TABLE I.

Performance Summary and Comparison with the State-of-the-Art

Specifications This Worka [7] [29] [30] [31]
Technology 180 nm HV BCD 350 nm HV CMOS 180 nm HV BCD 180 nm HV CMOSb 180 nm HV CMOS
Transducer CMUT CMUT CMUT CMUT PZT
Element Level Imp. Matching YES NO NO NO NO
TX/RX Array 16-TX/1-RX 56-TX/48-RX 64 TRX 4-TX/4-RX 7 TRX
Bandwidth (MHz) 2–4 40 7 5.2 5
Element Cap. Load 25 pF 90 fF 15 pF 40 pF 79.9 pF
Avg. Power (mW) 3.6 @ 1.8 V 20 4.9c 14.3 N/A
Area (mm) 2.6 1.54 29.7 1.5d 2
TX Max Pulse (V) 60 30 NA 30 32
RX Gain (dB) 74–90 106 106 96.6 N/A
Input-Referred Noise (pA/√Hz) 2–3 @3 MHz 0.31@20 MHze 2 @5 MHz 1.35 @3 MHzf N/A
Application Intracranial Ultrasound Intracardiac Echocardiography Intracardiac Echocardiography Medical Ultrasound Fitness Tracking
a

Includes BGR and bias generation circuitry.

b

Gate-oxide can handle 30-V swing.

c

It indicates the TIA power consumption.

d

Excluding the pads

e

TIA I-R noise with 90-fF CMUT

f

Calculated based on reported values