10
|
No product |
G0-C14 is stuck on the silica gel |
Increase the percentage of MeOH to increase the polarity of the eluent |
14
|
Low yield of G0-C14 product |
The amines of G0 are partially oxidized, resulting in a yellow G0 solution |
Always store the G0 solution in a 4 °C fridge and flush the bottle with argon after each use to protect the amines from oxidization. |
24
|
Low yield of DSPE-PEG-S2P product |
The peptide S2P is not completely dissolved. |
Warm up the solution at 37 °C in a water bath or add more HEPES buffer. |
42
|
Precipitation of siRNA |
The vigorous mixing process changes the homogenous state of the siRNA/G0-C14 solution |
Mixing the siRNA/G0-C14 solution by gentle pipetting. Do not vortex the solution |
57
|
Poor reproducibility of DLS results |
The siLuc NPs are too diluted and below the DLS detection limit or the extreme dilution destabilizes the NPs. |
Adjust the dilution factor according to the detection limit of the DLS instrument and do not dilute too much. |
68
|
Poor contrast of TEM images |
The volume of staining solution is not large enough or incubation time is too short |
Increase the volume of the staining solution or increase the incubation time |
Very few samples in TEM images |
The sample is too diluted |
Increase the concentration of the sample, or repeat the sample loading procedure several times |
The background is too dark in TEM images |
The incubation time for the staining solution is too long |
Decrease the incubation time of the staining solution |
91
|
The gel bands are too weak |
The siRNA concentration is too low |
Before gel electrophoresis, concentrate the siRNA samples using Amicon tubes (MWCO, 10 kDa) |
101
|
Low siRNA EE |
Fail to transfer all the Dy647-siLuc NPs from the Amicon tube |
Pipette the Dy647-siLuc NPs solution vigorously before transferring; Also, wash the tube one or two times with a small volume of PBS. |
121
|
Low fluorescence signals of Dy647 |
The concentration of Dy647-siLuc NPs is too low or the cell density is too high |
Raise the concentration of Dy647-siLuc NPs or decrease the cell density. |
131
|
Low bioluminescence signal |
The bioluminescence signal decays |
Prepare and warm up the imaging system before adding the substrate. Upon adding the substrate, finish the measurement within 10 min. |
Low gene silencing efficiency |
G0-C14 is degraded |
The G0-C14 can be easily degraded after dissolution in the solvent. Aliquot the G0-C14 solution and store aliquots at −20 °C. Do not use the same batch more than three times. |
152
|
All the samples are PI positive |
The PI staining process is too long |
Shorten the PI staining process; Finish the measurement within 1 h after the PI staining |
165
|
The centrifugation takes too much time |
The concentration of siRNA NPs is too high, making the ultrafiltration difficult |
Split the samples into more Amicon tubes for washing, combine and transfer all the samples to one Amicon tube to adjust the concentration of NPs for in vivo injection |
166
|
The volume of the NPs is too large for injection |
The siRNA concentration is too low |
Add less PBS to redisperse the NPs during the preparation of NPs |
176
|
Weak fluorescence signal of the sections |
Low activity of the primary antibody or the primary antibody concentration is too low |
Handle the primary antibodies on ice to retain their optimal activity; Decrease the dilution factor of the primary antibodies. |