Skip to main content
. 2022 Nov 24;2022:9112961. doi: 10.1155/2022/9112961

Table 3.

Bland–Altman analysis of the three evaluated glucometers with the reference method in dogs.

Coefficients
Glucometer Bias Std. deviation of bias Agreement (CI 95%)
Freedom Lite® 47,05 31,82 −25,39 119,49
Optium Neo® 6,23 17,08 −32,65 45,10
On Call Plus® 13,08 21,28 −35,36 61,52