Cross tabulation of the detection of Schistosoma haematobium eggs by the Schistoscope 5.0 and conventional microscopy performed on 487 urines collected at baseline screening

Conventional microscopy
Schistoscope 5.0 Positive (n = 166) Negative (n = 321) Total (N = 487)
Semi-automated digital microscope Positive 133 15 148
Negative 33 306 339
Fully automated digital microscope Positive 145 164 309
Negative 21 157 178