Table 2.

Summary of Comparison Between Devices

Parameter Device Mean ± SD Mean of the Difference* ± SD p value Correlation Coefficient
K Steep (D) Argos 44.90 ± 1.71 0.16 ± 0.28 <0.0001 0.986
IOLMaster 700 44.74 ± 1.70
K Flat (D) Argos 42.87 ± 1.60 −0.11± 0.28 0.0021 0.984
IOLMaster 700 42.75 ± 1.59
Corneal Astigmatism (D) Argos 2.03 ± 1.07 0.05± 0.29 0.20 0.964
IOLMaster 700 1.98 ± 1.08
Prediction Error SE (D) Argos 0.13 ± 0.39 0.12 ± 0.36 0.017 0.698
IOLMaster 700 0.25 ± 0.50
Prediction Error Cylinder (D) Argos −0.17 ± 0.43 0.29 ± 0.44 <0.0001; 0.634
IOLMaster 700 0.12 ± 0.56

Notes: *IOLMaster 700 – Argos. Note that P-values for keratometry are based on GEE models with covariates of AL and ACD. P-values for postoperative month 1 refraction are based on GEE models with covariates of lens choice, baseline delta K, AL, and ACD.

Abbreviations: D, diopters; SD, standard deviation; SE, spherical equivalent.