Table 2.

Parameters required for the proposed HPLC method's system suitability testing

Parameters PAP PCM PHA ETO Reference value (50)
Resolution (Rs)a 12.290 14.498 20.296 Rs > 2
Selectivity factor (α)b 2.372 1.545 1.504 α > 1
Capacity factor (K) 2.456 5.827 7.005 8.549 1 < K < 10
Number of theoretical plates (N) 2437 9838 19595 41923 N > 2000
Tailing factor (T) 1.550 1.352 1.352 1.280 T ≤ 2

a Resolution (Rs) = 2(tRB—tRA) / (WB+WA), where tR is the retention time and w is the peak width calculated for each of the two successive peaks

b Selectivity (α) = k’2/k’1 calculated for each of two successive peaks