Skip to main content
. 2022 Nov 24;2022:9112961. doi: 10.1155/2022/9112961

Table 5.

Percentage and absolute values of blood glucose concentration obtained using the three glucometers, according to ISO 15197 : 2013 standards in dogs.

Established values Freedom Lite OptiumNeo OnCall Plus
n Total % n Total % N Total %
<100 mg/dL: Difference of+/- 15 mg/dL 8 50 16 48 50 96 49 50 98
≥100 mg/dL: Difference of +/-15% 0 65 0.0 58 65 89.2 42 65 64.6
Total within acceptable limits 8 115 7.0 106 115 92.2 91 115 79.1