. 2022 Nov 28;80(1):18–29. doi: 10.1001/jamaneurol.2022.4166

Table 4. Spinal Cord Stimulator (SCS)–Related Complications and Revisions or Removals Among 1260 Patients Within 24 Months After Permanent Device Implantation.

Complications/revisions or removals No. of months after SCS placement No. (%)
Complications
Breakdown of lead/generator 1-12 56 (4.4)
13-24 16 (1.3)
Displacement of lead/generator 1-12 22 (1.8)
13-24 NA (NA)a
Infection/inflammation of lead/generator 1-12 26 (2.1)
13-24 NA (NA)
Other mechanical complications of lead/generator 1-12 117 (9.3)
13-24 51 (4.1)
Any complication of lead/generator 1-12 176 (14.0)
1-24 226 (17.9)
Revision or removals
Revision of lead/generator 1-12 184 (14.6)
13-24 75 (6.0)
Lead removal 1-12 95 (7.5)
13-24 50 (4.0)
Generator removal 1-12 23 (1.8)
13-24 NA (NA)
Any removal/revision of lead/generator 1-12 217 (17.2)
1-24 279 (22.1)

Abbreviation: NA, not applicable.

a

Small numbers (n <11) cannot be reported according to the Optum Labs cell size suppression policy.