Table VII -.

ACR DIR dose reference levels

Reference size range (cm) No. of patients Achieveable dose (mGy) Diagnostic reference level (mGy)
21-25 29691 7 9
25-29 82822 9 11
29-33 108921 12 15
33-37 76681 17 21
37-41 30640 21 24

Data collated from Kanal and colleagues [3]

Note- Size was measured as water-equivalent-diameter. Achievable dose is the 50th percentile among surveyed facilities. Diagnostic reference level is the 75th percentile among surveyed facilities.