TABLE I.
Performance Summary and Comparison with the State-of-the-Art
Specifications | This Worka | [7] | [29] | [30] | [31] |
---|---|---|---|---|---|
Technology | 180 nm HV BCD | 350 nm HV CMOS | 180 nm HV BCD | 180 nm HV CMOSb | 180 nm HV CMOS |
Transducer | CMUT | CMUT | CMUT | CMUT | PZT |
Element Level Imp. Matching | YES | NO | NO | NO | NO |
TX/RX Array | 16-TX/1-RX | 56-TX/48-RX | 64 TRX | 4-TX/4-RX | 7 TRX |
Bandwidth (MHz) | 2–4 | 40 | 7 | 5.2 | 5 |
Element Cap. Load | 25 pF | 90 fF | 15 pF | 40 pF | 79.9 pF |
Avg. Power (mW) | 3.6 @ 1.8 V | 20 | 4.9c | 14.3 | N/A |
Area (mm) | 2.6 | 1.54 | 29.7 | 1.5d | 2 |
TX Max Pulse (V) | 60 | 30 | NA | 30 | 32 |
RX Gain (dB) | 74–90 | 106 | 106 | 96.6 | N/A |
Input-Referred Noise (pA/√Hz) | 2–3 @3 MHz | 0.31@20 MHze | 2 @5 MHz | 1.35 @3 MHzf | N/A |
Application | Intracranial Ultrasound | Intracardiac Echocardiography | Intracardiac Echocardiography | Medical Ultrasound | Fitness Tracking |
Includes BGR and bias generation circuitry.
Gate-oxide can handle 30-V swing.
It indicates the TIA power consumption.
Excluding the pads
TIA I-R noise with 90-fF CMUT
Calculated based on reported values