Table 3.

Performance of total antibody screening assays in blind diagnostic evaluation using the validation set (n = 807 samples).

Specificity at defined thresholds (95% CI) Sensitivity for Acute COVID-19 at defined thresholds (95% CI) Sensitivity for Early Convalescent COVID-19 at defined thresholds (95% CI) Sensitivity for Late COVID-19 at defined thresholds (95% CI) Optimal Threshold Value (method)
Threshold* 1 2 3 1 2 3 1 2 3 1 2 3
N Pan ELISA 99.5
(98.4 - 99.8)
98.8
(97.5 - 99.4)
97.4
(95.8 - 98.4)
61.7
(47.2 - 74.4)
61.7
(47.2 - 74.4)
61.7
(47.2 - 74.4)
70.5
(61.1 - 78.4)
77.1
(68.1 - 84.2)
81.9
(73.3 - 88.2)
77.1
(65.9 - 85.5)
81.4
(70.6 - 88.9)
85.7
(75.4 - 92.1)
(2)
RBD Pan ELISA 99.3
(98.2 - 99.7)
99.3
(98.2 - 99.7)
99
(97.7 - 99.5)
38.3
(25.6 - 52.8)
46.8
(33.1 – 61)
57.4
(43.1 - 70.7)
63.8
(54.2 - 72.4)
65.7
(56.1 - 74.2)
73.3
(64.1 - 80.9)
80
(69 - 87.8)
81.4
(70.6 - 88.9)
87.1
(77.1 - 93.2)
(3)
Spike Pan ELISA 97.3
(95.9 - 98.5)
96.7
(95.2 – 98.1)
96.7
(95.2 – 98.1)
76.6
(62.5 - 86.6)
76.6
(62.5 - 86.6)
76.6
(62.5 - 86.6)
93.3
(88.6 – 97.1)
94.3(87.8 - 98.1) 94.3
(87.8 - 98.1)
95.7
(87.5 - 98.6)
95.7
(87.5 - 98.6)
95.7
(87.5 - 98.6)
(1)
Spike-RBD Bridging LIPS 97.6
(96 - 98.6)
96
(94.1 - 97.4)
98.5
(97.1 - 99.2)
80.9
(67.1 - 89.7)
80.9
(67.1 - 89.7)
78.7
(64.8 - 88.2)
92.4
(85.5 - 96.1)
92.4
(85.5 - 96.1)
92.4
(85.5 - 96.1)
94.3
(85.7 - 97.8)
94.3
(85.7 - 97.8)
94.3
(85.7 - 97.8)
(3)

*Threshold method key: 1 - 99th Percentile; 2 - 98th percentile; 3 - Jmax i.e., highest Youden IndexFigure headings.

For each candidate screening assays, one value for specificity of the assays is reported, and sensitivity estimates are provided for COVID-19 samples taken at different time periods (Acute, Early Convalescent and Late Convalescent) post infection are reported for each of the threshold methods. Confidence intervals (95% CI) are given for each estimate. The optimal threshold methods was determined by finding the value which provided at least (or as near to) 98% specificity, whilst offering the optimal sensitivity.