Table 2.

Measured FiO2 for varying tidal volume (Vt), respiratory rate (RR), I:E ratio (ti/te), and nasal cannula flow rate (QNC)

Vt (ml) RR (/min) ti/te QNC (l/min) FiO2 (%)
17.1 65 1:1 0.5 36.9 ± 1.4
17.1 65 1:1 1.0 53.8 ± 3.1
17.1 65 1:1 2.0 81.1 ± 1.7
17.1 65 3:4 0.5 36.5 ± 1.6
17.1 65 3:4 1.0 50.7 ± 1.5
17.1 65 3:4 2.0 76.6 ± 0.5
26.8 65 1:1 0.5 30.6 ± 0.7
26.8 65 1:1 1.0 42.1 ± 2.3
26.8 65 1:1 2.0 62.6 ± 2.7
26.8 65 3:4 0.5 30.2 ± 0.7
26.8 65 3:4 1.0 40.7 ± 2.0
26.8 65 3:4 2.0 58.2 ± 0.7
36.6 65 1:1 0.5 28.0 ± 0.7
36.6 65 1:1 1.0 36.4 ± 1.9
36.6 65 1:1 2.0 53.4 ± 1.8
36.6 65 3:4 0.5 27.9 ± 0.5
36.6 65 3:4 1.0 35.9 ± 1.1
36.6 65 3:4 2.0 49.8 ± 1.0
17.1 50 1:1 0.5 41.0 ± 4.2
17.1 50 1:1 1.0 61.9 ± 2.5
17.1 50 1:1 2.0 88.6 ± 2.7
17.1 50 3:4 0.5 40.0 ± 0.8
17.1 50 3:4 1.0 67.6 ± 2.7
17.1 50 3:4 2.0 83.9 ± 3.2
26.8 50 1:1 0.5 36.5 ± 2.4
26.8 50 1:1 1.0 49.4 ± 2.3
26.8 50 1:1 2.0 72.4 ± 2.7
26.8 50 3:4 0.5 33.8 ± 1.1
26.8 50 3:4 1.0 44.8 ± 3.6
26.8 50 3:4 2.0 66.9 ± 2.4
36.6 50 1:1 0.5 32.5 ± 2.2
36.6 50 1:1 1.0 42.6 ± 1.5
36.6 50 1:1 2.0 59.8 ± 0.3
36.6 50 3:4 0.5 42.3 ± 2.8
36.6 50 3:4 1.0 30.3 ± 1.1
36.6 50 3:4 2.0 56.5 ± 2.1

Values of FiO2 are reported as average ± standard deviation. The number of replicas (n) used for calculating average and standard deviation is n = 4 for QNC = 0.5 l/min, n = 7 for QNC = 1 l/min, and n = 3 for QNC = 2 l/min. Data are tabulated for fully inserted nasal cannula prongs the and sinusoidal breathing pattern