Table 2.
Mean differences between angular measurement results of parallel (0°) and varying off-z-axis (15° and 45° deviated) double oblique positioning of the femurs on the CT table in the gantry during scanning to test for the independency of femoral positioning.
|
Femurs: n = 13 Positioning: Three times CT-scans: 39 |
Subtraction of the results of different positioning angles | Mean | p -value |
|---|---|---|---|
| Femoral torsion angle-neck center | FTA-NC (0°) - FTA-NC (15°) | −0.626 | 0.075 |
| (FTA-NC) | FTA-NC (0°) - FTA-NC (45°) | −0.933 | 0.201 |
| FTA-NC (15°)–FTA-NC (45°) | −0.308 | 0.362 | |
| Femoral torsion angle-neck base | FTA-NBC (0°)–FTA-NBC (15°) | −0.400 | 0.145 |
| center (FTA-NBC) | FTA-NBC (0°)–FTA-NBC (45°) | −0.523 | 0.096 |
| FTA-NBC (15°)–FTA-NBC (45°) | −0.123 | 0.712 | |
| Femoral neck inclination angle-neck | FIA-NC (0°)–FIA-NC (15°) | −0.662 | 0.124 |
| center (FIA-NC) | FIA-NC (0°)–FIA-NC (45°) | −0.813 | 0.170 |
| FIA-NC (15°)–FIA-NC (45°) | −0.151 | 0.644 | |
| Femoral neck inclination angle-neck | FIA-NBC (0°)–FIA-NBC (15°) | 0.659 | 0.131 |
| base center (FIA-NBC) | FIA-NBC (0°)–FIA-NBC (45°) | 1.29 | 0.002 |
| FIA-NBC (15°)–FIA-NBC (45°) | 0.628 | 0.163 | |
| Femoral varus or valgus angle | FVA-td (0°)–FVA-td (15°) | 0.359 | 0.089 |
| total-distal (FVA-td) | FVA-td (0°)–FVA-td (45°) | 0.364 | 0.055 |
| FVA-td (15°)–FVA-td (45°) | 0.005 | 0.966 | |
| Femoral varus or valgus angle | FVA-pd (0°)–FVA-pd (15°) | 0.449 | 0.015 |
| proximal-distal (FVA-pd) | FVA-pd (0°)–FVA-pd (45°) | 0.31 | 0.154 |
| FVA-pd (15°)-FVA-pd (45°) | 0.139 | 0.283 |