Skip to main content
. Author manuscript; available in PMC: 2022 Dec 9.
Published in final edited form as: Nat Protoc. 2022 Feb 4;17(3):748–780. doi: 10.1038/s41596-021-00665-4

Table 3 ∣.

Troubleshooting table.

Steps Problem Possible reason Solution
10 No product G0-C14 is stuck on the silica gel Increase the percentage of MeOH to increase the polarity of the eluent
14 Low yield of G0-C14 product The amines of G0 are partially oxidized, resulting in a yellow G0 solution Always store the G0 solution in a 4 °C fridge and flush the bottle with argon after each use to protect the amines from oxidization.
24 Low yield of DSPE-PEG-S2P product The peptide S2P is not completely dissolved. Warm up the solution at 37 °C in a water bath or add more HEPES buffer.
42 Precipitation of siRNA The vigorous mixing process changes the homogenous state of the siRNA/G0-C14 solution Mixing the siRNA/G0-C14 solution by gentle pipetting. Do not vortex the solution
57 Poor reproducibility of DLS results The siLuc NPs are too diluted and below the DLS detection limit or the extreme dilution destabilizes the NPs. Adjust the dilution factor according to the detection limit of the DLS instrument and do not dilute too much.
68 Poor contrast of TEM images The volume of staining solution is not large enough or incubation time is too short Increase the volume of the staining solution or increase the incubation time
Very few samples in TEM images The sample is too diluted Increase the concentration of the sample, or repeat the sample loading procedure several times
The background is too dark in TEM images The incubation time for the staining solution is too long Decrease the incubation time of the staining solution
91 The gel bands are too weak The siRNA concentration is too low Before gel electrophoresis, concentrate the siRNA samples using Amicon tubes (MWCO, 10 kDa)
101 Low siRNA EE Fail to transfer all the Dy647-siLuc NPs from the Amicon tube Pipette the Dy647-siLuc NPs solution vigorously before transferring; Also, wash the tube one or two times with a small volume of PBS.
121 Low fluorescence signals of Dy647 The concentration of Dy647-siLuc NPs is too low or the cell density is too high Raise the concentration of Dy647-siLuc NPs or decrease the cell density.
131 Low bioluminescence signal The bioluminescence signal decays Prepare and warm up the imaging system before adding the substrate. Upon adding the substrate, finish the measurement within 10 min.
Low gene silencing efficiency G0-C14 is degraded The G0-C14 can be easily degraded after dissolution in the solvent. Aliquot the G0-C14 solution and store aliquots at −20 °C. Do not use the same batch more than three times.
152 All the samples are PI positive The PI staining process is too long Shorten the PI staining process; Finish the measurement within 1 h after the PI staining
165 The centrifugation takes too much time The concentration of siRNA NPs is too high, making the ultrafiltration difficult Split the samples into more Amicon tubes for washing, combine and transfer all the samples to one Amicon tube to adjust the concentration of NPs for in vivo injection
166 The volume of the NPs is too large for injection The siRNA concentration is too low Add less PBS to redisperse the NPs during the preparation of NPs
176 Weak fluorescence signal of the sections Low activity of the primary antibody or the primary antibody concentration is too low Handle the primary antibodies on ice to retain their optimal activity; Decrease the dilution factor of the primary antibodies.