Table 1.

Test agent generation parameters.

Chemical Name CAS # Vendor Catalog Number Gas Mixture or Lot # and Purity GC Conditions/Methodology Generation Method Preparation for Method
Acrolein 107-02-8 Airgas National Welders Custom Acrolein in Nitrogen at 1011 PPM Isothermal Oven Temperature: 28 °C held for 2.8 min; Column Flow: 10 mL/min; Splitless inlet at 200 °C; FID detector at 250C Gas Cylinder Acrolein in Nitrogen Target = 1000 PPM, Actual = 1011 PPM, Accuracy = ±2%
Acetaldehyde 75-07-0 Airgas National Welders Custom Acetaldehyde in Nitrogen at 2064 PPM Isothermal Oven Temperature: 45 °C held for 1.5 min; Column Flow: 10 mL/min; Splitless inlet at 200 °C; FID detector at 250C Gas Cylinder Acetaldehyde in Nitrogen, Target = 2000 PPM, Actual = 2064 PPM, Accuracy = ±2%
1-Bromopropane 106-94-5 Sigma Aldrich B78106 Lot #: WXBC0691V Purity: 99.8% Isothermal Oven Temperature: 45 °C held for 1.5 min; Column Flow: 10 mL/min; Splitless inlet at 200 °C; FID and ECD detector at 250C Liquid in vial Chemical placed in vial (25mL) with orifice which was placed into the dynamic headspace generator to carry VOCs in air stream into chamber
Formaldehyde (paraformaldhyede) 50-00-0 (30525-89-4) Sigma Aldrich 158127 Lot #: STBB9507V Purity: 94.1% Non-dispersive infrared analyzers Solid in vial Chemical placed in vial (25mL) with orifice which was placed into the dynamic headspace generator to carry VOCs in air stream into chamber
1,3-Butadiene 106-99-0 Airgas National Welders Custom 1,3 Butadiene in Air at 511 PPM Isothermal Oven Temperature: 27 °C held for 1.0 min; Column Flow: 10 mL/min; Splitless inlet at 200 °C; FID detector at 250C Gas Cylinder 1,3 Butadiene in Air Target = 500 PPM, Actual = 511 PPM, Accuracy = ±2%
Dichloromethane 75-09-2 Burdick & Jackson 300-4 Lot #: C0674 Purity: 99.9% Isothermal Oven Temperature: 45 °C held for 2.5 min; Column Flow: 10 mL/min; Splitless inlet at 200 °C; FID and ECD detector 250C Liquid in vial Chemical placed in vial (25mL) with orifice which was placed into the dynamic headspace generator to carry VOCs in air stream into chamber
Trichloroethylene 79.01-6 Sigma Aldrich 251402 Lot #: SHBJ3703 Purity: 99.98% Isothermal Oven Temperature: 60 °C held for 4 min; Column Flow: 10 mL/min; Splitless inlet at 200 °C; FID and ECD detector at 250C Liquid in vial Chemical placed in vial (25mL) with orifice which was placed into the dynamic headspace generator to carry VOCs in air stream into chamber
Carbon Tetrachloride 56-23-5 Sigma Aldrich 319961 Lot #: SHBJ0149 Purity: 99.90% Isothermal Oven Temperature: 60 °C held for 4 min; Column Flow: 10 mL/min; Splitless inlet at 200 °C; FID and ECD detectors at 250C Liquid in vial Chemical placed in vial (25mL) with orifice which was placed into the dynamic headspace generator to carry VOCs in air stream into chamber