. 2022 Nov 4;13:1039765. doi: 10.3389/fimmu.2022.1039765

Table 4.

Summarized characteristics of newly developed C1-INH complex assays.

C1s/C1-INH complex assay MASP-1/C1-INH complex assay
Standard range 100.0 – 1.6 ng/mL 25.0 – 0.4 ng/mL
Limit of quantification
ULoQ
LLoQ
100.0 ng/mL
0.2 ng/mL
50 ng/mL
0.06 ng/mL
cross-reactivity with uncomplexed components no no
cross-reactivity with animal serum
(mouse, rat, horse, pig, dog)
mouse no
Inter-assay variation < 10% < 10%
Intra-assay variation < 10% < 10%
Matrices to be used
(minimal required dilution)
EDTA plasma (100x)
Citrate plasma (100x)
Heparin plasma (200x)
Serum (200x)
EDTA plasma (5x)
Citrate plasma (5x)
Heparin plasma (5x)
Serum (10x)
Measured complex concentration in healthy individuals, n=96 1846 ± 1060 ng/mL
[mean ± 2SD]
40.5 (13.8-87.9) ng/mL
[median (2.5-97.5 percentile range)]

EDTA, ethylenediaminetetraacetic acid; ULoQ, upper limit of quantification; LLoQ, lower limit of quantification; SD, standard deviation.