Sensitivity and specificity of AF and sinus rhythm diagnosis with Kardia Mobile 6L and Istel HR 2000 against standard 12-lead ECG

Device Standard 12-lead ECG Kardia Mobile 6L Istel HR 2000
Sinus rhythm n = 59 (60%) n = 56 (58%) n = 60 (61%)
 Sensitivity 100% 88.1% 91.5%
 Specificity 100% 89.7% 84.6%
Atrial fibrillation n = 22 (22%) n = 21 (22%) n = 18 (18%)
 Sensitivity 100% 86.4% 77.3%
 Specificity 100% 97.4% 98.7%