | - | 8.12 | - | - | |
| Test IV-1 | 5 min. | 0 | - | 0.00 | ≥ 8.12 | Pass | |
| Test IV-2 | 0 | - | 0.00 | ≥ 8.12 | Pass | ||
| Test IV-3 | 0 | - | 0.00 | ≥ 8.12 | Pass | ||
Three test specimens (Test X-1,2,3) were subjected to ultrasonic cleaning once. The fourth test specimen remained uncleaned as positive control. CFU = colony forming units, log = logarithm [of colony forming units], TS = test sample, RF = reduction factor = (initial contamination [lg] - residual contamination [lg], PC = positive control, + = Turbidity due to bacteria growth in sub cultivation (enrichment), - = lack of turbidity due to lack of bacterial growth, * = Incubation of the eluate 7d at 36 ± 1℃.
The test soiling of blood and Enterococcus faecium proved to have good reproducibility in the evaluation of medical devices.48 In the present study, the addition of bacterial cultures and blood as test soil allowed the validation of the disinfection efficacy on the tested CAD-CAM implant abutments. Disinfection alone is not able to achieve the necessary reduction of microorganisms, and therefore efficient cleaning, e.g. by ultrasound, is also advised. Since customized abutments were sequentially immersed in three different solutions with ultrasound, the physical effects of ultrasonic cleaning and the effect of antimicrobial solutions were not separately evaluated in the limitation of this study. Thus, further studies will be necessary to assess each effect of antimicrobial solutions and ultrasonic procedure separately. Although the reproducibility of the in vitro investigation was ensured by the use of a defined contamination and ultrasonic cleaning protocol, the employment of an artificial soil could be considered a drawback of the present study. Another limitation concerns the relatively small number of samples per subgroup and their limited variety of material and size, which should be expanded in future studies with a larger number and selection of samples.
CONCLUSION
Based on the present findings, the following conclusions were obtained:
A three-step ultrasonic cleaning and disinfection process is reproducible and complies with the standardized reprocessing and disinfection specifications for custom one- or two-piece CAD-CAM implant abutments made of titanium or ceramic. The cleaning solution with the active ingredients chlorhexidine gluconate and cetrimide, followed by ethyl alcohol and medically pure water in the ultrasonic treatment, leads to effective bacterial decontamination. Ultrasonic reprocessing resulted in a substantial decrease of residual protein of less than 80 µg and a reduction in microbiota of more than 7 lg levels of colony-forming units for all abutment types, exceeding the effect required for disinfection.
ACKNOWLEDGEMENTS
The authors gratefully acknowledge Jochen Schlenker for his support and valuable advise as well as Patrick Mühlbeyer for his excellent technical assistance.
The authors thank BEGO Medical GmbH for providing the test samples for the experimental investigation.
References
- 1.Kapos T, Evans C. CAD/CAM technology for implant abutments, crowns, and superstructures. Int J Oral Maxillofac Implants. 2014;29:117–136. doi: 10.11607/jomi.2014suppl.g2.3. [DOI] [PubMed] [Google Scholar]
- 2.Zarauz C, Pitta J, Pjetursson B, Zwahlen M, Pradies G, Sailer I. Esthetic outcomes of implant-supported single crowns related to abutment type and material: a systematic review. Int J Prosthodont. 2021;34:229–249. doi: 10.11607/ijp.6314. [DOI] [PubMed] [Google Scholar]
- 3.Gehrke P, Bleuel K, Fischer C, Sader R. Influence of margin location and luting material on the amount of undetected cement excess on CAD/CAM implant abutments and cement-retained zirconia crowns: an in-vitro study. BMC Oral Health. 2019;19:111. doi: 10.1186/s12903-019-0809-2. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 4.Ivanovski S, Lee R. Comparison of peri-implant and periodontal marginal soft tissues in health and disease. Periodontol 2000. 2018;76:116–130. doi: 10.1111/prd.12150. [DOI] [PubMed] [Google Scholar]
- 5.Rupp F, Gittens RA, Scheideler L, Marmur A, Boyan BD, Schwartz Z, Geis-Gerstorfer J. A review on the wettability of dental implant surfaces I: theoretical and experimental aspects. Acta Biomater. 2014;10:2894–2906. doi: 10.1016/j.actbio.2014.02.040. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 6.Canullo L, Tallarico M, Radovanovic S, Delibasic B, Covani U, Rakic M. Distinguishing predictive profiles for patient-based risk assessment and diagnostics of plaque induced, surgically and prosthetically triggered peri-implantitis. Clin Oral Implants Res. 2016;27:1243–1250. doi: 10.1111/clr.12738. [DOI] [PubMed] [Google Scholar]
- 7.Welander M, Abrahamsson I, Berglundh T. The mucosal barrier at implant abutments of different materials. Clin Oral Implants Res. 2008;19:635–641. doi: 10.1111/j.1600-0501.2008.01543.x. [DOI] [PubMed] [Google Scholar]
- 8.Canullo L, Micarelli C, Lembo-Fazio L, Iannello G, Clementini M. Microscopical and microbiologic characterization of customized titanium abutments after different cleaning procedures. Clin Oral Implants Res. 2014;25:328–336. doi: 10.1111/clr.12089. [DOI] [PubMed] [Google Scholar]
- 9.Gehrke P, Tabellion A, Fischer C. Microscopical and chemical surface characterization of CAD/CAM zircona abutments after different cleaning procedures. A qualitative analysis. J Adv Prosthodont. 2015;7:151–159. doi: 10.4047/jap.2015.7.2.151. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 10.Gehrke P, Abazari C, Schlichter K, Fischer C, Duddeck D, Romanos GE, Weigl P. Qualitative and semi-quantitative assessment of processing-related surface contamination of one- and two-piece CAD/CAM abutments before and after ultrasonic cleaning. Materials (Basel) 2020;13:3225. doi: 10.3390/ma13143225. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 11.Canullo L, Micarelli C, Iannello G. Microscopical and chemical surface characterization of the gingival portion and connection of an internal hexagon abutment before and after different technical stages of preparation. Clin Oral Implants Res. 2013;24:606–611. doi: 10.1111/j.1600-0501.2012.02499.x. [DOI] [PubMed] [Google Scholar]
- 12.Mishra PK, Wu W, Rozo C, Hallab NJ, Benevenia J, Gause WC. Micrometer-sized titanium particles can induce potent Th2-type responses through TLR4-independent pathways. J Immunol. 2011;187:6491–6498. doi: 10.4049/jimmunol.1101392. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 13.ADA Council on Scientific Affairs and ADA Council on Dental Practice. Infection control recommendations for the dental office and the dental laboratory. J Am Dent Assoc. 1996;127:672–680. doi: 10.14219/jada.archive.1996.0280. [DOI] [PubMed] [Google Scholar]
- 14.ISO 17664-1. Processing of health care products - Information to be provided by the medical device manufacturer for the processing of medical devices - Part 1: Critical and semi-critical medical devices. Geneva; Switzerland: International Standard Organization (ISO); 2021. [Google Scholar]
- 15.Park JH, Olivares-Navarrete R, Baier RE, Meyer AE, Tannenbaum R, Boyan BD, Schwartz Z. Effect of cleaning and sterilization on titanium implant surface properties and cellular response. Acta Biomater. 2012;8:1966–1975. doi: 10.1016/j.actbio.2011.11.026. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 16.Lang R, Hiller KA, Kienböck L, Friedl K, Friedl KH. Influence of autoclave sterilization on bond strength between zirconia frameworks and Ti-base abutments using different resin cements. J Prosthet Dent. 2022;127:617.e1–617.e6. doi: 10.1016/j.prosdent.2022.01.028. [DOI] [PubMed] [Google Scholar]
- 17.Gehrke P, Sing T, Fischer C, Spintzyk S, Geis-Gerstorfer J. Marginal and internal adaptation of hybrid abutment assemblies after central and local manufacturing, respectively. Int J Oral Maxillofac Implants. 2018;33:808–814. doi: 10.11607/jomi.6131. [DOI] [PubMed] [Google Scholar]
- 18.Pils D, Baeppler RJ, Junker R, Kielbassa AM, Nothdurft FP. Application of a standard autoclaving protocol does not harm structural integrity of two-piece zirconia abutments under detachment force testing. Clin Oral Investig. 2019;23:3133–3137. doi: 10.1007/s00784-019-02926-9. [DOI] [PubMed] [Google Scholar]
- 19.Wiedenmann F, Liebermann A, Spintzyk S, Eichberger M, Stawarczyk B. Influence of different cleaning procedures on tensile bond strength between zirconia abutment and titanium base. Int J Oral Maxillofac Implants. 2019;34:1318–1327. doi: 10.11607/jomi.7638. [DOI] [PubMed] [Google Scholar]
- 20.DIN EN 14885:2018. Chemical disinfectants and antiseptics-Application of European Standards for chemical disinfectants and antiseptics. 2018. [Google Scholar]
- 21.Canullo L, Tallarico M, Chu S, Peñarrocha D, Özcan M, Pesce P. Cleaning, disinfection, and sterilization protocols employed for customized implant abutments: an international survey of 100 universities worldwide. Int J Oral Maxillofac Implants. 2017;32:774–778. doi: 10.11607/jomi.5402. [DOI] [PubMed] [Google Scholar]
- 22.Rupp F, Liang L, Geis-Gerstorfer J, Scheideler L, Hüttig F. Surface characteristics of dental implants: A review. Dent Mater. 2018;34:40–57. doi: 10.1016/j.dental.2017.09.007. [DOI] [PubMed] [Google Scholar]
- 23.Bollen CM, Papaioanno W, Van Eldere J, Schepers E, Quirynen M, van Steenberghe D. The influence of abutment surface roughness on plaque accumulation and peri-implant mucositis. Clin Oral Implants Res. 1996;7:201–211. doi: 10.1034/j.1600-0501.1996.070302.x. [DOI] [PubMed] [Google Scholar]
- 24.Canullo L, Penarrocha-Oltra D, Marchionni S, Bagan L, Penarrocha-Diago MA, Micarelli C. Soft tissue cell adhesion to titanium abutments after different cleaning procedures: Preliminary results of a randomized clinical trial. Med Oral Patol Oral Cir Bucal. 2014;19:e177–e183. doi: 10.4317/medoral.19329. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 25.Canullo L, Tallarico M, Peñarrocha-Oltra D, Monje A, Wang HL, Peñarrocha-Diago M. Implant abutment cleaning by plasma of argon: 5-year follow-up of a randomized controlled trial. J Periodontol. 2016;87:434–442. doi: 10.1902/jop.2015.150549. [DOI] [PubMed] [Google Scholar]
- 26.Kern M. On the scientific evidence that the sterilization of customized implant abutments is required. Eur J Oral Implantol. 2015;8:219. [PubMed] [Google Scholar]
- 27.German guideline for the validation of manual cleaning and manual chemical disinfection of medical devices compiled by DGKH-German Society for Hospital Hygiene. DGSV-German Society for Sterile Supply, AKI-working group instrument reprocessing in cooperation with the VAH-association for applied hygiene. 5th ed. Supplement. 2017. pp. 5–14. [Google Scholar]
- 28.Gehrke P, Smeets R, Gosau M, Friedrich RE, Madani E, Duddeck D, Fischer C, Tebbel F, Sader R, Hartjen P. The influence of an ultrasonic cleaning protocol for cad/cam abutment surfaces on cell viability and inflammatory response in vitro. In Vivo. 2019;33:689–698. doi: 10.21873/invivo.11527. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 29.DIN EN ISO 15883-5. Washer-disinfectors - Part 5: Performance requirements and test method criteria for demonstrating cleaning efficacy. Geneva; Switzerland: International Standard Organization (ISO); 2021. [Google Scholar]
- 30.Technical information report AMI TIR30:2011 (R2016). A compendium of processes, materials, test methods, and acceptance criteria for cleaning reusable medical devices. Textbook, Association for the Advancement of Medical Instrumentation. 2011. pp. 9–17. [Google Scholar]
- 31.Guideline compiled by German Society of Hospital Hygiene (DGKH, DGSV and AK). Validation and routine monitoring of automated cleaning and thermal disinfection processes for medical devices. 5th ed. 2017. pp. 1–4. [Google Scholar]
- 32.Kampf G, Bloss R, Martiny H. Surface fixation of dried blood by glutaraldehyde and peracetic acid. J Hosp Infect. 2004;57:139–143. doi: 10.1016/j.jhin.2004.02.004. [DOI] [PubMed] [Google Scholar]
- 33.Happe A, Röling N, Schäfer A, Rothamel D. Effects of different polishing protocols on the surface roughness of Y-TZP surfaces used for custom-made implant abutments: a controlled morphologic SEM and profilometric pilot study. J Prosthet Dent. 2015;113:440–447. doi: 10.1016/j.prosdent.2014.12.005. [DOI] [PubMed] [Google Scholar]
- 34.Piattelli A, Pontes AE, Degidi M, Iezzi G. Histologic studies on osseointegration: soft tissues response to implant surfaces and components. A review. Dent Mater. 2011;27:53–60. doi: 10.1016/j.dental.2010.10.019. [DOI] [PubMed] [Google Scholar]
- 35.Nakajima K, Odatsu T, Shinohara A, Baba K, Shibata Y, Sawase T. Effects of cleaning methods for custom abutment surfaces on gene expression of human gingival fibroblasts. J Oral Sci. 2017;59:533–539. doi: 10.2334/josnusd.16-0681. [DOI] [PubMed] [Google Scholar]
- 36.Geng D, Mao H, Wang J, Zhu X, Huang C, Chen L, Yang H, Xu Y. Protective effects of COX-2 inhibitor on titanium-particle-induced inflammatory osteolysis via the down-regulation of RANK/RANKL. Acta Biomater. 2011;7:3216–3221. doi: 10.1016/j.actbio.2011.05.007. [DOI] [PubMed] [Google Scholar]
- 37.Yang SY, Zhang K, Bai L, Song Z, Yu H, McQueen DA, Wooley PH. Polymethylmethacrylate and titanium alloy particles activate peripheral monocytes during periprosthetic inflammation and osteolysis. J Orthop Res. 2011;29:781–786. doi: 10.1002/jor.21287. [DOI] [PubMed] [Google Scholar]
- 38.Canullo L, Penarrocha D, Micarelli C, Massidda O, Bazzoli M. Hard tissue response to argon plasma cleaning/sterilisation of customised titanium abutments versus 5-second steam cleaning: results of a 2-year post-loading follow-up from an explanatory randomised controlled trial in periodontally healthy patients. Eur J Oral Implantol. 2013;6:251–260. [PubMed] [Google Scholar]
- 39.Kim S, Choi C, Cha Y, Chang JS. The efficacy of convenient cleaning methods applicable for customized abutments: an in vitro study. BMC Oral Health. 2021;21:78. doi: 10.1186/s12903-021-01436-z. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 40.Mehl C, Kern M, Zimmermann A, Harder S, Huth S, Selhuber-Unkel C. Impact of cleaning procedures on adhesion of living cells to three abutment materials. Int J Oral Maxillofac Implants. 2017;32:976–984. doi: 10.11607/jomi.5630. [DOI] [PubMed] [Google Scholar]
- 41.Özcan M, Bernasconi M. Adhesion to zirconia used for dental restorations: a systematic review and meta-analysis. J Adhes Dent. 2015;17:7–26. doi: 10.3290/j.jad.a33525. [DOI] [PubMed] [Google Scholar]
- 42.Inokoshi M, De Munck J, Minakuchi S, Van Meerbeek B. Meta-analysis of bonding effectiveness to zirconia ceramics. J Dent Res. 2014;93:329–334. doi: 10.1177/0022034514524228. [DOI] [PubMed] [Google Scholar]
- 43.Bergamo E, Zahoui A, Luri Amorin Ikejiri L, Marun M, Peixoto da Silva K, G Coelho P, Soares S, A Bonfante E. Retention of zirconia crowns to Ti-base abutments: effect of luting protocol, abutment treatment and autoclave sterilization. J Prosthodont Res. 2021;65:171–175. doi: 10.2186/jpr.JPOR_2019_537. [DOI] [PubMed] [Google Scholar]
- 44.Yan S, Li M, Komasa S, Agariguchi A, Yang Y, Zeng Y, Takao S, Zhang H, Tashiro Y, Kusumoto T, Kobayashi Y, Chen L, Kashiwagi K, Matsumoto N, Okazaki J, Kawazoe T. Decontamination of titanium surface using different methods: an in vitro study. Materials (Basel) 2020;13:2287. doi: 10.3390/ma13102287. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 45.Jatzwauk L, Schöne H, Pietsch H. How to improve instrument disinfection by ultrasound. J Hosp Infect. 2001;48:S80–S83. doi: 10.1016/s0195-6701(01)90019-2. [DOI] [PubMed] [Google Scholar]
- 46.Bentley EM. The value of ultrasonic cleaners in dental practice. Br Dent J. 1994;177:53–56. doi: 10.1038/sj.bdj.4808506. [DOI] [PubMed] [Google Scholar]
- 47.Guan A, Wang Y, Phillips KS. An extraction free modified o-phthalaldehyde assay for quantifying residual protein and microbial biofilms on surfaces. Biofouling. 2018;34:925–934. doi: 10.1080/08927014.2018.1521959. [DOI] [PubMed] [Google Scholar]
- 48.Zuhlsdorf B, Martiny H. Intralaboratory reproducibility of the German test method of prEN ISO 15883-1 for determination of the cleaning efficacy of washer-disinfectors for flexible endoscopes. J Hosp Infect. 2005;59:286–291. doi: 10.1016/j.jhin.2004.09.033. [DOI] [PubMed] [Google Scholar]



