| Acceptability criteria | ||
|---|---|---|
| MS calibration ≤ 7 days | Day of injection (11.04.19) | Passed |
| Bias on Pascal Triangle sample ≤ 5% | PEP: −1.4% max | Passed |
| FruBP: −2.5% max | Passed | |
| ATP: 2.4% max | Passed |
| Acceptability criteria | ||
|---|---|---|
| MS calibration ≤ 7 days | Day of injection (11.04.19) | Passed |
| Bias on Pascal Triangle sample ≤ 5% | PEP: −1.4% max | Passed |
| FruBP: −2.5% max | Passed | |
| ATP: 2.4% max | Passed |