Table 4.

The comparison of refraction spherical, refraction cylinder and refraction equivalent between Myopia master, ARK-1 and manifest refraction

Device Mean±SD (min,max) Compared test
(p value)
Correlation test ICC Difference of
the means
95% LOA
r P Lower Upper
Spherical Refraction Myopia master -0.95±2.13(-8.00,7.75) 0.01 0.954

all

<0.001

0.976

(0.970,0.982)

0.36 -0.91 1.63
Manifest refraction -0.57±2.16(-7.75,8.00)
Myopia master as above 0.016 0.988

0.994

(0.992,0.995)

0.37 -0.29 1.04
ARK-1 -0.57±2.16(-7.50,8.50)
Manifest refraction as above 0.901 0.953

0.976

(0.969,0.981)

-0.01 -1.30 1.28
ARK-1 as above
Cylindrical Refraction Myopia master -0.79±0.71(-4.50,0) 0.002 0.748

0.858

(0.817,0.890)

0.17 -0.76 1.11
Manifest refraction -0.6±0.63(-3.50,0)
Myopia master as above 0.375 0.852

0.933

(0.914,0.948)

0.05 -0.63 0.73
ARK-1 -0.73±0.68(-4.00,0)
Manifest refraction as above 0.024 0.812

0.908

(0.881,0.928)

0.13 -0.61 0.86
ARK-1 as above
Spherical Equivalent Myopia master -1.35±2.12(-8.63,6.75) 0.001 0.960

0.939

(0.799,0.972)

0.47 -0.77 1.71
Manifest refraction -0.86±2.11(-8.50,7.13)
Myopia master as above 0.008 0.984

0.968

(0.768,0.989)

0.40 -0.36 1.16
ARK-1 -0.93±2.13(-8.38,7.38)
Manifest refraction as above 0.646 0.960

0.959

(0.947,0.968)

0.06 -1.12 1.24
ARK-1 as above

Values with statistical significance are shown in bold

SD standard deviation, ICC (A, 1) intraclass correlation coefficients (inter-rater reliability, two-way random effect model), LOA limit of agreements