. 2022 Nov 12;11(22):6705. doi: 10.3390/jcm11226705

Table 2.

Protocols for 0.55T and 1.5T for the three most imaged body regions (head, lumbar spine, hip).

Repitition Time (TR) (ms) Echo time (TE) (ms) Slice thickness (ST) (mm) Resolution (mm2) Field-of-view (FOV) (mm2) Time of acquisition (TA) (min)
Lumbar spine T1 Turbo spin echo (TSE) sagittal 1.5T 625 11 4 0.7 × 0.7 300 × 300 02:29
0.55T 454 13 4 0.8 × 0.8 320 × 320 05:26
T2 TSE sagittal 1.5T 3600 102 4 0.7 × 07 300 × 300 01:44
0.55T 3500 99 4 0.8 × 0.8 320 × 320 03:34
T2 TSE axial 1.5T 4210 107 4 0.5 × 0.5 200 × 200 04:40
0.55T 5910 84 4 0.5 × 0.5 200 × 200 04:51
T2 DIXON cor 1.5T 6630 90 5 0.8 × 0.8 300 × 300 05:00
Hip TIRM cor 1.5T 4000 34 3.5 1.1 × 0.9 220 × 220 01:48
0.55T 4440 35 3.5 1.6 × 1.1 220 × 220 02:48
T2 TSE cor 1.5T 3400 73 3.5 0.6 × 0.5 220 × 220 01:25
0.55T 3220 77 3.5 1.0 × 0.7 220 × 220 02:21
TIRM axial 1.5T 5210 54 5 0.8 × 0.6 180 × 180 03:33
0.55T 4260 25 5 1.1 × 0.9 220 × 220 04:58
T1 TSE axial 1.5T 661 9.5 5 0.9 × 0.8 200 × 200 05:44
0.55T 517 9.4 5 1 × 0.7 220 × 220 03:40
T1 TSE sagittal 1.5T 596 8.5 5 0.9 × 0.7 220 × 220 02:32
0.55T 517 9.4 5 1 × 0.7 220 × 220 03:40
Head Fluid attenuated inversion recovery (FLAIR) axial 1.5T 8510 112 3 0.9 x 0.9 187 × 230 03:26
0.55T 7780 96 3 1.3 × 1 209 × 230 05:28
Susceptibility weighted imaging (SWI) axial 1.5T 48 40 3 1.1 × 0.96 194 × 230 02:17
0.55T 172 100 3 0.9 x 0.8 201 × 230 02:23
Diffusion weighted imaging (DWI) (echo-planar imaging(EPI)) axial 1.5T 6200 103 3 1.4 × 1.4 220 × 220 02:04
0.55T 7400 102 3 1.7 × 1.7 230 × 230 04:35