Table 1.
Troubleshooting guide for imaging of fixed and immunostained cells with DNA-PAINT-ERS
| Problem | Possible Cause | Solution |
|---|---|---|
| Low conjugation efficiency for DS-DBCO or IS-dye | Loss of NHS-ester activity or improper pH for the reaction buffer | Use fresh reagents Confirm pH of the reaction buffer |
| Low recovery of DS-DBCO or IS-dye conjugates | Potential problem with DNA precipitation | Repeat DNA precipitation with DNA LoBind tubes Keep sample at −80°C degrees for longer Spin sample down at higher rpm or for prolonged time |
| Concentration of azide-tagged antibody is low | Problem with ultrafiltration | Passivate the spin filter with TWEEN 20 Check spin filter’s expiration date – the membranes used in spin filters occasionally fail |
| Low recovery of DS-conjugated antibody | Problem with ultrafiltration Failed conjugation |
Check spin filter’s expiration date – the membranes used in spin filters occasionally fail Use fresh reagents Check pH of the reaction buffer |
| Unspecific staining | DBCO non-specifically bound to cellular targets Dirty coverglass |
Quench DBCO with azide repeatedly Validate specificity of primary antibody Optimize blocking steps (% BSA, salmon DNA, signal enhancer) during low resolution imaging experiments Etch and clean coverglass |
| Low staining signal | Insufficient permeabilization Insufficient concentration of antibodies Too much blocking Solutions not fresh |
Use triton instead of saponin (disadvantageous for membrane targets) Increase antibody concentration Block with lower % BSA Prepare fresh reagents |
| Nuclear staining | DBCO bound to nucleus | Add signal enhancer incubation step (see Basic protocol 2 step 13) Add salmon sperm DNA for additional blocking (see Basic protocol 2 step 15) |
| Uneven illumination after sample mounting | Bubbles in immersion oil Old oil smear on bottom of the coverglass from previous imaging session |
Clean coverglass with isopropanol and objective lens with lens cleaning paper Remount sample with new immersion oil |
| Low signal to noise ratio during imaging | High IS concentration Wrong TIRF angle |
Lower IS concentration to reduce background signal Adjust TIRF angle to select signal relevant to imaging plane of the target |
| Too slow/Fast DS/IS kinetics | Low/high room temperature EC concentration Orthogonal DS-IS pairs |
Adjust EC concentration (5–15 %) Mix EC properly Monitor room temperature Check specific kinetics of DS-IS sequences Confirm the correct DS-IS pair is used |
| Unstable temperature in microscope room | Central climate control | Adjust EC concentration (5–15 %) Monitor room temperature and image during certain time windows |
| Focus drift | No focus stabilization Temperature changes during image acquisition Missing lid on the 8-well chambered coverglass |
Introduce focus stabilization Equilibrate samples and imaging buffers at room temperature if previously stored in the fridge Place correct lid on top of the chambered coverglass to minimize evaporation of imaging solution |
| Lack of gold fiducials | Gold solution needs to be more concentrated or incubated for longer | Apply gold nanoparticles when sample is mounted on microscope, with laser illumination to assess proper amount of gold fiducials for a successful drift correction |