Skip to main content
. 2022 Mar 18;17(16):1340–1347. doi: 10.4244/EIJ-D-21-00419

Table 3. Incidence of device thrombosis up to 5-year follow-up.

Absorb BVS (n=924) XIENCE EES (n=921) Hazard ratio (95% CI) p-value
Definite 38 (4.3%) 9 (1.0%) 4.24 (2.05-8.77) <0.001
Probable 5 (0.5%) 4 (0.5%) 1.24 (0.33-4.62) 0.747
Possible 16 (1.8%) 25 (3.0%) 0.63 (0.34-1.18) 0.150
Definite/probable 43 (4.8%) 13 (1.5%) 3.32 (1.78-6.17) <0.001
≤24 hours (acute) 3 3
>24 hours to 30 days (subacute) 10 2
31 days to 1 year (late) 8 1
1-2 years (very late) 9 2
2-3 years (very late) 4 0
3-4 years (very late) 6 3
4-5 years (very late) 3 2
Any device thrombosis 58 (6.5%) 38 (4.4%) 1.53 (1.02-2.31) 0.039
p-values were calculated by the log-rank test. BVS: bioresorbable vascular scaffold; EES: everolimus-eluting stent