Skip to main content
. 2022 Dec 1;18(12):2703–2712. doi: 10.5664/jcsm.10194

Table 3.

Diagnostic performance of ANNE sleep vs PSG at AHI thresholds.

AHI Cutoff (events/h) Sensitivity (95% CI) Specificity (95% CI) AUC (95% CI) Accuracy PPV (95% CI) NPV (95% CI) LR+ LR−
5 0.96 (0.92–0.99) 0.86 (0.77–0.92) 0.91 (0.87– 0.95) 0.92 0.91 (0.85–0.95) 0.94 (0.87–0.98) 6.7 0.04
15 0.90 (0.8–0.96) 0.97 (0.94–0.99) 0.94 (0.90–0.97) 0.95 0.94 (0.85–0.98) 0.96 (0.91–0.98) 35.2 0.11
30 0.65 (0.45–0.81) 0.99 (0.96–0.99) 0.82 (0.73–0.90) 0.94 0.91 (0.71–0.99) 0.95 (0.91–0.97) 62.6 0.36

AHI = apnea-hypopnea index, AUC = area under the curve, CI = confidence interval, LR = likelihood ratio, NPV = negative predictive value, PPV = positive predictive value, PSG = polysomnography.