Table 2.

The comparison of axial lengths (AL) between Myopia master, IOLMaster 700 and IOLMaster 500

Device Mean±SD (min,max) Compared test
(p value)
Correlation test ICC Difference of
the means
95% LOA
r P Lower Upper
Myopia master 23.67±1.26(20.50,26.86) 0.741

all

0.999

all

<0.001

0.999

(0.998,0.999)

0.02 -0.07 0.12
IOL Master 700 23.70±1.25(20.49,26.89)
Myopia master as above 0.908

0.999

(0.999,0.999)

0.01 -0.09 0.12
IOL Master 500 23.68±1.26(20.52,26.85)
IOL Master 700 as above 0.819

0.999

(0.999,1)

0.01 -0.07 0.09
IOL Master 500 as above

SD standard deviation, ICC (A, 1) intraclass correlation coefficiens (inter-rater reliability, two-way random effect model), LOA limit of agreements