. 2021 Aug 27;17(6):481–488. doi: 10.4244/EIJ-D-20-01000

Table 2. Angiographic* and procedural characteristics.

Super high-pressure balloon (n=37) Scoring balloon (n=37) p-value
Angiographic characteristics
Target lesion location
Left anterior descending 56.8% (21/37) 51.4% (19/37) 0.66
Left circumflex 13.5% (5/37) 21.6% (8/37)
Right coronary artery 29.7% (11/37) 27.0% (10/37)
Bifurcation 40.5% (15/37) 37.8% (14/37) >0.99
Moderate/severe tortuosity 13.5% (5/37) 21.6% (8/37) 0.74
Chronic total occlusion 2.7% (1/37) 5.4% (2/37) >0.99
B2/C lesion 100% (37/37) 100% (37/37) >0.99
Severe calcification 100% (37/37) 97.3% (36/37) >0.99
Vessel diameter, mm 3.1±0.5 2.9±0.4 0.11
Lesion length, mm 23.3±11.5 24.8±12.1 0.58
Diameter stenosis, pre, % 69.1±10.2 70.6±10.7 0.53
Minimal lumen diameter, pre, mm 0.85±0.36 0.82±0.33 0.12
Procedural characteristics
Guiding catheter
6 Fr 43.2% (16/37) 41.7% (15/37) >0.99
7 Fr 51.4% (19/37) 55.6% (20/37)
Predilation
Maximal predilation balloon diameter, mm 3.1±0.4 3.0±0.4 0.18
Maximal predilation balloon pressure, bar 33.8±6.5 17.0±4.0 <0.001
Large residual dissection (>5 mm) 8.1% (3/37) 13.5% (5/37) 0.71
Stent/lesion 1.5±0.8 1.6±0.7 0.88
Total stent length, mm 39.6±18.0 41.0±21.5 0.75
Minimal stent diameter, mm 3.2±0.5 3.1±0.5 0.38
Maximal stent diameter, mm 3.4±0.5 3.2±0.4 0.24
Maximal stent implantation pressure, bar 16.5±3.5 16.1±2.9 0.62
Post-dilation 78.4% (29/37) 83.8% (31/37) 0.77
Maximal post-dilation balloon diameter, mm 3.7±0.5 3.4±0.6 0.11
Maximal post-dilation balloon pressure, bar 21.3±7.9 19.9±3.8 0.38
Diameter stenosis, post, % 11.6±4.8 14.4±5.6 0.02
Minimal lumen diameter, post, mm 2.83±0.34 2.65±0.36 0.03
Acute lumen gain, mm 1.89±0.42 1.83±0.45 0.60
Compromised side branch 13.5% (5/37) 13.5% (5/37) >0.99
Values are n (%) or mean±standard deviation. *As adjudicated by the angiographic core laboratory. Study device.