. 2022 Jul 7;4(3):100293. doi: 10.1016/j.ocarto.2022.100293

Table 3.

Assessment of intra and inter plate coefficient of variation (CV) using synovial fluid, vendor plasma quality control (QC) and serum QC samples. Only data from biomarkers that had values above lower limit detection are presented.

Synovial fluid Vendor plasma QC
Intra plate Biomarkers
Total, n 63–72 71–73
Intra plate CV %, mean (range) 6.3 (0.0–25.0) 6.2 (0.0–30.1)
With CV ​< ​20%, n (%) 61–72 (97–100) 59–72 (80–100)
aExperiments, n 2 3
Samples, n (dilution) 4 (4x) 1 (1x)
Repeats/plate, n 2 or 4 2
Plates, n 2 32
Synovial fluid Vendor plasma QC Serum QC
Inter plate Biomarkers
Total, n 66 67–73 78
Inter plate CV %, mean (range) 19.1 (6.7–49.6) 13.5 (0.4–79.3) 14.8 (5.3–92.5)
With CV ​< ​20%, n (%) 42 (64) 59–70 (81–99) 68 (87)
Experiments, n 1 3 1
Samples, n (dilution) 1 (4x) 1 (1x) 1 (1x)
Repeats/plate, n 1 2 1
Plates, n 9–12 34 15–19
a

Experiments ​= ​analyses done at one, two or three different occasions at Olink.