Skip to main content
. Author manuscript; available in PMC: 2023 Sep 1.
Published in final edited form as: Magn Reson Med. 2022 May 23;88(3):1419–1433. doi: 10.1002/mrm.29288

Figure 8:

Figure 8:

Inverse G-factors for representative slice obtained from the 64-channel head-neck and the 32-channel coils. In the brain region, the 64-channel head-neck coil shows slightly lower noise amplifications at all acceleration stages when compared to the 32-channel head coil. The green and red numbers in brackets indicate the mean and maximum G-factors, respectively (non-inverted values). Information for slices in the neck area can be found in Supporting Information Figures S3.