Table 1.

mCycle wish-list for auto-planning of cervical cancer (50 Gy in 25 fractions).

Clinical Constraints
Structure Cost Function
(parameter)
Shrink margin (cm) Limit
PTV Quadratic Overdose (52 Gy) / <0.02 Gy
Bowel bag Overdose DVH (45 Gy) / <195.0 cm3
External Maximum Dose / 53.4 Gy
Planning Constraints
Structure Cost Function
(parameter values)
Shrink margin (cm) Limit
External Quadratic Overdose (50 Gy) 0.0 <0.1 Gy
External Quadratic Overdose (45 Gy) 0.3 <0.2 Gy
External Quadratic Overdose (40 Gy) 0.6 <0.2 Gy
External Quadratic Overdose (35 Gy) 0.9 <0.2 Gy
External Quadratic Overdose (25 Gy) 2.5 <0.3 Gy
Objectives
Priority Structure Cost Function
(parameter values)
Shrink margin (cm) Goal value (sufficient)
1 PTV Target EUD (0.5) 50.0 Gy
1 PTV Target Penalty (99%) 50.0 Gy
2 Rectum Parallel (40 Gy, k = 3) 0.3 <30.0%
2 Bladder Parallel (40 Gy, k = 3) 0.3 <33.5%
3 Rectum Serial (k = 15) <46.0 Gy
3 Bowel bag Parallel (40 Gy, k = 3) 0.3 <20.0%
3 Bowel bag Serial (k = 15) <43.0 Gy
3 Bladder Serial (k = 15) <47.0 Gy
4 Bowel bag Overdose DVH (45 Gy) <14.0% (7.0%)
4 Right femoral head Serial (k = 15) <38.0 Gy
4 Left femoral head Serial (k = 15) <38.0 Gy
5 External Conformality <0.75
6 Right femoral head Serial (k = 1) <30.0 Gy
6 Left femoral head Serial (k = 1) <30.0 Gy
7 Rectum Serial (k = 1) <30.0 Gy
8 Bladder Serial (k = 1) <35.0 Gy

Priority: order list according to which the objectives (cost functions) are optimized. Shrink margin: creates a buffer zone between the PTV and overlapping structures to avoid conflict between the applied cost functions of each of the structures. PTV, planning target volume; DVH, dose volume histogram; EUD, equivalent uniform dose.