Skip to main content
. 2022 Dec 8;33:101408. doi: 10.1016/j.bbrep.2022.101408

Table 1.

Bioanalytical method summary for the analyte DRV and the internal standard RTV.

Optimized mass parameters
Analyte Transitions (m/z) CE (eV) CXP (eV)
Darunavir (DRV) 547.7/392 20 30
Ritonavir
721.3/296.1
27
29
Other parameters: DP: 100, EP: 10, CUR: 20, CAD: 8, IS: 5500, GS1: 50, GS2: 50, Temp: 500
Liquid chromatography conditions
HPLC Column: Waters Xterra® MS C18 Column, 125 Å, 3.5 μm,4.6 mm × 50 mm
Mobile Phase: acetonitrile and water with 0.1% formic acid Injection volume: 1 μl
Total flow: 1 ml/min
Gradient Time Organic solvent (%)
0.10 30
1.00 50
1.50 60
5.00 60
5.10 30

MS parameters are abbreviated as follows: Capillary electrophoresis (CE), Cell exit potential (CXP), Declustering potential (DP), Entrance potential (EP), Curtain gas (CUR), Collision gas (CAD), Ion spray velocity (IS), Ion source gas 1 (GS1), Ion source gas 2 (GS2), Temperature (Temp).