Skip to main content
. 2022 Dec 2;22:300. doi: 10.1186/s12902-022-01216-y

Table 2.

Aldosterone concentration with inferred effects of cross-reactions

Laboratory examinations Baseline SPAC®-S Aldosterone kit
Crossover rates (%) Concentrations that could have been measured as aldosterone (nmol/L)
Progesterone (nmol/L) 21.6 0.008 0.001728
Deoxycorticosterone (pmol/L) 5870 0.05 0.002935
Corticosterone (nmol/L) 308 0.03 0.0924
The total concentration that could have been measured as aldosterone (nmol/L) 0.097063
Large discrepancy between the theoretical and measured values
Measured aldosterone concentration (nmol/L) 0.4551